Ballard* Turbo-Cleaning Closed Suction System 92481Recalls & reported problems
Avanos Medical, Inc. · Turbo-Cleaning Closed Suction Catheter, T-Piece · Class 1
All models of Ballard* Turbo-Cleaning Closed Suction System (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00609038924816), received between Apr 14, 2022 and Feb 19, 2025 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Suction Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Suction Problem4
- Increase in Suction3
Patient problems
- Insufficient Information7
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2025Increase in Suction; Increase in Suction
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 8030647-2025-00017 FOR THE SECOND REPORT. REFER TO 8030647-2025-00018 FOR THE THIRD REPORT. IT WAS REPORTED, WHEN THE SUCTION ACTIVATION BUTTON WAS PRESSED TO APPLY SUCTION, IT REMAINED PRESSED, RESULTING IN CONTINUOUS SUCTION; THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 21413013 ↗MalfunctionFeb 19, 2025Increase in Suction; Increase in Suction
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 8030647-2025-00016 FOR THE FIRST REPORT. REFER TO 8030647-2025-00017 FOR THE SECOND REPORT. IT WAS REPORTED, WHEN THE SUCTION ACTIVATION BUTTON WAS PRESSED TO APPLY SUCTION, IT REMAINED PRESSED, RESULTING IN CONTINUOUS SUCTION; THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 21412996 ↗MalfunctionFeb 19, 2025Increase in Suction; Increase in Suction
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 8030647-2025-00016 FOR THE FIRST REPORT. REFER TO 8030647-2025-00018 FOR THE THIRD REPORT. IT WAS REPORTED, WHEN THE SUCTION ACTIVATION BUTTON WAS PRESSED TO APPLY SUCTION, IT REMAINED PRESSED, RESULTING IN CONTINUOUS SUCTION; THERE WAS NO REPORTED INJURY.
Patient problems: Insufficient Information
MAUDE report 21412992 ↗MalfunctionApr 14, 2022Suction Problem; Suction Problem
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FOURTH OF FOUR REPORTS. REFER TO 8030647-2022-00116 FOR THE FIRST REPORT. REFER TO 8030647-2022-00117 FOR THE SECOND REPORT. REFER TO 8030647-2022-00118 FOR THE THIRD REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED.
Patient problems: Insufficient Information
MAUDE report 14115934 ↗MalfunctionApr 14, 2022Suction Problem; Suction Problem
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF FOUR REPORTS. REFER TO 8030647-2022-00116 FOR THE FIRST REPORT. REFER TO 8030647-2022-00118 FOR THE THIRD REPORT. REFER TO 8030647-2022-00119 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED.
Patient problems: Insufficient Information
MAUDE report 14115933 ↗MalfunctionApr 14, 2022Suction Problem; Suction Problem
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF FOUR REPORTS. REFER TO 8030647-2022-00117 FOR THE SECOND REPORT. REFER TO 8030647-2022-00118 FOR THE THIRD REPORT. REFER TO 8030647-2022-00119 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED.
Patient problems: Insufficient Information
MAUDE report 14115923 ↗InjuryApr 14, 2022Suction Problem; Suction Problem
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE THIRD OF FOUR REPORTS. REFER TO 8030647-2022-00116 FOR THE FIRST REPORT. REFER TO 8030647-2022-00117 FOR THE SECOND REPORT. REFER TO 8030647-2022-00119 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED. ADDITIONALLY, IT WAS REPORTED THAT AN UNSPECIFIED ADVERSE EVENT OCCURRED; HOWEVER, NO MEDICAL INTERVENTION WAS NEEDED.
Patient problems: Insufficient Information
MAUDE report 14115915 ↗
Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices
- Ballard* Closed Suction System 93826Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall70 reports
- Ballard* Closed Suction System 92431Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL78 reports
- Ballard* Turbo-Cleaning Closed Suction System 98965Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Medline DYNCPDS14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Ballard* Turbo-Cleaning Closed Suction System 98263Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall11 reports
- Ballard* Closed Suction System for Adults 98310Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall1 death9 reports
- Medline DYNCPTP14TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall6 reports
- Medline DYNCPTP14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page