BiPAP A40 AUX3100S19Recalls & reported problems
Respironics, Inc. · BiPAP A40 Pro, Australia · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00606959059177), received between Sep 15, 2021 and Nov 28, 2024 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (9 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 9, 2025Open, Classified
This device does not indicate for use in patients with respiratory failure.
Z-2037-2025 · Root cause: Device Design · FDA record ↗
- Mar 26, 2024Open, Classified
Potential for devices to experience interruption/loss of therapy during a Ventilator Inoperative alarm. This may lead to hypoventilation, mild to severe hypoxemia, hypercarbia, respiratory failure/insufficiency, or potentially death in the most vulnerable patients.
Z-1813-2024 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Electrical /Electronic Property Problem9
- Unexpected Shutdown2
- No Device Output1
- Noise, Audible1
- Therapeutic or Diagnostic Output Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- Hypoventilation1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 28, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. PRIOR TO IDENTIFYING THE ISSUE DESCRIBED IN FSN 2023-CC-SRC-039, COMPLAINTS ASSOCIATED WITH THE ISSUE DID NOT MEET REQUIREMENTS FOR VIGILANCE REPORTING AS A REPORTABLE INCIDENT. AS PART OF THE HHE PROCESS, A RETROSPECTIVE REVIEW OF COMPLAINTS ASSOCIATED WITH THE ISSUE DESCRIBED IN FSN 2023-CC-SRC-039 WAS NECESSARY TO REASSESS REPORTABILITY. UPON THIS FURTHER ASSESSMENT, THESE COMPLAINTS ARE BEING REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20807126 ↗MalfunctionNov 28, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. PRIOR TO IDENTIFYING THE ISSUE DESCRIBED IN FSN 2023-CC-SRC-039, COMPLAINTS ASSOCIATED WITH THE ISSUE DID NOT MEET REQUIREMENTS FOR VIGILANCE REPORTING AS A REPORTABLE INCIDENT. AS PART OF THE HHE PROCESS, A RETROSPECTIVE REVIEW OF COMPLAINTS ASSOCIATED WITH THE ISSUE DESCRIBED IN FSN 2023-CC-SRC-039 WAS NECESSARY TO REASSESS REPORTABILITY. UPON THIS FURTHER ASSESSMENT, THESE COMPLAINTS ARE BEING REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20807106 ↗MalfunctionNov 20, 2024Unexpected Shutdown; Unexpected Shutdown
THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE DEVICE TURNED OFF DURING THERAPY. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. PRIOR TO IDENTIFYING THE ISSUE DESCRIBED IN FSN 2023-CC-SRC-039, COMPLAINTS ASSOCIATED WITH THE ISSUE DID NOT MEET REQUIREMENTS FOR VIGILANCE REPORTING AS A REPORTABLE INCIDENT. AS PART OF THE HHE PROCESS, A RETROSPECTIVE REVIEW OF COMPLAINTS ASSOCIATED WITH THE ISSUE DESCRIBED IN FSN 2023-CC-SRC-039 WAS NECESSARY TO REASSESS REPORTABILITY. UPON THIS FURTHER ASSESSMENT, THESE COMPLAINTS ARE BEING REPORTED. THE BIPAP A40 PRO IS SUBSTANTIALLY SIMILAR TO THE BIPAP A40 AND WILL BE REPORTED IN THE UNITED STATES UNDER BIPAP A40 PRO, 501K NUMBER: K121623.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20735676 ↗MalfunctionAug 5, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. THE BIPAP A40 PRO IS SUBSTANTIALLY SIMILAR TO THE BIPAP A40 AND WILL BE REPORTED IN THE UNITED STATES UNDER BIPAP A40, 501K NUMBER: K121623.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19911418 ↗MalfunctionAug 5, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. THE BIPAP A40 PRO IS SUBSTANTIALLY SIMILAR TO THE BIPAP A40 AND WILL BE REPORTED IN THE UNITED STATES UNDER BIPAP A40, 501K NUMBER: K121623 .
