Alexis® Orthopaedic Protector HR104Recalls & reported problems

Applied Medical Resources Corporation · Flexible Alexis Orthopaedic Protector · Class 2

All models of Alexis Orthopaedic Protector (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00607915138950), received between Apr 24, 2024 and Feb 20, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 3 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Split, Cut or Torn5
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 20, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    PROCEDURE PERFORMED: PTH VOIE ANTERIEUR. EVENT DESCRIPTION: THE ALEXIS WAS IN POSITION, AFTER USING THE SHARP TOOLS LIKE THE RASP, THE ALEXIS TEARED FROM THE SIDE. THE PATIENT WAS OBESE, SO ADDITIONAL FORCES WAS NEEDED. ADDITIONAL INFORMATION RECEIVED FROM APPLIED MEDICAL REPRESENTATIVE VIA 30JAN25: THE PATIENT'S STATUS IS GOOD, NO IMPACT WHATSOEVER, AND NO INJURY. INTERVENTION: NO ADDITIONAL INTERVENTION WAS NEEDED. PATIENT STATUS: THE PATIENT'S STATUS IS GOOD, NO IMPACT WHATSOEVER, AND NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21426704 ↗
  • MalfunctionJul 3, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    PROCEDURE PERFORMED: ANTERIOR HIP ARTHROPLASTY. DESCRIPTION OF EVENT: COMPLAINT 1 OF 3: (B)(4) - MFR# 2027111-2024-00706; COMPLAINT 2 OF 3: (B)(4); COMPLAINT 3 OF 3: (B)(4). ANTERIOR HIP ARTHROPLASTY. THE ALEXIS ORTHO WAS PLACED ACCORDING TO THE IFU UNDER THE TFL AND THE SARTORIUS. AFTER REMOVING THE ALEXIS ORTHO AT THE END OF THE PROCEDURE, THERE WAS A TEAR IN THE SHEATH OF THE ALEXIS ORTHO. PRODUCT TEST WITH (NAME REDACTED) FOR AN ANTERIOR HIP ARTHROPLASTY AT [FACILITY] IN [CITY]. THE SURGEON FINISHED THE SURGERY WITHOUT COMPLICATIONS. THE TEAR IN THE SHEATH WAS FIRST DISCOVERED AFTER THE SURGERY. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REP. VIA E-MAIL ON 19JUN2024: THE MALFUNCTION OCCURRED IN THE MIDDLE OF THE OR WHILE INTRODUCING THE ACETABULAR SOCKET IN TO THE PELVIC. THE REGULAR INSTRUMENTS USED FOR ANTERIOR HIP ARTHROPLASTIES WERE USED DURING THIS PROCEDURE TOGETHER WITH THE ALEXIS WERE BREACHER, REAMER, RETRACTORS. THE INSTRUMENTS WERE USED AS NORMAL DURING ANTERIOR HIP ARTHROPLASTIES. WE ARE NOT SURE IF THE DEVICE PARTICULATED, THE SURGEON MEANT THAT THE ALEXIS PARTICULATED AND PART OF THE ALEXIS SHEATH WAS NOW IMPLANTED ON THE ACETABULAR SOCKET. BUT THIS WAS JUST HIS

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19671622 ↗
  • MalfunctionJun 21, 2024Material Fragmentation; Material Fragmentation

