Medtronic CareLink® Programmer 2090W (UDI 00613994127631)Recalls & reported problems
Medtronic, Inc. · PROGRAMMER 2090W Tel C UPGRADE US · Class 3
All models of Medtronic CareLink Programmer (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic CareLink 2090W programmer is used to program pacemakers. Many of its 208 reports describe screens that do not display or respond to finger touch, cursor movement on its own, and software problems. The reports include 2 injuries and no deaths. There are 3 recalls: 2 are open, covering cursor movement on the touchscreen and cybersecurity weaknesses in telemetry, and 1 is closed, covering an error in how battery life is displayed.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 208
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 3
Summary
The FDA's MAUDE database contains 208 adverse event reports that list this exact device (DI 00613994127631), received between Apr 6, 2017 and Aug 31, 2026 — classified as 2 injuries, 206 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (53 reports).
3 FDA recall records name this device's identifier, 2 still open.
10 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 11, 2024Open, Classified
Medtronic CareLink 2090 programmers with serial number prefixes PKK0 and PKK1 have the potential for autonomous cursor motion when Finger Touchscreen capability is enabled by software version 3.2 or higher.
Z-3152-2024 · Root cause: Device Design · Recall details · FDA record ↗
- Jan 30, 2020Open, Classified
Medtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access control and the cleartext transmission of sensitive information.
Z-1524-2020 · Root cause: Software design · Recall details · FDA record ↗
- Jan 17, 2019Terminated
There is an error in how the programmer calculates and displays the remaining longevity value during a period of time (up to 2 years) prior to the device reaching its Recommended Replacement Time (RRT).
Z-0795-2019 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- No Display/Image53
- Computer Operating System Problem51
- Device Difficult to Program or Calibrate51
- Computer Software Problem29
- Output Problem23
- Problem with Software Installation19
- Unexpected Shutdown19
- Display or Visual Feedback Problem17
Patient problems
- No Clinical Signs, Symptoms or Conditions89
- No Patient Involvement60
- No Known Impact Or Consequence To Patient55
- No Consequences Or Impact To Patient6
- No Code Available1
Recent adverse event reports
The 25 most recent of 208 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 31, 2026Unexpected Shutdown
It was reported that the programmer powered on without the Radiofrequency(RF) head being connected and operated normally. When the RF head was connected, the programmer immediately powered off. The power switch was cycled with the RF head remaining connected; however, the programmer did not power on. The power cord was verified to be securely connected. The RF head was then disconnected, after which the programmer powered on and functioned normally. A different RF head was subsequently connected to the programmer, and the programmer continued to operate normally without powering off. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50517300 ↗MalfunctionApr 29, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER CURSOR FAILED TO FOLLOW FINGER TOUCH MOVEMENTS AND JUMPED AROUND, CLICKING AUTONOMOUSLY WITHOUT PROMPTING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25024850 ↗MalfunctionFeb 20, 2026No Display/Image
IT WAS REPORTED THAT THE PROGRAMMER DISPLAY WAS UNRESPONSIVE TO FINGER TOUCH AFTER THE SOFTWARE UPDATE AND WAS ABLE TO WORK WITH THE STYLUS PEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24398144 ↗MalfunctionJan 27, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT DURING AN ELECTRICAL SAFETY TEST, THE PROGRAMMER FAILED THE RESISTANCE PART. MULTIPLE DIFFERENT POWER CORDS WERE TRIED, BUT THE ISSUE STILL PERSISTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24187043 ↗MalfunctionDec 9, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER FAILED THE FINGER TOUCH TEST FOR CURSOR MISALIGNMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23755978 ↗MalfunctionNov 12, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER FAILED THE FINGER TOUCH TEST FOR CURSOR MISALIGNMENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23529847 ↗MalfunctionNov 11, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER EXHIBITED AUTONOMOUS CURSOR MOVEMENT AFTER THE SOFTWARE UPDATE. THERE WAS NO PATIENT INVOLVEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23520038 ↗MalfunctionSep 18, 2025Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Display or Visual Feedback Problem; Device Difficult to Program or Calibrate
IT WAS REPORTED THAT THE PROGRAMMER'S STYLUS WAS UNABLE TO WORK INTERMITTENTLY OR DID NOT TRACK. IT WAS ALSO NOTED THAT THE TOUCHSCREEN UPDATE FAILED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23084299 ↗MalfunctionMar 11, 2025Display or Visual Feedback Problem; Display or Visual Feedback Problem