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19911270 ↗MalfunctionAug 5, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE. THE BIPAP A40 PRO AUX3100S19 IS SUBSTANTIALLY SIMILAR TO THE BIPAP A40 AND WILL BE REPORTED IN THE UNITED STATES UNDER BIPAP A40, K121623.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19904965 ↗MalfunctionFeb 23, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE AND A LOW-PRESSURE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE VENTILATOR WAS REPLACED WITH AN ALTERNATIVE DEVICE. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18772950 ↗MalfunctionFeb 15, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR INOPERATIVE CONDITION OCCURRED. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18711647 ↗MalfunctionDec 8, 2023No Device Output; No Device Output
THE MANUFACTURER RECEIVED INFORMATION ALLEGING THAT VENTILATOR DEVICE WAS PERIODICALLY CUTTING OUT. THERE WAS NO HARM OR INJURY REPORTED. THE DEVICE HAS NOT BEEN RETURNED TO THE MANUFACTURER. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18295387 ↗MalfunctionDec 5, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING DEVICE RESTARTING BY ITSELF EVERY 5-15 MINUTES. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18268916 ↗MalfunctionNov 30, 2023Noise, Audible; Unexpected Shutdown; Noise, Audible; Unexpected Shutdown
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A DEVICE GAVE AUDITORY TONE AND THEN STOPPED WHILE IN USE. THERE WAS NO HARM OR INJURY REPORTED. THE MANUFACTURER'S INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18243464 ↗MalfunctionNov 1, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THE MANUFACTURER RECEIVED INFORMATION ALLEGING THE DEVICE TURNED ITSELF OFF DURING NIGHT FOR ABOUT 30 SECONDS, THEN RESUMED THERAPY. PATIENT WOKE GASPING FOR AIR. PATIENT DESCRIBED IT AS A DANGEROUS SITUATION AS HIS LUNG CAPACITY CONTINUES TO DROP HE RELIES HEAVILY ON THE DEVICE FOR UNINTERRUPTED SLEEP. NO MEDICAL INTERVENTION SPECIFIED. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER'S SERVICE CENTER FOR EVALUATION AND IT WAS FOUND THAT DEVICE RECORDED AN O2 FAN FAILURE ALARM. SERVICE CENTER DETERMINED THAT ONCE THE POWER SUPPLY WAS RE-CONNECTED, THE FAN CONTINUED TO RUN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18048686 ↗InjurySep 15, 2021Therapeutic or Diagnostic Output Failure
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A BIPAP A40 PRO WAS INADEQUATELY VENTILATING THE PATIENT AND THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR HYPERCAPNIA. THE DEVICE HAS YET TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. AT THIS TIME, WE ARE UNABLE TO CONFIRM THE ALLEGED MALFUNCTION. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE MANUFACTURER'S INVESTIGATION IS COMPLETE.
Patient problems: Hypoventilation
MAUDE report 12474510 ↗
Other Ventilator, continuous, minimal ventilatory support, facility use devices
- Philips Respironics V60Respironics California, Inc. · Ventilator, Continuous, Minimal Ventilatory Support,Facility Use3 recalls30 deaths13,525 reports
- PHILIPS 1076716Respironics California, Inc. · Ventilator, continuous, minimal ventilatory support, facility use2 recalls1 death2,946 reports
- PHILIPS 1076709Respironics California, Inc. · Ventilator, continuous, minimal ventilatory support, facility use8 deaths2,174 reports
- PHILIPS 1053614Respironics California, Inc. · Ventilator, continuous, minimal ventilatory support, facility use2 recalls4 deaths1,203 reports
- BiPAP A40 EEX3100S19Respironics, Inc. · Ventilator, continuous, minimal ventilatory support, facility use2 recalls422 reports
- BiPAP A40 DEX3100S13Respironics, Inc. · Ventilator, continuous, minimal ventilatory support, facility use2 recalls164 reports
- BiPAP A40 ITX3100S21Respironics, Inc. · Ventilator, continuous, minimal ventilatory support, facility use2 recalls163 reports
- PHILIPS 1053613Respironics California, Inc. · Ventilator, continuous, minimal ventilatory support, facility use2 recalls1 death161 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.