Ballard* Turbo-Cleaning Closed Suction System 94701Recalls & reported problems
Avanos Medical, Inc. · Turbo-Cleaning Closed Suction Catheter, T-Piece · Class 1
All models of Ballard* Turbo-Cleaning Closed Suction System (16)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Avanos Ballard Turbo-Cleaning closed suction catheter is used to clear secretions from the airway through a breathing tube. The 33 reports mostly describe the connection coming loose and popping off from the patient, sometimes repeatedly during a shift. No deaths and no injuries were reported, and no recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00609038947013), received between Sep 14, 2020 and Dec 21, 2020 — classified as 33 malfunctions.
The most frequently reported specific device problem is “Disconnection” (33 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Disconnection33
Patient problems
- No Consequences Or Impact To Patient33
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 21, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT PATIENTS. THIS IS THE FOURTH OF FOUR REPORTS. REFER TO 8030647-2020-00119 FOR THE FIRST REPORT; REFER TO 8030647-2020-00120 FOR THE SECOND REPORT ; REFER TO 8030647-2020-00121 FOR THE THIRD REPORT. IT WAS REPORTED THAT THE CLOSED SUCTION CATHETER HAD A LOOSE CONNECTION AND "POPPED OFF" AT PATIENT CONNECTION. THERE WAS ALARM AWARENESS AND THE CATHETER WAS REAPPLIED OR SWITCHED OUT FOR ANOTHER LIKE CLOSED SUCTION CATHETER. NO PATIENT INJURY, OR ADVERSE EVENTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049558 ↗MalfunctionDec 21, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT PATIENTS. THIS IS THE FIRST OF FOUR REPORTS. REFER TO 8030647-2020-00120 FOR THE SECOND REPORT. REFER TO 8030647-2020-00121 FOR THE THIRD REPORT. REFER TO 8030647-2020-00122 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE CLOSED SUCTION CATHETER HAD A LOOSE CONNECTION AND "POPPED OFF" AT PATIENT CONNECTION. THERE WAS ALARM AWARENESS AND THE CATHETER WAS REAPPLIED OR SWITCHED OUT FOR ANOTHER LIKE CLOSED SUCTION CATHETER. NO PATIENT INJURY OR ADVERSE EVENTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049542 ↗MalfunctionDec 21, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT PATIENTS. THIS IS THE SECOND OF FOUR REPORTS. REFER TO 8030647-2020-00119 FOR THE FIRST REPORT; REFER TO 8030647-2020-00121 FOR THE THIRD REPORT; REFER TO 8030647-2020-00122 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE CLOSED SUCTION CATHETER HAD A LOOSE CONNECTION AND "POPPED OFF" AT PATIENT CONNECTION. THERE WAS ALARM AWARENESS, AND THE CATHETER WAS REAPPLIED OR SWITCHED OUT FOR ANOTHER LIKE CLOSED SUCTION CATHETER. NO PATIENT INJURY, OR ADVERSE EVENTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049530 ↗MalfunctionDec 21, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED FOUR DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FOUR DIFFERENT PATIENTS. THIS IS THE THIRD OF FOUR REPORTS. REFER TO 8030647-2020-00119 FOR THE FIRST REPORT. REFER TO 8030647-2020-00120 FOR THE SECOND REPORT. REFER TO 8030647-2020-00122 FOR THE FOURTH REPORT. IT WAS REPORTED THAT THE CLOSED SUCTION CATHETER HAD A LOOSE CONNECTION AND "POPPED OFF" AT PATIENT CONNECTION. THERE WAS ALARM AWARENESS AND THE CATHETER WAS REAPPLIED OR SWITCHED OUT FOR ANOTHER LIKE CLOSED SUCTION CATHETER. NO PATIENT INJURY OR ADVERSE EVENTS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11049473 ↗MalfunctionDec 12, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED AT LEAST TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING DIFFERENT PATIENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 8030647-2020-00081 FOR THE SECOND REPORT. IT WAS REPORTED THAT THE SUCTION CATHETER KEEPS DISCONNECTING FROM THE PATIENT. DURING A 12 HOURS SHIFT THE NURSE ESTIMATES THAT THE SAME DEVICE DISCONNECTS 5-10 TIMES. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10995928 ↗MalfunctionNov 19, 2020Disconnection; Disconnection
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE FIRST OF FIVE REPORTS. REFER TO 8030647-2020-00093 FOR THE SECOND EVENT. REFER TO 8030647-2020-00094 FOR THE THIRD EVENT. REFER TO 8030647-2020-00095 FOR THE FOURTH EVENT. REFER TO 8030647-2020-00096 FOR THE FIFTH EVENT. IT WAS REPORTED THE CLOSED SUCTION CATHETER WAS POPPING OFF THE TRACH DURING USE AND POPPED OFF DURING USE ON A COVID-19 PATIENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10870668 ↗MalfunctionNov 19, 2020Disconnection; Disconnection
