CD HORIZON® Spinal System 7570905Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SLEEVE 7570905 REDUX INNER · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00613994302755), received between Feb 23, 2018 and May 15, 2025 — classified as 3 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Problem4
  • Bent3
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Known Impact Or Consequence To Patient4

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM MULTIPLE SOURCES (MANUFACTURER REPRESENTATIVE, USER FACILITY) REGARDING A PATIENT WHO HAD VERTEBRAL FR ACTURE DUE TO MOTOR VEHICLE ACCIDENT. IT WAS REPORTED THAT THE EXTENDER WON¿T LET THE SCREWDRIVER GO DOWN THE SHAFT. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. 2025 MAY 6, UPDATE RECEIVED: THE SCREWDRIVER DOES NOT FIT DOWN THE SHAFT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22024993 ↗
  • InjuryMay 2, 2024Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, THE EXTENDER AND SLEEVES DOES NOT HOLD THE SCREW. THE EXTERNAL SLEEVE OF THE LONGITUDE SYSTEM WITHOUT PROPER PRESSURE, TIGHTENING AND/OR TORQUE FOR FITTING THE PEDICLE SCREW. INNER SLEEVE OF THE LONGITUDE SYSTEM DILATED WITHOUT PROPER PRESSURE, TIGHTENING AND/OR TOUCH TO FIT THE PEDICLE SCREW, THE SCREW WAS LOOSE IN THE ASSEMBLED DEVICE, AND RISK OF THE SCREW ESCAPING AND TO FINALIZE THE PROPOSED SURGICAL TECHNIQUE. LONGITUDE SYSTEM DULY TESTED IN PRE-SURGERY AND INTERNAL OPERATIONAL PROCEDURES. HOWEVER, DURING THE INTRAOPERATIVE PERIOD, IT WAS IDENTIFIED THAT THE PRESSURE SUFFERED IN THE SURGICAL CAVITY CAUSED THE TOWERS TO DILATE IN ORDER TO SUPPORT THE LOAD OF THE PEDICLE SCREWS AND THE BIOMECHANICAL FORCE REQUIRED TO ASSEMBLE THE SYSTEM, WHICH LED THE SURGEON TO CHANGE THE SURGICAL STRATEGY DURING THE INTRAOPERATIVE PERIOD WITHOUT HARMING THE PATIENT. PROCEDURE WAS CHANGED FROM MINIMALLY INVASIVE SURGERY TO OPEN TECHNIQUE. NO PATIENT SYMPTOMS REPORTED, NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19227705 ↗
  • InjuryMay 2, 2024Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, THE EXTENDER AND SLEEVES DOES NOT HOLD THE SCREW. THE EXTERNAL SLEEVE OF THE LONGITUDE SYSTEM WITHOUT PROPER PRESSURE, TIGHTENING AND/OR TORQUE FOR FITTING THE PEDICLE SCREW. INNER SLEEVE OF THE LONGITUDE SYSTEM DILATED WITHOUT PROPER PRESSURE, TIGHTENING AND/OR TOUCH TO FIT THE PEDICLE SCREW, THE SCREW WAS LOOSE IN THE ASSEMBLED DEVICE, AND RISK OF THE SCREW ESCAPING AND TO FINALIZE THE PROPOSED SURGICAL TECHNIQUE. LONGITUDE SYSTEM DULY TESTED IN PRE-SURGERY AND INTERNAL OPERATIONAL PROCEDURES. HOWEVER, DURING THE INTRAOPERATIVE PERIOD, IT WAS IDENTIFIED THAT THE PRESSURE SUFFERED IN THE SURGICAL CAVITY CAUSED THE TOWERS TO DILATE IN ORDER TO SUPPORT THE LOAD OF THE PEDICLE SCREWS AND THE BIOMECHANICAL FORCE REQUIRED TO ASSEMBLE THE SYSTEM, WHICH LED THE SURGEON TO CHANGE THE SURGICAL STRATEGY DURING THE INTRAOPERATIVE PERIOD WITHOUT HARMING THE PATIENT. PROCEDURE WAS CHANGED FROM MINIMALLY INVASIVE SURGERY TO OPEN TECHNIQUE. NO PATIENT SYMPTOMS REPORTED, NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19227696 ↗
  • InjuryMay 2, 2024Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, THE EXTENDER AND SLEEVES DOES NOT HOLD THE SCREW. THE EXTERNAL SLEEVE OF THE LONGITUDE SYSTEM WITHOUT PROPER PRESSURE, TIGHTENING AND/OR TORQUE FOR FITTING THE PEDICLE SCREW. INNER SLEEVE OF THE LONGITUDE SYSTEM DILATED WITHOUT PROPER PRESSURE, TIGHTENING AND/OR TOUCH TO FIT THE PEDICLE SCREW, THE SCREW WAS LOOSE IN THE ASSEMBLED DEVICE, AND RISK OF THE SCREW ESCAPING AND TO FINALIZE THE PROPOSED SURGICAL TECHNIQUE. LONGITUDE SYSTEM DULY TESTED IN PRE-SURGERY AND INTERNAL OPERATIONAL PROCEDURES. HOWEVER, DURING THE INTRAOPERATIVE PERIOD, IT WAS IDENTIFIED THAT THE PRESSURE SUFFERED IN THE SURGICAL CAVITY CAUSED THE TOWERS TO DILATE IN ORDER TO SUPPORT THE LOAD OF THE PEDICLE SCREWS AND THE BIOMECHANICAL FORCE REQUIRED TO ASSEMBLE THE SYSTEM, WHICH LED THE SURGEON TO CHANGE THE SURGICAL STRATEGY DURING THE INTRAOPERATIVE PERIOD WITHOUT HARMING THE PATIENT. PROCEDURE WAS CHANGED FROM MINIMALLY INVASIVE SURGERY TO OPEN TECHNIQUE. NO PATIENT SYMPTOMS REPORTED, NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19227671 ↗
  • MalfunctionDec 29, 2019Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LUMBAR ARTHRODESIS SURGERY DUE TO SPINE PAIN. INTRA-OP, THE INSTRUMENT WAS DAMAGED, SO THE SCREW DID NOT FIT PROPERLY. THE PART OF THE INSTRUMENT THAT FITS THE SCREW WAS BENT, WHICH DID NOT ALLOW THE PERFECT FIT OF THE SAME, MAKING THE IMPLANTATION IN THE PATIENT UNFEASIBLE. NO ANY FRAGMENTS OF THE IMPLANT REMAINING IN THE PATIENT BODY. NO PATIENT COMPLICATIONS WERE REPORTED DUE TO THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9528859 ↗
  • MalfunctionFeb 23, 2018Bent; Bent

    IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH DEGENERATIVE DISC DISEASE AND UNDERWENT MINIMALLY INVASIVE LUMBAR DECOMPRESSION AND FUSION AT L3-S1. DURING SURGERY, THE SLEEVES EACH ¿POPPED¿ OFF WHEN SURGEON WAS ATTEMPTING TO REDUCE ROD INTO SCREW HEAD. THE SLEEVES APPEAR BENT. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS THE RESULT OF THE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7291593 ↗
  • MalfunctionFeb 23, 2018Bent; Bent

    IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH DEGENERATIVE DISC DISEASE AND UNDERWENT MINIMALLY INVASIVE LUMBAR DECOMPRESSION AND FUSION AT L3-S1. DURING SURGERY, THE SLEEVES EACH ¿POPPED¿ OFF WHEN SURGEON WAS ATTEMPTING TO REDUCE ROD INTO SCREW HEAD. THE SLEEVES APPEAR BENT. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS THE RESULT OF THE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7291592 ↗
  • MalfunctionFeb 23, 2018Bent; Bent

    IT WAS REPORTED THAT THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH DEGENERATIVE DISC DISEASE AND UNDERWENT MINIMALLY INVASIVE LUMBAR DECOMPRESSION AND FUSION AT L3-S1. DURING SURGERY, THE SLEEVES EACH ¿POPPED¿ OFF WHEN SURGEON WAS ATTEMPTING TO REDUCE ROD INTO SCREW HEAD. THE SLEEVES APPEAR BENT. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS THE RESULT OF THE EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7291589 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →