This DreamStation Auto CPAP with humidifier is a breathing machine. Reports mostly describe degraded foam and dust inside the unit. There are 7,041 reports, including 44 deaths and 513 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 5,793
- Deaths reported
- 44
- Injuries reported
- 513
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5,793 adverse event reports that list this exact device (DI 00606959424159), received between Nov 7, 2022 and Sep 30, 2026 — classified as 44 deaths, 513 injuries, 5,236 malfunctions.
The most frequently reported specific device problem is “Degraded” (5,758 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded5,758
- Contamination33
- Adverse Event Without Identified Device or Use Problem27
- Material Integrity Problem6
- No Apparent Adverse Event5
- Smoking5
- Contamination /Decontamination Problem4
- Mechanical Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions5,117
- Unspecified Respiratory Problem354
- Cancer225
- Headache215
- Dizziness207
- Insufficient Information133
- Asthma124
- Unspecified Kidney or Urinary Problem122
Recent adverse event reports
The 25 most recent of 5,793 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Degraded
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953072 ↗MalfunctionSep 25, 2026Degraded
A DreamStation Auto CPAP device was returned to a third-party service center in reference to the voluntary Field Safety Notice / recall notification related to the sound abatement foam in certain CPAP, BiPAP, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and visible foam particles were observed. In addition, foam particles inside the blower kit and dust were found in contamination findings. Additionally, the device was scrapped.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50898014 ↗MalfunctionJul 7, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THIRD PARTY EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND FOAM PARTICLES, FLUID FAIL WERE OBSERVED INSIDE THE ASSEMBLY. ADDITIONALLY, THE DEVICE WAS SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25740019 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. DURING THE EVALUATION, THERE WAS DUST CONTAMINATION FOUND IN THE DEVICE. THE DEVICE WAS SCRAPPED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25673834 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25672705 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED DUST CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25671271 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THERE WAS NO INITIAL ALLEGED COMPLAINT FOUND. DURING THE EVALUATION CONDUCTED AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM PARTICLES WAS IDENTIFIED. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND DUST CONTAMINATION AND THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25670794 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25670716 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED, AND DUST CONTAMINATION WAS FOUND IN THE DEVICE. THE DEVICE WAS SCRAPPED. A FINAL REPORT IS BEING SUBMITTED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25670305 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO THE THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE AT THE SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND FOUND EVIDENCE OF FOAM DEGRADATION. DURING THE EVALUATION, FOAM PARTICLES WERE OBSERVED, AND DUST CONTAMINATION WAS FOUND IN THE DEVICE. THE DEVICE WAS SCRAPPED. A FINAL REPORT IS BEING SUBMITTED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25670303 ↗MalfunctionJun 30, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM DEGRADATION WAS FOUND. ADDITIONALLY, THERE WAS NO POWER AS TECHNICAL FINDING, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25670109 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE TECHNICIAN OBSERVED DUST FAIL CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25669986 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR SERVICE AS PART OF THE SOUND ABATEMENT FOAM RECALL PROCESS. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES INSIDE THE BLOWER KIT. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE SERVICE TECHNICIAN OBSERVED FLUID CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25669973 ↗MalfunctionJun 30, 2026Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND TRASH CONTAMINATION. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25669255 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THERE WAS NO INITIAL ALLEGED COMPLAINT FOUND. DURING THE EVALUATION CONDUCTED AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM PARTICLES WAS IDENTIFIED. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND DUST CONTAMINATION AND THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED. THIS EVENT IS REPORTABLE UNDER FDA 21 CFR PART 803 AS IT MEETS THE CRITERIA FOR A SERIOUS INJURY AND/OR PRODUCT MALFUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668876 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING EVALUATION AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS OBSERVED INSIDE THE BLOWER KIT. ADDITIONALLY, UNRELATED TO THE REPORTABLE MALFUNCTION, THE SERVICE TECHNICIAN NOTED FLUIDS CONTAMINATION. FOLLOWING COMPLETION OF THE EVALUATION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER. THERE WAS NO REPORT OF DEATH, SERIOUS OR PERMANENT HARM, INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. THE OBSERVED FLUIDS CONTAMINATION WAS UNRELATED TO THE REPORTABLE MALFUNCTION AND WAS NOT DETERMINED TO HAVE CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. BASED ON THE IDENTIFICATION OF FOAM DEGRADATION/PARTICLES INSIDE THE