Skip to content

AVANOS 01-EC-459-1401Recalls & reported problems

Avanos Medical, Inc. · BALLARD* Liberator CS, Closed Suction System with manifold, 14 Fr · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00609038000138), received between Mar 31, 2021 and Dec 6, 2022 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Gas/Air Leak3
  • Fluid/Blood Leak2
  • Decrease in Suction1
  • Suction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Insufficient Information2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 6, 2022Decrease in Suction; Decrease in Suction

    THE RESPIRATORY THERAPIST (RT) MANAGER REPORTED, WHILE SUCTIONING WITH THE LIBERATOR CATHETER THEY DID NOT GET MUCH RETURN; THEY REMOVED THE LIBERATOR CLOSED SUCTION CATHETER (CSC) AND REPLACED IT WITH A BALLARD CSC AND REMOVED A LARGE AMOUNT OF SECRETIONS. THE RT MANAGER ADDITIONALLY REPORTED, THAT IN THE PROCESS OF CHANGING THE CATHETER, THE PATIENT STARTED WAKING UP AND COUGHED WHICH MAY HAVE LOOSENED SOME SECRETIONS CAUSING THEM TO BE EASIER TO SUCTION. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: Insufficient Information

    MAUDE report 15924402 ↗
  • MalfunctionDec 6, 2022Suction Problem; Suction Problem

    IT WAS REPORTED THE "SUCTION POWER SEEMS TO BE AN ISSUE WITH THE SET-UP WE ARE USING." THERE WAS NO REPORTED INJURY.

    Patient problems: Insufficient Information

    MAUDE report 15922112 ↗
  • MalfunctionMar 11, 2022Gas/Air Leak; Gas/Air Leak

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 3011270181-2022-00017 FOR THE SECOND REPORT IT WAS REPORTED THAT THE DEVICE MANIFOLD WAS "LEAKY" AND WAS REPLACED. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13732101 ↗
  • MalfunctionMar 11, 2022Gas/Air Leak; Gas/Air Leak

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 3011270181-2022-00016 FOR THE FIRST REPORT. IT WAS REPORTED THAT THE DEVICE MANIFOLD ON THE REPLACEMENT DEVICE WAS ALSO "LEAKY" AND WAS REPLACED AGAIN. THERE WAS NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13732090 ↗
  • MalfunctionFeb 4, 2022Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE RESPIRATORY THERAPIST WAS USING THE LIBERATOR MANIFOLD WITH THE LIBERATOR BRONCH ADAPTER AND NOTICED THE MANIFOLD HAD A LEAK AND IT SPRAYED WHEN IN-USE. THE RESPIRATORY THERAPIST AND THE ANESTHESIOLOGIST (WHO WAS IN THE PATIENT'S ROOM) BOTH HEARD THE LEAK AND WITNESSED THE SPRAY. IT APPEARED THE MANIFOLD WAS THE ISSUE, NOT THE BRONCH ADAPTER. THIS CONTINUED WHEN THE BRONCH ADAPTER WAS SWITCHED OUT FOR THE LIBERATOR CLOSED SUCTION CATHETER. THE PATIENT WAS ON A POSITIVE END-EXPIRATORY PRESSURE (PEEP) OF 15. THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13453246 ↗
  • MalfunctionMar 31, 2021Fluid/Blood Leak; Fluid/Blood Leak

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 3011270181-2021-00019 FOR THE SECOND REPORT. IT WAS REPORTED THAT THE LIBERATOR CSC [CLOSED SUCTION CATHETER] SLEEVE FILLED WITH SECRETIONS AND SALINE WITHIN A FEW HOURS OF USE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11599648 ↗
  • MalfunctionMar 31, 2021Fluid/Blood Leak; Fluid/Blood Leak

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF TWO REPORTS. REFER TO 3011270181-2021-00018 FOR THE FIRST REPORT IT WAS REPORTED THAT THE LIBERATOR CSC [CLOSED SUCTION CATHETER] SLEEVE FILLED WITH SECRETIONS AND SALINE WITHIN A FEW HOURS OF USE. THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11599612 ↗

FDA clearance and classification

The FDA regulates product code BSY (catheters, suction, tracheobronchial) under 21 CFR 868.6810 as a Class I device. 21 CFR 868.6810 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AVANOS 01-EC-459-1401

Has the AVANOS 01-EC-459-1401 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00609038000138), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AVANOS 01-EC-459-1401?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Mar 31, 2021 and Dec 6, 2022. In those reports, the most common specific device problems are “Gas/Air Leak” (3), “Fluid/Blood Leak” (2) and “Decrease in Suction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AVANOS 01-EC-459-1401?

None of the reports are classified as deaths, 0 as injuries and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AVANOS 01-EC-459-1401 FDA approved?

Devices of this type (product code BSY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Avanos Medical, Inc.

Who makes the AVANOS 01-EC-459-1401, and is it still sold?

The FDA's device database lists Avanos Medical, Inc. as the company responsible for it (brand AVANOS, model 01-EC-459-1401). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices

All CATHETERS, SUCTION, TRACHEOBRONCHIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page