CD HORIZON® Spinal System 7896545Recalls & reported problems
Medtronic Sofamor Danek, Inc. · SCREW 7896545 CDH PEEK 6.5X45 NON CANN · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994291509), received between May 25, 2018 and Dec 12, 2018 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Migration or Expulsion of Device2
Patient problems
- Failure of Implant2
- No Known Impact Or Consequence To Patient2
- Pain2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 12, 2018Break; Break
TYPE OF PROCEDURE: LUMBAR FUSION LEVELS IMPLANTED: L2-S1 PRODUCT STATUS: PARTIALLY EXPLANTED IT WAS REPORTED THAT ON AN UNKNOWN DATE, POST-OP, THE SCREWS AT S1 BROKE DUE TO WHICH THERE WAS NON-FUSION AT L5-S1. THE PATIENT ALSO REPORTED BACK PAIN DUE TO THIS EVENT. ALL THE OTHER LEVELS OF THE OPERATED SPINE HAD FUSED. ON (B)(6) 2018, A REVISION WAS PERFORMED WHEREIN THE BROKEN SCREWS WERE PARTIALLY EXPLANTED. THE BROKEN PIECES OF SCREW REMAINED IN S1 AND COULD NOT BE EXPLANTED. THE SCREWS BROKE 8-10MM BELOW THE SURFACE OF THE BONE. TREPHINES AND CONICAL EXTRACTION BOLTS WERE READY BUT THE SURGEON THOUGHT IT WOULD BE SAFER AND LESS INVASIVE TO PLACE THE TWO NEW SCREWS NEXT TO THE BROKEN SCREWS IS A SAFE AREA OF S1. THE TWO NEW SCREWS WERE PLACED IN A SAFE AREA AND WERE WELL FIXED. A SECOND IMAGING WAS PERFORMED TO CONFIRM PROPER PLACEMENT OF THE TWO NEW SCREWS. NO COMPLICATIONS WERE REPORTED AFTER THE REVISION SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 8154321 ↗InjuryDec 12, 2018Break; Break
TYPE OF PROCEDURE: LUMBAR FUSION LEVELS. IMPLANTED: L2-S1. PRODUCT STATUS: PARTIALLY EXPLANTED. IT WAS REPORTED THAT ON AN UNKNOWN DATE, POST-OP, THE SCREWS AT S1 BROKE DUE TO WHICH THERE WAS NON-FUSION AT L5-S1. THE PATIENT ALSO REPORTED BACK PAIN DUE TO THIS EVENT. ALL THE OTHER LEVELS OF THE OPERATED SPINE HAD FUSED. ON (B)(6) 2018, A REVISION WAS PERFORMED WHEREIN THE BROKEN SCREWS WERE PARTIALLY EXPLANTED. THE BROKEN PIECES OF SCREW REMAINED IN S1 AND COULD NOT BE EXPLANTED. THE SCREWS BROKE 8-10MM BELOW THE SURFACE OF THE BONE. TREPHINES AND CONICAL EXTRACTION BOLTS WERE READY, BUT THE SURGEON THOUGHT IT WOULD BE SAFER AND LESS INVASIVE TO PLACE THE TWO NEW SCREWS NEXT TO THE BROKEN SCREWS IS A SAFE AREA OF S1. THE TWO NEW SCREWS WERE PLACED IN A SAFE AREA AND WERE WELL FIXED. A SECOND IMAGING WAS PERFORMED TO CONFIRM PROPER PLACEMENT OF THE TWO NEW SCREWS. NO COMPLICATIONS WERE REPORTED AFTER THE REVISION SURGERY.
Patient problems: Failure of Implant; Pain
MAUDE report 8154311 ↗MalfunctionMay 25, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT OPEN POSTERIOR LUMBAR INTER-BODY FUSION (PLIF) SURGERY AT L5-S1. POST-OP, S1 SCREW LOOSENING WAS OBSERVED. PER RADIOGRAPHIC REVIEW, THE PATIENT HAD A SUBTLE S1 SCREW LOOSENING THAT PROGRESSED ON FOLLOW-UP. PER INVESTIGATOR, THE SCREW LOOSENING HAS A ¿CAUSAL RELATIONSHIP¿ TO THE SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7543886 ↗MalfunctionMay 25, 2018Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT OPEN POSTERIOR LUMBAR INTER-BODY FUSION (PLIF) SURGERY AT L5-S1. POST-OP, S1 SCREW LOOSENING WAS OBSERVED. PER RADIOGRAPHIC REVIEW, THE PATIENT HAD A SUBTLE S1 SCREW LOOSENING THAT PROGRESSED ON FOLLOW-UP. PER INVESTIGATOR, THE SCREW LOOSENING HAS A ¿CAUSAL RELATIONSHIP¿ TO THE SCREW.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7543884 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page