ATLANTIS® Anterior Cervical Plate System 3120313Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 3120313 4.0 X 13 SELF TAP VAR · Class 2 · Implant
All models of ATLANTIS Anterior Cervical Plate System (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994239679), received between Dec 1, 2017 and Mar 21, 2018 — classified as 5 injuries.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (5 reports).
3 other recalls cover devices cleared under the same FDA submission (K063100); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration or Expulsion of Device5
Patient problems
- Neck Pain5
- Inadequate Pain Relief1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMar 21, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT C5/6 ANTERIOR CERVICAL DISCECTOMY FUSION. POST-OP, PATIENT HAD PAIN AFTER THE SURGERY. ALLEGEDLY, "PATIENT FEELS THAT THE SCREWS STICKING OUT OF HER NECK AS WELL AS WHEN SHE TOUCHES THAT AREA OF HER NECK IT THROBS".
Patient problems: Inadequate Pain Relief; Neck Pain
MAUDE report 7359029 ↗InjuryDec 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR DISCECTOMY AND FUSION AT C3-C7. POST-OP, THE LEVELS AT C4-5 AND C5-6 COLLAPSED AND THE IMPLANTED SCREWS AT C5-C6 LOOSENED AND CAME OUT OF THE PLATE EVEN THOUGH THE LOCKING MECHANISM WAS LOCKED. PATIENT SUFFERED DUE TO NECK PAIN. PATIENT UNDERWENT REVISION SURGERY (ANTERIOR CERVICAL DISCECTOMY AND FUSION) FOR REMOVAL OF THE IMPLANTS.
Patient problems: Neck Pain
MAUDE report 7074073 ↗InjuryDec 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR DISCECTOMY AND FUSION AT C3-C7. POST-OP, THE LEVELS AT C4-5 AND C5-6 COLLAPSED AND THE IMPLANTED SCREWS AT C3-C4 LOOSENED AND CAME OUT OF THE PLATE EVEN THOUGH THE LOCKING MECHANISM WAS LOCKED. ONE OF THE 13MM SCREWS COMPLETELY CAME OUT OF THE PLATE AND WAS FLOATING FREE IN THE PATIENT DUE TO WHICH PATIENT SUFFERED WITH NECK PAIN. THE PATIENT UNDERWENT REVISION SURGERY (ANTERIOR CERVICAL DISCECTOMY AND FUSION) TO REMOVE THE IMPLANTS.
Patient problems: Neck Pain
MAUDE report 7074072 ↗InjuryDec 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR DISCECTOMY AND FUSION AT C3-C7. POST-OP, THE LEVELS AT C4-5 AND C5-6 COLLAPSED AND THE IMPLANTED SCREWS AT C3-C4 LOOSENED AND CAME OUT OF THE PLATE EVEN THOUGH THE LOCKING MECHANISM WAS LOCKED. ONE OF THE 13MM SCREWS COMPLETELY CAME OUT OF THE PLATE AND WAS FLOATING FREE IN THE PATIENT DUE TO WHICH PATIENT SUFFERED WITH NECK PAIN. THE PATIENT UNDERWENT REVISION SURGERY (ANTERIOR CERVICAL DISCECTOMY AND FUSION) TO REMOVE THE IMPLANTS.
Patient problems: Neck Pain
MAUDE report 7074069 ↗InjuryDec 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR DISCECTOMY AND FUSION AT C3-C7. POST-OP, THE LEVELS AT C4-5 AND C5-6 COLLAPSED AND THE IMPLANTED SCREWS AT C3-C4 LOOSENED AND CAME OUT OF THE PLATE EVEN THOUGH THE LOCKING MECHANISM WAS LOCKED. ONE OF THE 13MM SCREWS COMPLETELY CAME OUT OF THE PLATE AND WAS FLOATING FREE IN THE PATIENT. PATIENT SUFFERED DUE TO NECK PAIN. PATIENT UNDERWENT REVISION SURGERY (ANTERIOR CERVICAL DISCECTOMY AND FUSION) TO REMOVE THE IMPLANTS.
Patient problems: Neck Pain
MAUDE report 7074068 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K063100. One submission often covers many models, so check the product and lot details in each record.
- Z-0037-2013 · Aug 30, 2012 · TerminatedATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38
- Z-0036-2013 · Aug 30, 2012 · TerminatedATLANTIS(R) Anterior Cervical Plate System VARIABLE ANGLE SCREW MODULE, REF 176-503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee
- Z-0038-2013 · Aug 30, 2012 · TerminatedPREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthoped
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page