Skip to content

AVANOS B70811Recalls & reported problems

Avanos Medical, Inc. · Replacement Catheter for Neonates / Pediatrics · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00609038001043), received between Dec 31, 2022 and Dec 31, 2022 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Material Integrity Problem2
  • Gas/Air Leak1
  • Insufficient Flow or Under Infusion1

Patient problems

  • Insufficient Information2
  • Low Oxygen Saturation2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 31, 2022Material Integrity Problem; Material Integrity Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FOURTH OF FOUR REPORTS. REFER TO 8030647-2022-00237 FOR THE FIRST REPORT. REFER TO 8030647-2022-00238 FOR THE SECOND REPORT. REFER TO 8030647-2022-00239 FOR THE THIRD REPORT . IT WAS REPORTED, THE MANIFOLD CRACKED ON THE BALLARD, IN THE CRITICAL CARE UNIT (CCU), AND IT CAUSED INADEQUATE AIRFLOW. THERE WAS NO REPORTED INJURY TO THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 16081255 ↗
  • MalfunctionDec 31, 2022Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    (B)(4) RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF FOUR REPORTS. REFER TO 8030647-2022-00237 FOR THE FIRST REPORT. REFER TO 8030647-2022-00238 FOR THE SECOND REPORT. REFER TO 8030647-2022-00240 FOR THE FOURTH REPORT. IT WAS REPORTED, THE MANIFOLD CRACKED ON THE BALLARD, IN THE CRITICAL CARE UNIT (CCU), AND CAUSED INADEQUATE AIRFLOW. THERE WAS NO REPORTED INJURY TO THE PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 16081254 ↗
  • MalfunctionDec 31, 2022Material Integrity Problem; Material Integrity Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF FOUR REPORTS. REFER TO 8030647-2022-00238 FOR THE SECOND REPORT. REFER TO 8030647-2022-00239 FOR THE THIRD REPORT. REFER TO 8030647-2022-00240 FOR THE FOURTH REPORT. THE NEONATAL INTENSIVE CARE UNIT (NICU) NURSE MANGER REPORTED, THE BALLAD WAS ON THE CHILD AND THEIR OXYGEN WAS AT 80% WHEN THEY NOTICED THE MANIFOLD WAS CRACKED, (THE MANIFOLD HAD BEEN REPLACED ON (B)(6) 2022 AND THE CATHETER WAS CHANGED ON (B)(6) 2022); THEY REPLACED IT.

    Patient problems: Low Oxygen Saturation

    MAUDE report 16081253 ↗
  • MalfunctionDec 31, 2022Gas/Air Leak; Gas/Air Leak

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF FOUR REPORTS. REFER TO 8030647-2022-00237 FOR THE FIRST REPORT, REFER TO 8030647-2022-00239 FOR THE THIRD REPORT, REFER TO 8030647-2022-00240 FOR THE FOURTH REPORT. IT WAS REPORTED, THEY REPLACED ¿IT¿ [PART OF THE VENT CIRCUIT] WITH A NEW ONE AND THE OXYGEN LEVELS WENT DOWN TO 40%. ADDITIONAL INFORMATION RECEIVED 12DEC2022 REPORTED, THE SUCTION BALLARD AND MANIFOLD WERE REPLACED, AND THE PROBLEM CORRECTED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 16081252 ↗

FDA clearance and classification

The FDA regulates product code BSY (catheters, suction, tracheobronchial) under 21 CFR 868.6810 as a Class I device. 21 CFR 868.6810 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the AVANOS B70811

Has the AVANOS B70811 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00609038001043), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the AVANOS B70811?

The FDA's MAUDE database lists 4 adverse event reports for this device, received on Dec 31, 2022. In those reports, the most common specific device problems are “Material Integrity Problem” (2), “Gas/Air Leak” (1) and “Insufficient Flow or Under Infusion” (1); the most common patient problem is “Low Oxygen Saturation” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the AVANOS B70811?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the AVANOS B70811 FDA approved?

Devices of this type (product code BSY) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Avanos Medical, Inc.

Is the AVANOS B70811 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the AVANOS B70811, and is it still sold?

The FDA's device database lists Avanos Medical, Inc. as the company responsible for it (brand AVANOS, model B70811). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices

All CATHETERS, SUCTION, TRACHEOBRONCHIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page