Ballard* Turbo-Cleaning Closed Suction System 92481Recalls & reported problems

Avanos Medical, Inc. · Turbo-Cleaning Closed Suction Catheter, T-Piece · Class 1

All models of Ballard* Turbo-Cleaning Closed Suction System (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00609038924816), received between Apr 14, 2022 and Feb 19, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Suction Problem4
  • Increase in Suction3

Patient problems

  • Insufficient Information7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2025Increase in Suction; Increase in Suction

    SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 8030647-2025-00017 FOR THE SECOND REPORT. REFER TO 8030647-2025-00018 FOR THE THIRD REPORT. IT WAS REPORTED, WHEN THE SUCTION ACTIVATION BUTTON WAS PRESSED TO APPLY SUCTION, IT REMAINED PRESSED, RESULTING IN CONTINUOUS SUCTION; THERE WAS NO REPORTED INJURY.

    Patient problems: Insufficient Information

    MAUDE report 21413013 ↗
  • MalfunctionFeb 19, 2025Increase in Suction; Increase in Suction

    SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 8030647-2025-00016 FOR THE FIRST REPORT. REFER TO 8030647-2025-00017 FOR THE SECOND REPORT. IT WAS REPORTED, WHEN THE SUCTION ACTIVATION BUTTON WAS PRESSED TO APPLY SUCTION, IT REMAINED PRESSED, RESULTING IN CONTINUOUS SUCTION; THERE WAS NO REPORTED INJURY.

    Patient problems: Insufficient Information

    MAUDE report 21412996 ↗
  • MalfunctionFeb 19, 2025Increase in Suction; Increase in Suction

    SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 8030647-2025-00016 FOR THE FIRST REPORT. REFER TO 8030647-2025-00018 FOR THE THIRD REPORT. IT WAS REPORTED, WHEN THE SUCTION ACTIVATION BUTTON WAS PRESSED TO APPLY SUCTION, IT REMAINED PRESSED, RESULTING IN CONTINUOUS SUCTION; THERE WAS NO REPORTED INJURY.

    Patient problems: Insufficient Information

    MAUDE report 21412992 ↗
  • MalfunctionApr 14, 2022Suction Problem; Suction Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FOURTH OF FOUR REPORTS. REFER TO 8030647-2022-00116 FOR THE FIRST REPORT. REFER TO 8030647-2022-00117 FOR THE SECOND REPORT. REFER TO 8030647-2022-00118 FOR THE THIRD REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 14115934 ↗
  • MalfunctionApr 14, 2022Suction Problem; Suction Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE SECOND OF FOUR REPORTS. REFER TO 8030647-2022-00116 FOR THE FIRST REPORT. REFER TO 8030647-2022-00118 FOR THE THIRD REPORT. REFER TO 8030647-2022-00119 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 14115933 ↗
  • MalfunctionApr 14, 2022Suction Problem; Suction Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF FOUR REPORTS. REFER TO 8030647-2022-00117 FOR THE SECOND REPORT. REFER TO 8030647-2022-00118 FOR THE THIRD REPORT. REFER TO 8030647-2022-00119 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED.

    Patient problems: Insufficient Information

    MAUDE report 14115923 ↗
  • InjuryApr 14, 2022Suction Problem; Suction Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE THIRD OF FOUR REPORTS. REFER TO 8030647-2022-00116 FOR THE FIRST REPORT. REFER TO 8030647-2022-00117 FOR THE SECOND REPORT. REFER TO 8030647-2022-00119 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE IN-LINE SUCTION CATHETER WAS NOT APPLYING VACUUM WHEN THE SUCTION BUTTON WAS PRESSED; IT APPEARED THAT THE HOLE IN THE BUTTON TO ALLOW VACUUM TO BE GENERATED WAS NOT THERE. THERE WAS NO REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED ON 23MAR2022 REPORTED, THAT THE CATHETER WAS REPLACED. ADDITIONALLY, IT WAS REPORTED THAT AN UNSPECIFIED ADVERSE EVENT OCCURRED; HOWEVER, NO MEDICAL INTERVENTION WAS NEEDED.

    Patient problems: Insufficient Information

    MAUDE report 14115915 ↗

Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices

All CATHETERS, SUCTION, TRACHEOBRONCHIAL devices →