The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00609038983110), received between Jun 18, 2021 and Jul 16, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2
  • Inflation Problem2
  • Detachment of Device or Device Component1
  • Loose or Intermittent Connection1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information2

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 16, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED: THE INLINE SUCTION CATHETER WAS PLACED ON THE PATIENT WHILE ATTACHED TO THE VENTILATOR. [THE] PATIENT INDICATED SOMETHING WASN'T NORMAL WITH HIS BREATHING AND IT WAS NOTED THAT AIR WAS LEAVING OUT OF THE OPENING WHERE FLUSH/LAVAGE TUBING IS NORMALLY ATTACHED. [THE] PATIENT [WAS] GIVEN [A] NEW INLINE SUCTION [CATHETER]; THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22527447 ↗
  • MalfunctionAug 10, 2023Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED, THAT WHEN ATTEMPTING TO ATTACH THE [IN-LINE SUCTION CATHETER] TO A FRESH TRACH TUBE, THE IN-LINE WOULD NOT CONNECT PROPERLY, AND A LEAK KEPT HAPPENING. AFTER EXAMINING THE IN-LINE, IT WAS FOUND THAT THE PLASTIC WAS THE WRONG SIZE NOT ALLOWING THE CATHETER TO ATTACH PROPERLY TO TRACHEOSTOMY TUBE. THERE WAS NO REPORTED INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17503495 ↗
  • MalfunctionJan 10, 2023Inflation Problem; Inflation Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE SECOND OF TWO REPORTS. REFER TO 8030647-2023-00003 FOR THE FIRST REPORT. IT WAS REPORTED THAT THE IN-LINE CATHETER WAS PLACED IN THE EVENING. THE NEXT DAY, IT INFLATED LIKE A BALLOON. NO INJURY OR MEDICAL INTERVENTIONS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 16134502 ↗
  • MalfunctionJan 10, 2023Inflation Problem; Inflation Problem

    AVANOS MEDICAL INC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS. THIS IS THE FIRST OF TWO REPORTS. REFER TO 8030647-2023-00004 FOR THE SECOND REPORT. IT WAS REPORTED THAT THE IN-LINE CATHETER WAS PLACED IN THE MORNING. BY THE AFTERNOON, THE CATHETER INFLATED LIKE A BALLOON. NO INJURY OR MEDICAL INTERVENTIONS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 16134501 ↗
  • MalfunctionOct 30, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED, "THE INSIDE PIECE OF PLASTIC (THAT ATTACHES TO THE TRACH) CAME UNATTACHED FROM THE OUTER PLASTIC PIECE (THAT SWIVELS). THE INNER PORTION WAS STUCK ON THE PATIENT'S TRACH, AND TOOK SOME EFFORT TO GET IT OFF." THERE WAS NOT REPORTED INJURY. ADDITIONAL INFORMATION RECEIVED 13-OCT-2021 STATES THE DEVICE WAS IN PLACE FOR LESS THAN A WEEK. THE DEVICE WAS NOT ALTERED FROM ITS ORIGINAL FORM. RESPIRATORY THERAPY EXTRACTED THE BROKEN PIECE FROM THE BIVONA TTS TRACH TUBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12725035 ↗
  • MalfunctionJun 18, 2021Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT THE THUMB VALVE WAS STUCK IN THE DOWN POSITION RESULTING IN REDUCED VENTILATION TO THE PATIENT. THE VENTILATOR ALARMED AND NOTED A LOW CIRCUIT LEAK AND SHORTLY AFTER A LOW MINIMAL VENTILATION ALARM WAS TRIGGERED. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12027230 ↗

Other CATHETERS, SUCTION, TRACHEOBRONCHIAL devices

All CATHETERS, SUCTION, TRACHEOBRONCHIAL devices →