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE THIRD OF FIVE REPORTS. REFER TO 8030647-2020-00092 FOR THE FIRST EVENT. REFER TO 8030647-2020-00093 FOR THE SECOND EVENT. REFER TO 8030647-2020-00095 FOR THE FOURTH EVENT. REFER TO 8030647-2020-00096 FOR THE FIFTH EVENT. IT WAS REPORTED THE CLOSED SUCTION CATHETER WAS POPPING OFF THE TRACH DURING USE. THERE WAS NO REPORTED INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10870648 ↗MalfunctionNov 19, 2020Disconnection; Disconnection
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE SECOND OF FIVE REPORTS. REFER TO 8030647-2020-00092 FOR THE FIRST EVENT. REFER TO 8030647-2020-00094 FOR THE THIRD EVENT. REFER TO 8030647-2020-00095 FOR THE FOURTH EVENT. REFER TO 8030647-2020-00096 FOR THE FIFTH EVENT. IT WAS REPORTED THE CLOSED SUCTION CATHETER WAS POPPING OFF THE TRACH DURING USE. THERE WAS NO REPORTED INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10870626 ↗MalfunctionNov 19, 2020Disconnection; Disconnection
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE FIFTH OF FIVE REPORTS. REFER TO 8030647-2020-00092 FOR THE FIRST EVENT. REFER TO 8030647-2020-00093 FOR THE SECOND EVENT. REFER TO 8030647-2020-00094 FOR THE THIRD EVENT. REFER TO 8030647-2020-00095 FOR THE FOURTH EVENT. IT WAS REPORTED THE CLOSED SUCTION CATHETER WAS POPPING OFF THE TRACH DURING USE. THERE WAS NO REPORTED INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10870585 ↗MalfunctionNov 19, 2020Disconnection; Disconnection
AVANOS MEDICAL, INC. RECEIVED A SINGLE REPORT THAT REFERENCED FIVE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING FIVE DIFFERENT EVENTS. THIS IS THE FOURTH OF FIVE REPORTS. REFER TO 8030647-2020-00092 FOR THE FIRST EVENT. REFER TO 8030647-2020-00093 FOR THE SECOND EVENT. REFER TO 8030647-2020-00094 FOR THE THIRD EVENT. REFER TO 8030647-2020-00096 FOR THE FIFTH EVENT. IT WAS REPORTED THE CLOSED SUCTION CATHETER WAS POPPING OFF THE TRACH DURING USE. THERE WAS NO REPORTED INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10870579 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE ELEVENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT . REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10536867 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE TWELFTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10536860 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE THIRTEENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10536853 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE FOURTEENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10536845 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE TENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT . REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT . REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPOR
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535517 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE EIGHTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT.REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535515 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE NINTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535514 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED AT LEAST TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING DIFFERENT PATIENTS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 8030647-2020-00080 FOR THE FIRST REPORT. IT WAS REPORTED THAT THE SUCTION CATHETER KEEPS DISCONNECTING FROM THE PATIENT. DURING A 12 HOURS SHIFT THE NURSE ESTIMATES THAT THE SAME DEVICE DISCONNECTS 5-10 TIMES. NO PATIENT INJURY. ADDITIONAL INFORMATION REQUESTED BUT NOT YET PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535511 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED AT LEAST TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING DIFFERENT PATIENTS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 8030647-2020-00081 FOR THE SECOND REPORT. IT WAS REPORTED THAT THE SUCTION CATHETER KEEPS DISCONNECTING FROM THE PATIENT. DURING A 12 HOURS SHIFT THE NURSE ESTIMATES THAT THE SAME DEVICE DISCONNECTS 5-10 TIMES. NO PATIENT INJURY. ADDITIONAL INFORMATION REQUESTED BUT NOT YET PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535506 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE TWENTY-FIRST OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT.REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPO
