Skip to content

Ballard* Turbo-Cleaning Closed Suction System 98255Recalls & reported problems

Avanos Medical, Inc. · Turbo-Cleaning Closed Suction Catheter, Double Swivel Elbow - Adult · Class 1

All models of Ballard* Turbo-Cleaning Closed Suction System (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00609038982557), received between Jul 28, 2020 and Apr 22, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Material Integrity Problem1
  • Retraction Problem1

Patient problems

  • Aspiration/Inhalation1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 22, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED, THE CLOSED SUCTION SYSTEM BROKE WHILE IT WAS CONNECTED TO A PATIENT WHO WAS BEING MOBILIZED TO THE PRONE POSITION. THERE WAS NO REPORTED INJURY TO THE PATIENT. NO MEDICAL INTERVENTION WAS REQUIRED. ADDITIONAL INFORMATION RECEIVED 02APR2024, PER GOOGLE TRANSLATE REPORTED, "THE GUIDE BECAME FRAYED DURING INSTALLATION OF THE KTA WHEN REMOVING THE GUIDE. THE GUIDE DID NOT BREAK SHARPLY. THE WIRES THAT HAVE BROKEN ARE MORE LOCATED ON THE MIDDLE PART OF THE GUIDE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19152291 ↗
  • MalfunctionJan 18, 2024Break; Break

    IT WAS REPORTED, "ON WARNING SIGN: SPASTIC INTUBATED PATIENT, CALL TO THE DOCTOR BY THE NURSE." A PIECE OF PLASTIC WAS VISUALIZED IN THE INTUBATION TUBE. "MATERIAL INSPIRED BY PATIENT." FIBROSCOPY PERFORMED ON (B)(6) 2023 TO RETRIEVE RETAINED PIECE OF PLASTIC WHICH APPEARED TO CORRESPOND WITH THE VALVE ON THE CLOSED SUCTION CATHETER. "THE SYSTEM HAD BEEN IN PLACE FOR LESS THAN 72 HOURS AND RINSED WITH SALINE AS RECOMMENDED." THE CLOSED SUCTION SYSTEM WAS REMOVED; THERE WAS NO REPORTED INJURY TO THE PATIENT.

    Patient problems: Aspiration/Inhalation

    MAUDE report 18535609 ↗
  • MalfunctionJul 28, 2020Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE USER WAS UNABLE TO "REASSEMBLE" THE SUCTION CATHETER. THE SUCTION TUBE WAS "HALF BLOCKED" IN THE PATIENT'S TRACHEA AND BRONCHUS. THE PATIENT WAS THEN DISCONNECTED FROM THE VENT CIRCUIT TO REMOVE THE SYSTEM. THERE WERE NO OTHER CONSEQUENCES TO THE PATIENT. PER ADDITIONAL INFORMATION RECEIVED ON 15 JUL 2020, THE ISSUE WAS THAT THE CATHETER COULD NOT BE RETURNED TO THE ORIGINAL POSITION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10332368 ↗

Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices

All CATHETERS, SUCTION, TRACHEOBRONCHIAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page