Skip to content

BiPAP AVAPS BiPAP AVAPS C Series, ITRecalls & reported problems

Respironics, Inc. · BiPAP AVAPS C Series, Italy · Class 2

All models of BiPAP AVAPS (13)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00606959048171), received between Oct 10, 2023 and Oct 10, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Degraded” (1 report).

1 other recall cover devices cleared under the same FDA submission (K102465); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1

Patient problems

  • Headache1
  • Skin Inflammation/ Irritation1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 10, 2023Degraded; Degraded

    THE MANUFACTURER BECAME AWARE OF A USER OF A BIPAP AVAPS C SERIES DEVICE WAS EXPERIENCING HEADACHES AND SKIN IRRITATION, ESPECIALLY OF THE HEAD SINCE (B)(6) 2021. THERE WAS NO REPORT OF VISIBLE PARTICLES SEEN OF ANY KIND. THE DEVICE IS CURRENTLY AT THE PATIENT'S HOME. THE PRODUCT HAS NOT BEEN RETURNED FOR INVESTIGATION. THE MANUFACTURER'S INVESTIGATION IS ON-GOING. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.

    Patient problems: Headache; Skin Inflammation/ Irritation

    MAUDE report 17908724 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K102465. One submission often covers many models, so check the product and lot details in each record.

  • Z-1973-2021 · Jun 14, 2021 · Open, Classified
    DreamStation ASV, DreamStation ST, AVAPS, SystemOne ASV4, C-Series ASV, C-Series S/T and AVAPS, OmniLab Advanced+ Continuous Ventilator, Non-life Supporting The Philips Respironic

Other VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING devices

All VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page