ROHO QUADTRO SELECT HIGH PROFILE Cushion QS1010CRecalls & reported problems
Roho, Inc. · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613732054304), received between Nov 13, 2025 and May 15, 2026 — classified as 3 injuries.
The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole2
- Insufficient Device Problem Information1
- Material Integrity Problem1
- Material Perforation1
Patient problems
- Pressure Sore/Ulcer3
- Tissue Breakdown1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 15, 2026Material Puncture/Hole
USER'S CAREGIVER CALLED THAT HER HUSBANDS ROHO QUAD SECECT CUSHION HAD STOPPED HOLDING AIR IN THE RIGHT REAR QUADRANT. DURING THE TIME THE CAREGIVER WAS AWAY FROM THE HOUSE FOR WORK, HER HUSBAND DEVELOPED A PRESSURE SORE FROM SITTING ON THE FLAT QUADRANT.
Patient problems: Pressure Sore/Ulcer
MAUDE report 25191060 ↗InjuryFeb 20, 2026Material Puncture/Hole; Material Perforation; Material Integrity Problem
END USER RECEIVED A NEW ROHO CUSHION THROUGH THEIR DEALER, NATIONAL SEATING AND MOBILITY. ON FIRST DAY OF USE, THE CUSHION DID NOT HOLD AIR. THE USER SAT ON THE IMPROPERLY INFLATED CUSHION ALL DAY AND DEVELOPED A STAGE 1 PRESSURE SORE, CAUSING HIM TO BE RESTRICTED TO BED REST AND MISS WORK. THE CUSHION WAS REPLACED AND THE SUSPECT CUSHION WAS RETURNED TO ROHO, INC. FOR INSPECTION. UPON INSPECTION, A MANUFACTURING DEFECT WAS IDENTIFIED WHICH LED TO THE LEAKING CUSHION. THE DEFECT OBSERVED IS CLASSIFIED AS A BLISTER. A BLISTER IS AN AIR POCKET WHICH FORMS DURING THE NEOPRENE FORMING PROCESS. BLISTERS ARE TYPICALLY FOUND DUIRNG THE INSPECTION PROCESS THROUGH VISUAL INSPECTIONS AND/OR THE OVERINFLATION TEST. OVER INFLATION TESTING IS DESIGNED TO SIMULATE THE PRESSURE OF A PERSON SITTING ON THEIR CUSHION, WHICH WILL TYPICALLY CAUSE BLISTERS TO RUPTURE AND LEAK.
Patient problems: Pressure Sore/Ulcer; Tissue Breakdown
MAUDE report 24400017 ↗InjuryNov 13, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
USER CALLED IN WITH A COMPLAINT THAT HIS QS1010C CUSHION IS LEAKING. HE HAS DEVELOPED A PRESSURE SORE FROM SITTING ON A FLAT CUSHION.
Patient problems: Pressure Sore/Ulcer
MAUDE report 23550220 ↗
Other Cushion, Flotation devices
- MEDLINE MSCBUB1818Medline Industries, Inc. · CUSHION, FLOTATION10 reports
- MEDLINE MSCBUB1818VLMedline Industries, Inc. · CUSHION, FLOTATION4 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS1011CRoho, Inc. · Cushion, Flotation3 reports
- Medline MSCBUB2822VLMedline Industries, Inc. · CUSHION, FLOTATION3 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS910CRoho, Inc. · Cushion, Flotation2 reports
- ROHO QUADTRO SELECT HIGH PROFILE Cushion QS89CRoho, Inc. · Cushion, Flotation2 reports
- MEDLINE MSCBUB2822Medline Industries, Inc. · CUSHION, FLOTATION2 reports
- ROHO HIGH PROFILE Single Compartment Cushion 1R109CRoho, Inc. · Cushion, Flotation2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.