Hollister 9786Recalls & reported problems
Hollister Incorporated · Feeding tube attachment device- Accommodates 5Fr to 18Fr tubes · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00610075167760), received between Jul 5, 2023 and Jan 15, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Improper or Incorrect Procedure or Method1
- Inadequate or Insufficient Training1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Aspiration/Inhalation1
- Hypoxia1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 15, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT A HOLLISTER FEEDING TUBE ATTACHMENT DEVICE PEELED COMPLETELY AWAY FROM THE PATIENT'S SKIN DISLODGING THE NASOGASTRIC TUBE IT WAS SECURING. IT WAS REPORTED THAT THERE WAS NO INJURY BUT THE PATIENT REQUIRED A REINSERTION OF THE NG TUBE AND ADDITIONAL X-RAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21157448 ↗InjuryNov 14, 2023Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated; Improper or Incorrect Procedure or Method; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE TAPE ON THE HOLLISTER FEEDING TUBE ATTACHMENT DEVICE DID NOT STAY SECURED TO THE PATIENT'S NOSE AND THE NASOGASTRIC TUBE WAS ABLE TO DISLODGE. IT WAS FURTHER REPORTED THAT THE TAPE ADHERED TO THE NOSE FOR LESS THAN 24 HOURS. THE ACCOUNT WENT ON TO REPORT THAT THIS IMPACTED THE PATIENT'S LENGTH OF STAY AS THEY REQUIRED ANOTHER PROCEDURE TO INSERT A TUBE. THEY ALSO REPORTED THAT THE TUBE IS A NASOJEJUNAL TUBE FOR FEEDING AND REQUIRED INSERTION UNDER SEDATION AND UNTIL REINSERTED, THEY WERE NOT MEETING THEIR NUTRITIONAL REQUIREMENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18129514 ↗InjuryJul 5, 2023Inadequate or Insufficient Training; Inadequate or Insufficient Training
IT WAS REPORTED THAT A PATIENT WHO HAD A 12 FR FEEDING TUBE IN PLACE SECURED BY A HOLLISTER FEEDING TUBE ATTACHMENT DEVICE EXPERIENCED A TUBE DISLODGEMENT BECAUSE THE CLAMP DID NOT SECURELY HOLD THE TUBE IN PLACE AND THE TUBE WAS ABLE TO SLIDE THROUGH THE CLAMP. IT WAS REPORTED THAT THE DISLODGEMENT LED TO ASPIRATION LEADING TO RESPIRATORY DISTRESS, HYPOXIA, AND TACHYPNEA. IT WAS FURTHER REPORTED THAT CRACKLES WERE HEARD IN THE CHEST AND CONSOLIDATION IN LUNG WAS SEEN ON X-RAY. IT WAS REPORTED THAT THE PATIENT NEEDED REPEATED DEEP SUCTIONING, SEVERAL DAYS OF OXYGEN THERAPY, AND IV ANTIBIOTICS. THE ACCOUNT REPORTED THAT THE PATIENT RECOVERED FROM THE ASPIRATION EVENT BUT REMAINS IN THE HOSPITAL FOR OTHER REASONS. THE ACCOUNT REPORTED THAT THEY SWITCHED TO THE HOLLISTER TUBE ATTACHMENT DEVICE A COUPLE OF MONTHS AGO AND HAD NO EDUCATION OR TRAINING WHEN THIS SWITCH WAS MADE.
Patient problems: Aspiration/Inhalation; Hypoxia
MAUDE report 17255978 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-05055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-06055-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.