ATLANTIS® Anterior Cervical Plate System 3120517Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 3120517 4.0 X 17 SELF DRILL VAR · Class 2 · Implant
All models of ATLANTIS Anterior Cervical Plate System (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994241160), received between Dec 1, 2017 and Aug 9, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K063100); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration or Expulsion of Device2
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- Neck Pain2
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 9, 2023Break; Break
INFORMATION WAS RECEIVED FROM A (HEALTHCARE PROFESSIONAL, RISK MANAGER, REGULATORY BODY) VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING A ANTERIOR CERVICAL DISCECTOMY AND FUSION. IT WAS REPORTED THAT THE SCREW BROKE. THE DEVICE WAS EXPLANTED AND THE PATIENT WAS HOSPITALIZED. THERE WERE NO PATIENTS SYMPTOMS OR FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17486318 ↗InjuryNov 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH NON UNION AND UNDERWENT ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) SURGERY AT C4-7 LEVELS. POST-OPERATIVELY, A SCREW WAS BROKEN IN THE BODY OF C7. AS A RESULT, THE BROKEN PORTION OF THE SCREW WAS CAPTURED AND REMOVED FROM THE C7 VERTEBRAL BODY. THE PRODUCTS CAME IN CONTACT WITH THE PATIENT. NO FRAGMENT OF THE SCREW REMAINED IN THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8109428 ↗InjuryDec 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR DISCECTOMY AND FUSION AT C3-C7. POST-OP, THE LEVELS AT C4-5 AND C5-6 COLLAPSED AND THE IMPLANTED SCREWS AT C7 LOOSENED AND CAME OUT OF THE PLATE EVEN THOUGH THE LOCKING MECHANISM WAS LOCKED. PATIENT SUFFERED WITH NECK PAIN AND HENCE UNDERWENT REVISION SURGERY (ANTERIOR CERVICAL DISCECTOMY AND FUSION) FOR REMOVAL OF THE IMPLANTS.
Patient problems: Neck Pain
MAUDE report 7074081 ↗InjuryDec 1, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT ANTERIOR DISCECTOMY AND FUSION AT C3-C7. POST-OP, THE LEVELS AT C4-5 AND C5-6 COLLAPSED AND THE IMPLANTED SCREWS AT C7 LOOSENED AND CAME OUT OF THE PLATE EVEN THOUGH THE LOCKING MECHANISM WAS LOCKED. PATIENT SUFFERED DUE TO NECK PAIN. THE PATIENT UNDERWENT REVISION SURGERY (ANTERIOR CERVICAL DISCECTOMY AND FUSION) TO REMOVE THE IMPLANTS.
Patient problems: Neck Pain
MAUDE report 7074071 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K063100. One submission often covers many models, so check the product and lot details in each record.
- Z-0038-2013 · Aug 30, 2012 · TerminatedPREMIER(R) Anterior Cervical Plate System SCREW BLOCK, REF 1860503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132. Orthoped
- Z-0037-2013 · Aug 30, 2012 · TerminatedATLANTIS(R) Anterior Cervical Plate System FIXED ANGLE SCREW MODULE, REF 176-504, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38
- Z-0036-2013 · Aug 30, 2012 · TerminatedATLANTIS(R) Anterior Cervical Plate System VARIABLE ANGLE SCREW MODULE, REF 176-503, QTY: 1 EA, Rx only, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page