    PROCEDURE PERFORMED: ANTERIOR HIP ARTHROPLASTY. DESCRIPTION OF EVENT: COMPLAINT 1 OF 3: (B)(4) - MFR# 2027111-2024-00706. COMPLAINT 2 OF 3: (B)(4) - MFR# 2027111-2024-00728. COMPLAINT 3 OF 3: (B)(4) - MFR# 2027111-2024-00726. THE ALEXIS ORTHO WAS PLACED ACCORDING TO THE IFU UNDER THE TFL AND THE SARTORIUS. DURING THE PROCEDURE, WHILE INTRODUCING THE ACETABULAR CUP, THE SURGEON GOT IN CONTACT WITH THE SHEATH OF THE ALEXIS WHICH CAUSED A TEAR IN THE PRODUCT. PRODUCT TEST WITH (NAME REDACTED) FOR AN ANTERIOR HIP ARTHROPLASTY AT [FACILITY] IN ZÜRICH. THE SURGEON CONTINUED AND FINISHED THE PROCEDURE ACCORDINGLY WITHOUT ANY COMPLICATIONS. ADDITIONAL INFORMATION RECEIVED FROM COMPANY REP. VIA E-MAIL ON 19JUN24: THE PATIENT IS FINE. THE MALFUNCTION OCCURRED IN THE MIDDLE OF THE OR WHILE INTRODUCING THE ACETABULAR SOCKET IN TO THE PELVIC. THE REGULAR INSTRUMENTS USED FOR ANTERIOR HIP ARTHROPLASTIES WERE USED DURING THIS PROCEDURE TOGETHER WITH THE ALEXIS BREACHER, REAMER, RETRACTORS, THE INSTRUMENTS WERE USED AS NORMAL DURING ANTERIOR HIP ARTHROPLASTIES WE ARE NOT SURE. THE SURGEON MEANT THAT THE ALEXIS PARTICULATED AND PART OF THE ALEXIS SHEATH WAS NOW IMPLANTED ON THE ACETABULAR SOCKET. B

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19586998 ↗
  • MalfunctionApr 24, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    PROCEDURE PERFORMED: TOTAL HIP ARTHROPLASTY. DESCRIPTION OF EVENT: [NAME] HOSPITAL. THIS IS A COMPLAINT FROM THE MARKET. REPORT FROM THE SALES REP. WHEN USING A REAMER, IT HIT THE PROTECTOR AND TORE THE SHEATH. CLINICAL SAMPLE AND NOT ACCUSTOMED TO USING THE PROTECTOR. SO, REAMER HIT IT. THIS UNIT WILL NOT BE RETURNED. DISCARDED AT HOSPITAL DUE TO INFECTION (HBS ANTIGEN +). TYPE OF INTERVENTION: THE PRODUCT WAS REMOVED AND SURGERY CONTINUED. PATIENT STATUS: NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19177276 ↗
  • MalfunctionApr 24, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    PROCEDURE PERFORMED: THA. DESCRIPTION OF EVENT: (B)(6) HOSPITAL. THIS IS A COMPLAINT FROM THE MARKET. REPORT FROM THE SALES REP VIA EMAIL; THE SEAM OF THE ALEXIS SHEATH RIPPED SEVERAL CENTIMETERS. THE SURGEON NOTICED THAT THE SHEATH WAS RIPPED WHEN HE REMOVED IT. NO OTHER INSTRUMENTS ARE IN CONTACT WITH THE SHEATH. SEE ATTACHED FOR A PHOTO OF THE SHEATH TEAR. THIS UNIT WILL NOT BE RETURNED. TYPE OF INTERVENTION: NA. PATIENT STATUS: NO IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19176054 ↗
  • MalfunctionApr 24, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    PROCEDURE PERFORMED: THA. DESCRIPTION OF EVENT: [NAME] HOSPITAL. THIS IS A COMPLAINT FROM THE MARKET. REPORT FROM THE SALES REP VIA EMAIL; THE SEAM OF THE ALEXIS SHEATH RIPPED SEVERAL CENTIMETERS. THE SURGEON NOTICED THAT THE SHEATH WAS RIPPED WHEN HE REMOVED IT. NO OTHER INSTRUMENTS ARE IN CONTACT WITH THE SHEATH. SEE ATTACHED FOR A PHOTO OF THE SHEATH TEAR. THIS UNIT WILL NOT BE RETURNED. TYPE OF INTERVENTION: NA. PATIENT STATUS: NO IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19176044 ↗

Other Ring (Wound Protector), Drape Retention, Internal devices

All Ring (Wound Protector), Drape Retention, Internal devices →