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535472 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE TWENTIETH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535469 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE NINETEENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535418 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE SEVENTEENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT, REFER TO 8030647-2020-00060 FOR THE SECOND REPORT, REFER TO 8030647-2020-00061 FOR THE THIRD REPORT, REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT, REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT, REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT, REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT, REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT, REFER TO 8030647-2020-00067 FOR THE NINTH REPORT, REFER TO 8030647-2020-00068 FOR THE TENTH REPORT, REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT, REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT, REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT, REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT, REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT, REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT, REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT, REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT, REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT,
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535417 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE SIXTEENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00076 FOR THE EIGHTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535416 ↗MalfunctionSep 15, 2020Disconnection; Disconnection
AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED 21 DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING 21 DIFFERENT PATIENTS. THIS IS THE EIGHTEENTH OF TWENTY-ONE REPORTS. REFER TO 8030647-2020-00059 FOR THE FIRST REPORT. REFER TO 8030647-2020-00060 FOR THE SECOND REPORT. REFER TO 8030647-2020-00061 FOR THE THIRD REPORT. REFER TO 8030647-2020-00062 FOR THE FOURTH REPORT. REFER TO 8030647-2020-00063 FOR THE FIFTH REPORT. REFER TO 8030647-2020-00064 FOR THE SIXTH REPORT. REFER TO 8030647-2020-00065 FOR THE SEVENTH REPORT. REFER TO 8030647-2020-00066 FOR THE EIGHTH REPORT. REFER TO 8030647-2020-00067 FOR THE NINTH REPORT. REFER TO 8030647-2020-00068 FOR THE TENTH REPORT. REFER TO 8030647-2020-00069 FOR THE ELEVENTH REPORT. REFER TO 8030647-2020-00070 FOR THE TWELFTH REPORT. REFER TO 8030647-2020-00071 FOR THE THIRTEENTH REPORT. REFER TO 8030647-2020-00072 FOR THE FOURTEENTH REPORT. REFER TO 8030647-2020-00073 FOR THE FIFTEENTH REPORT. REFER TO 8030647-2020-00074 FOR THE SIXTEENTH REPORT. REFER TO 8030647-2020-00075 FOR THE SEVENTEENTH REPORT. REFER TO 8030647-2020-00077 FOR THE NINETEENTH REPORT. REFER TO 8030647-2020-00078 FOR THE TWENTIETH REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10535415 ↗
Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices
- Ballard* Closed Suction System 93826Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall70 reports
- Ballard* Closed Suction System 92431Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL78 reports
- Ballard* Turbo-Cleaning Closed Suction System 98965Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Medline DYNCPDS14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall16 reports
- Ballard* Turbo-Cleaning Closed Suction System 98263Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall11 reports
- Ballard* Closed Suction System for Adults 98310Avanos Medical, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall1 death9 reports
- Medline DYNCPTP14TMedline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall6 reports
- Medline DYNCPTP14Medline Industries, Inc. · CATHETERS, SUCTION, TRACHEOBRONCHIAL1 recall4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page