THE PROGRAMMER WAS RETURNED AS PART OF AN INVENTORY REDUCTION INITIATIVE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21574041 ↗MalfunctionOct 31, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER WAS BEEPING AND THE CURSOR WAS AUTONOMOUSLY MOVING FROM THE UPPER LEFT-HAND CORNER OF THE SCREEN TO THE RIGHT-HAND CORNER, BACK AND FORTH, AND AT ONE POINT MOVED DOWN TO THE BOTTOM OF THE SCREEN BEFORE RETURNING TO THE TOP LEFT CORNER. EVENTUALLY, IT STOPPED AND REMAINED IN PLACE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20576483 ↗MalfunctionOct 9, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER FAILED THE TOUCH TEST FOR CURSOR MISALIGNMENT AFTER A SOFTWARE UPDATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20406095 ↗MalfunctionAug 26, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS FURTHER REPORTED THAT THE RADIOFREQUENCY (RF) HEAD WAS RETURNED FOR EVALUATION AS AN ASSOCIATE PRODUCT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20062859 ↗MalfunctionAug 23, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER FAILED AN AUTONOMOUS CURSOR MOVEMENT TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20049422 ↗MalfunctionAug 22, 2024Display or Visual Feedback Problem; Failure to Interrogate; Display or Visual Feedback Problem; Failure to Interrogate
IT WAS REPORTED THAT THE PROGRAMMER FAILED THE TOUCHSCREEN TEST AFTER THE RECENT SOFTWARE UPDATE. IT WAS NOTED THAT THE PROGRAMMER ALSO HAD AN ISSUE WITH INTERROGATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20031124 ↗MalfunctionAug 20, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER¿S CURSOR WAS MOVED AND JUMPED RANDOMLY WITHOUT BEING TOUCHED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20002210 ↗MalfunctionAug 8, 2024Signal Artifact/Noise; Display or Visual Feedback Problem; Signal Artifact/Noise; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER DEMONSTRATED AUTONOMOUS CURSOR MOVEMENT WITHOUT INPUT FROM EITHER THE STYLUS PEN OR FINGER TOUCH AFTER THE SOFTWARE UPDATE VIA A UNIVERSAL SERIAL BUS (USB). IT WAS NOTED THAT THERE WAS NOISE ON THE ELECTROCARDIOGRAM (EKG) PORT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19940617 ↗MalfunctionJul 26, 2024Display or Visual Feedback Problem; Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PROGRAMMER FAILED CURSOR TESTING FOLLOWING AN UPDATE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19836148 ↗MalfunctionDec 12, 2023Overheating of Device; Connection Problem; Overheating of Device; Connection Problem
IT WAS REPORTED THAT THE VIDEO GRAPHICS ARRAY (VGA) VIDEO OUTPUT OF THE PROGRAMMER DID NOT WORK. THE PROGRAMMER HAS BEEN RETURNED FOR EVALUATION AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18308968 ↗MalfunctionOct 24, 2023No Display/Image; Device Difficult to Program or Calibrate; No Display/Image; Device Difficult to Program or Calibrate
IT WAS REPORTED THAT THE PROGRAMMER SCREEN WOULD NOT RESPOND TO THE STYLUS, THE INCORRECT ITEMS WERE OPENED AND THE CURSOR JUMPED AROUND THE SCREEN. THE PROGRAMMER HAS BEEN RETURNED FOR EVALUATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17995634 ↗MalfunctionOct 13, 2023Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
IT WAS REPORTED THAT THE PROGRAMMER WAS UNABLE TO RESPOND TO THE STYLUS. TROUBLESHOOTING STEPS INCLUDED A REBOOT OF THE PROGRAMMER AND RESETTING THE PROGRAMMER STYLUS PLUGIN WHICH RESOLVED THE ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17929386 ↗MalfunctionAug 18, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE PROGRAMMER WHICH WAS RETURNED FOR PREVENTATIVE MAINTENANCE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17570519 ↗MalfunctionMay 9, 2023No Display/Image; No Display/Image
IT WAS REPORTED THAT THE PROGRAMMER SCREEN WAS NOT RESPONDING TO THE STYLUS BY THE HEART RATE NUMBERS AND THE NUMBERS COULD NOT BE CHANGED. THE PROGRAMMER WAS RETUNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16893797 ↗MalfunctionApr 24, 2023No Display/Image; No Display/Image
THE PROGRAMMER WAS RETURNED AS PART OF AN INVENTORY REDUCTION INITIATIVE AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16793558 ↗MalfunctionMar 31, 2023Device Emits Odor; Overheating of Device; Device Emits Odor; Overheating of Device
IT WAS REPORTED THAT THE PROGRAMMERS MOTOR WAS SMOKING AND HAD A BURNING SMELL. THE PROGRAMMER WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16654457 ↗MalfunctionMar 21, 2023Overheating of Device; Failure to Power Up; Noise, Audible; Unexpected Shutdown; Overheating of Device; Failure to Power Up; Noise, Audible; Unexpected Shutdown
IT WAS REPORTED THAT THE PROGRAMMER MADE A NOISE AND SMOKE CAME OUT OF THE FAN AFTER POWERING ON. THE PROGRAMMER THEN POWERED OFF AND WAS UNABLE TO BE TURNED BACK ON. THE PROGRAMMER WAS RETURNED FOR SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16584095 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.
- Z-2169-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
- Z-2168-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2506-2020 · Jun 4, 2020 · Open, ClassifiedPatient Connector, Model Number 24967
- Z-2505-2020 · Jun 4, 2020 · Open, ClassifiedCareLink SmartSync Device Manager, Model Number 24970A
- Z-0845-2019 · Jan 17, 2019 · TerminatedMedtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Z-0399-2018 · Jul 14, 2017 · TerminatedMyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with
- Z-0400-2018 · Jul 14, 2017 · TerminatedMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
- Z-2209-2010 · Jun 11, 2010 · TerminatedMedtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w
Other Programmer, pacemaker devices
- Viva™ XT CRT-D DTBA1D1Medtronic, Inc. · Programmer, pacemaker1 recall25 deaths893 reports
- Medtronic CareLink® Programmer 2090WMedtronic, Inc. · Programmer, pacemaker3 recalls489 reports
- Viva™ Quad XT CRT-D DTBA1Q1Medtronic, Inc. · Programmer, pacemaker2 recalls6 deaths236 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker2 recalls198 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Programmer, pacemaker3 recalls108 reports
- Viva™ Quad S CRT-D DTBB1QQMedtronic, Inc. · Programmer, pacemaker2 recalls1 death38 reports
- Viva™ Quad XT CRT-D DTBA1QQMedtronic, Inc. · Programmer, pacemaker1 recall47 reports
- NA 2067LMedtronic, Inc. · Programmer, pacemaker1 recall45 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.