Skip to content

JACKSON REES 3771Y-WSNRecalls & reported problems

King Systems Corporation · JACKSON REES MODIFICATION · Class 1

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00612649188683), received between Sep 5, 2025 and Sep 5, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Restricted Flow rate” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Restricted Flow rate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2025Restricted Flow rate; Restricted Flow rate

    A BREATHING CIRCUIT (JACKSON-REES) WAS REPORTED TO EXHIBIT INCREASED RESISTANCE DURING CLINICAL USE. THE HEALTHCARE PROFESSIONAL NOTED THE BAG WAS NOT DEFLATING PROPERLY AND UPON INSPECTION, OBSERVED THAT THE TRIANGULAR OPENINGS ON THE APL (ADJUSTABLE PRESSURE-LIMITING) VALVE SEEMED TO BE PARTIALLY OCCLUDED, LIMITING AIRFLOW AND CAUSING INCREASED RESISTANCE. NO PATIENT INJURY OCCURRED. THE REPORT INDICATED THAT THIS CONDITION COULD LEAD TO ELEVATED PRESSURES THAT COULD POTENTIALLY RESULT IN LUNG DAMAGE IF NOT DETECTED TIMELY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22985603 ↗

Other Circuit, breathing (w connector, adaptor, y piece) devices

All Circuit, breathing (w connector, adaptor, y piece) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.