Skip to content

Ballard* Mini-BAL Sampling Catheter 11139Recalls & reported problems

Avanos Medical, Inc. · BALLARD* Mini-BAL Sampling Catheter · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00609038111391), received between Sep 21, 2021 and Sep 21, 2021 — classified as 1 injury.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • Bradycardia1
  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 21, 2021Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE PATIENT WAS UNDERGOING A MINI BAL [ALVEOLAR LAVAGE] PROCEDURE ON THE RIGHT LUNG WHEN THE PATIENT'S HEART RATE DROPPED AND VENTILATOR VOLUMES WOULD NOT RETURN TO PREPROCEDURAL VOLUMES. ET [ENDOTRACHEAL] TUBE HAD TO BE REMOVED FROM PATIENT'S AIRWAY. AT THAT TIME THE CSC [CLOSED SUCTION CATHETER] TIP WAS DETERMINED TO BE MISSING. IT WAS LATER LOCATED AND REMOVED. ACCORDING TO THE USER FACILITY, "FOLLOWED PROCEDURE GUIDELINES, 100% FIO2 [OXYGEN CONCENTRATION] SXN [SUCTION] PRIOR, TURN SUCTION PRESSURE DOWN TO 60. ADVANCED MINI BAL CATHETER INSIDE PT [PATIENT] ETT [ENDOTRACHEAL TUBE], TILL NUMBERS MATCHED. ADVANCING INNER CATHETER WITH INSPIRATION. DURING INSTILLATION 30ML OF SALINE WAS ADVANCED TO R [RIGHT] SIDE, TURNED STOPCOCK WITH NO RETURNED SPUTUM, EXTRA PRESSURE NOTED. PT STARTED TO BRADY TO 51. VOLUMES ON VENTILATOR NOT RETURNING. IMMEDIATELY WITHDREW ETT, ALLOWING PT TO RECOVER. DURING THIS TIME WITH CATHETER WITHDRAWN NOTED TIP OF CATHETER MISSING. FOUND MISSING CATHETER TIP PULLED BACK INTO CATHETER." ADDITIONAL INFORMATION RECEIVED 02-SEP-2021 INDICATED THE MINI BAL PROCEDURE WAS PERFORMED FOR BRONCHIAL SAMPLING. NO INTERVENTIONS WERE NEEDED; THE PAT

    Patient problems: Bradycardia; Foreign Body In Patient

    MAUDE report 12509684 ↗

Sources and updates