CAPSTONE® Spinal System 2990826Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SPACER 2990826 CAPSTONE PEEK 08X26 · Class 2 · Implant

All models of CAPSTONE Spinal System (21)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994291035), received between Jul 27, 2020 and Nov 19, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 19, 2025Break; Break

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (HEALTHCARE PROVIDER, DISTRIBUTOR) REGARDING A PATIENT IMPLANTED WITH SPACER HAVING TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) THERAPY FOR L3-L4 HERNIATED INTERVERTEBRAL DISC AND GRADE 2 SPONDYLOLISTHESIS. IT WAS REPORTED THAT AFTER TRIALING, THE SURGEON FULLY INSERTED THE CAGE. WHEN ATTEMPTING TO IMPACT THE CAGE FURTHER INTO POSITION, THE SURGEON NOTED THAT THE CAGE HOLDER HAD DETACHED FROM THE CAGE. THE SURGEON REMOVED THE IMPLANT AND ASSOCIATED FRAGMENTS COMPLETELY. A NEW CAGE WAS THEN OPENED, IMPLANTED, AND THE PROCEDURE WAS COMPLETED UNEVENTFULLY, NO PATIENT SYMPTOM WERE REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23592113 ↗
  • MalfunctionSep 14, 2020Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH A CAGE FOR A LUMBAR FUSION. IT WAS REPORTED THAT THE POSTERIOR THIRD OF THE LUMBAR FUSION CAGE WAS BROKEN. THE IMPLANT BROKE AND THERE WAS NO FRAGMENT OF THE INSTRUMENT REMAINING IN THE PATIENT'S BODY. THE FUSION CAGE WAS BROKEN DURING THE SURGERY OF THE CUSTOMER DURING IMPLANTATION. THE PATIENT IS ASSOCIATED WITH THE EVENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THE EVENT. THE PRE- OP DIAGNOSIS WAS A LUMBAR DISC HERNIATION. DETAILS OF LEVELS IMPLANTED: IN THE FIRST IMPLANTATION, THE SURGEON HIT THE HOLDER WITH A BONE HAMMER. DURING THE PROCESS OF IMPLANTING THE FUSION CAGE, THE FUSION CAGE HAD NOT BEEN COMPLETELY IMPLANTED IN THE INTERVERTEBRAL SPACE. THERE WAS NO ADDITIONAL SURGERY OR TREATMENT PERFORMED. THE PRODUCT WAS EXPLANTED AND REPLACED WITH A MEDTRONIC PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10527258 ↗
  • MalfunctionJul 27, 2020Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH A CAGE FOR A LUMBAR DISCECTOMY. IT WAS REPORTED THAT WHEN THE SURGEON COMPLETED THE LUMBAR DISCECTOMY AND THEN PREPARED TO IMPLANT THE CAGE, HE PUT THE CAGE ON THE CAGE HOLDER AND PUNCHED INTO THE L4-5, THE CAGE WAS BROKEN AT THE CAGE HOLDER AND CAGE CONNECT PLACE. THE IMPLANT BROKE AND THERE WERE NO FRAGMENTS OF THE IMPLANT REMAINING IN THE PATIENT BODY. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THE EVENT. THE PRE-OP DIAGNOSIS WAS LUMBAR DDD. LEVELS IMPLANTED WERE L4-L5. DATE OF IMPLANT AND EXPLANT WAS (B)(6) 2020. THE DEVICE WAS EXPLANTED. THE PRODUCT WAS REPLACED BY A MEDTRONIC PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10324762 ↗