BLOWER KIT DURING DEVICE EVALUATION, THIS EVENT WAS DETERMINED TO BE REPORTABLE UNDER FDA 21 CFR PART 803 AS A PRODUCT MALFUNCTION AND/OR POTENTIAL SERIOUS INJURY EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668873 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER FOR EVALUATION. DURING EVALUATION AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM PARTICLES WAS OBSERVED INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN NOTED HUMIDITY CONTAMINATION. FOLLOWING COMPLETION OF THE EVALUATION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER. THERE WAS NO REPORT OF DEATH, SERIOUS OR PERMANENT HARM, INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. BASED ON THE IDENTIFICATION OF FOAM PARTICLES INSIDE THE BLOWER KIT DURING DEVICE EVALUATION, THIS EVENT WAS DETERMINED TO BE REPORTABLE UNDER FDA 21 CFR PART 803 AS A PRODUCT MALFUNCTION AND/OR POTENTIAL SERIOUS INJURY EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668871 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING EVALUATION AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION/PARTICLES WAS OBSERVED INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN NOTED DUST CONTAMINATION. FOLLOWING COMPLETION OF THE EVALUATION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER. THERE WAS NO REPORT OF DEATH, SERIOUS OR PERMANENT HARM, INJURY, OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. BASED ON THE IDENTIFICATION OF FOAM DEGRADATION/PARTICLES INSIDE THE BLOWER KIT DURING DEVICE EVALUATION, THIS EVENT WAS DETERMINED TO BE REPORTABLE UNDER FDA 21 CFR PART 803 AS A PRODUCT MALFUNCTION AND/OR POTENTIAL SERIOUS INJURY EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668860 ↗MalfunctionJun 30, 2026Degraded
THE MANUFACTURER WAS CONTACTED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER WITH NO INITIAL ALLEGED COMPLAINT. DURING THE EVALUATION OF THE DEVICE AT THE THIRD-PARTY SERVICE CENTER THE DEVICE WAS VISUALLY INSPECTED AND EVIDENCE OF FOAM DEGRADATION WAS FOUND INSIDE THE BLOWER KIT. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND FLUID CONTAMINATION. IN ADDITION, THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25668832 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP WAS RETURNED TO A THIRD-PARTY SERVICE CENTER. THERE WAS NO INITIAL ALLEGED COMPLAINT FOUND. DURING THE EVALUATION CONDUCTED AT THE THIRD-PARTY SERVICE CENTER, THE DEVICE WAS VISUALLY INSPECTED, AND EVIDENCE OF FOAM PARTICLES WAS IDENTIFIED. ADDITIONALLY, THE SERVICE TECHNICIAN FOUND DUST CONTAMINATION AND THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25666969 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25666683 ↗MalfunctionJun 30, 2026Degraded
A DREAMSTATION AUTO CPAP DEVICE WAS RETURNED IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. DURING THE EVALUATION OF THE DEVICE, THE SERVICE TECHNICIAN OBSERVED VISIBLE FOAM PARTICLES. ADDITIONALLY, THE SERVICE TECHNICIAN ALSO NOTED CONTAMINATION. THE DEVICE WAS SCRAPPED BY THE SERVICE CENTER AFTER THE EVALUATION WAS COMPLETED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25666659 ↗MalfunctionJun 30, 2026Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND AN UNSPECIFIED FLUID CONTAMINATION. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25665312 ↗MalfunctionJun 30, 2026Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND DUST CONTAMINATION. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE FILED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25665137 ↗MalfunctionJun 30, 2026Degraded
A DEVICE WAS RETURNED TO A THIRD-PARTY SERVICE CENTER IN REFERENCE TO THE VOLUNTARY FIELD SAFETY NOTICE / RECALL NOTIFICATION RELATED TO THE SOUND ABATEMENT FOAM IN CERTAIN CPAP, BIPAP, AND MECHANICAL VENTILATOR DEVICES. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE, THE THIRD-PARTY SERVICE CENTER VISUALLY INSPECTED THE DEVICE AND FOUND EVIDENCE OF FOAM DEGRADATION AND DUST AN UNSPECIFIED FLUID CONTAMINATION. ADDITIONALLY, THE EVALUATION OF THE DEVICE DATA IDENTIFIED UNRELATED ERROR CODES. THESE ERROR CODES ARE CONSISTENT WITH NORMAL DEVICE OPERATION AND EXPECTED USE CONDITIONS AND ARE NOT CONSIDERED REPORTABLE UNDER MEDICAL DEVICE REPORTING REQUIREMENTS. NO EVIDENCE WAS IDENTIFIED TO SUGGEST THAT THESE ERROR CODES CONTRIBUTED TO OR CAUSED THE REPORTED EVENT. THE DEVICE HAS BEEN SCRAPPED. AT THIS TIME, NO FURTHER INVESTIGATION CAN BE PERFORMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25664632 ↗
Other VENTILATOR, NON-CONTINUOUS (RESPIRATOR) devices
- REMstar System One 60 Series REMstar Auto A-Flex, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)117 deaths10,075 reports
- DreamStation DreamStation Auto BiPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)106 deaths5,937 reports
- REMstar System One 60 Series REMstar Auto A-Flex System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)64 deaths5,138 reports
- REMstar System One 60 Series REMstar Pro C-Flex+, w/Humid System One 60 Series USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)65 deaths4,520 reports
- REMstar System One 60 Series REMstar Auto A-Flex w/HT Humid System One 60 Series, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)43 deaths4,149 reports
- DreamStation DreamStation Auto CPAP w/Humid w/Heated Tube, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)30 deaths3,981 reports
- DreamStation DreamStation Auto CPAP w/Humidifier, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)86 deaths3,516 reports
- DreamStation DreamStation Auto CPAP w/Humid, w/H Tube, w/Cell, USRespironics, Inc. · VENTILATOR, NON-CONTINUOUS (RESPIRATOR)52 deaths2,424 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.