CRESCENT™ Spinal System 9393608Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SPACER 9393608 CRESCENT VERT BODY 36X8 · Class 2 · Implant
All models of CRESCENT Spinal System (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994187703), received between Apr 13, 2018 and Oct 26, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 26, 2020Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH UNKNOWN INDICATION FOR SPINAL THERAPY SUGGESTED FOR TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF). EVENT OCCURRED INTRA-OP. LEVELS IMPLANTED L4-5. IT WAS REPORTED THAT THERE WAS A BREAK IN INSTRUMENT AND FRAGMENTS WERE LEFT IN PATIENT. LEFT THE CAGE IN THE DISC SPACE REASONABLY PLACE AND WOULD LIKELY FUSE. IMPLANT REMAINS IN PATIENT. NO FURTHER SYMPTOMS OR COMPLICATIONS REPORTED. UPDATE 2020-OCT-13 E1(REP); THE CAGE BROKE WHILE IMPLANTING THE TIP WAS ON THE INSERTER, THE REST OF THE IMPLANT 90% OF IT WAS IN THE DISC SPACE SURGEON TAMPED IT IN FURTHER AND WAS HAPPY WITH THE MIDLINE FINAL POSITION OF THE IMPLANT. THE CRESCENT INSERTER DID NOT BREAK THE IMPLANT BROKE. CRESCENT PEEK CAGE WAS A PLANNED PART OF THE PROCEDURE. DECOMPRESS THE LEVEL AND FUSE THE LEVEL. PRE-OP DIAGNOSIS WAS CENTRAL STENOSIS AND A TLIF TO ADDRESS THE PATHOLOGY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10732632 ↗MalfunctionApr 13, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT TRANSFORAMINAL LUMBAR INTERBODY FUSION AT L4-5. INTRA-OP, THE CAGE BROKE DURING INSERTION AT THE L4-5 LEVEL. NO FRAGMENTS OF THE DEVICE REMAINED INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7430525 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.