N/A 9733450 (UDI 00613994378477)Recalls & reported problems
Medtronic Navigation, Inc. · CURVED SUCTION 70 9733450 EM ENT · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00613994378477), received between Dec 15, 2016 and Apr 21, 2025 — classified as 16 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (6 reports).
28 other recalls cover devices cleared under the same FDA submission (K001284); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem6
- Material Deformation4
- Device Operates Differently Than Expected3
- Communication or Transmission Problem2
- Device Sensing Problem2
- Break1
- Imprecision1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Consequences Or Impact To Patient5
- No Known Impact Or Consequence To Patient4
- No Patient Involvement4
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE INACCURACY WAS 0.2MM OR LESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21875320 ↗MalfunctionDec 27, 2024Device Sensing Problem; Device Sensing Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THIS SITE WAS HAVING ISSUES VERIFYING ALL THREE SUCTIONS. ALL OTHER INSTRUMENTS VERIFY WITHOUT ISSUE. NO PATIENT WAS PRESENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21026241 ↗MalfunctionDec 13, 2024Use of Device Problem; Device Sensing Problem; Use of Device Problem; Device Sensing Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE INSTRUMENTS COULD NO LONGER BE REFERENCED. IT WAS NOTED THAT THEY WERE "PROBABLY BENT." THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20926439 ↗MalfunctionJun 17, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS). IT WAS REPORTED THAT THE 70 DEGREE SUCTION WAS BENT BY THE HEALTH CARE PROFESSIONAL INSIDE OF THE PATIENT ANATOMY. THERE WAS NO DELAY TO THE PROCEDURE. NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8704100 ↗MalfunctionJun 17, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM EVENT HAVING OCCURRED OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT A DAMAGED STRAIGHT SUCTION WAS RECEIVED IN MATERIALS. NO PATIENT PRESENT.
Patient problems: No Patient Involvement
MAUDE report 8702692 ↗MalfunctionJun 13, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED FOR A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE. IT WAS REPORTED THAT THE SITE HAD THREE STRAIGHT SUCTIONS THAT WERE DAMAGED. THERE WAS LESS THEN AN HOUR DELAY TO THE PROCEDURE. THERE WAS NO IMPACT ON THE PATIENT¿S OUTCOME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8698251 ↗MalfunctionMar 25, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION THAT, WHILE OUTSIDE OF A PROCEDURE, THE CURVED SUCTION WAS DAMAGED AS THE INSTRUMENT WOULD RETURN A HIGH DISTANCE TO DIVOT VALUE, >2.0MM, WHEN VERIFYING. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 8448465 ↗MalfunctionDec 6, 2018Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED INTRA/PERI-OPERATIVELY OF A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS) PROCEDURE. IT WAS REPORTED THAT THE SITE'S CURVED SUCTION INSTRUMENT WAS FOUND TO BE DAMAGED DURING THE CASE. THE SITE USED ANOTHER INSTRUMENT TO COMPLETE THE SURGERY. THERE WAS A LESS THAN 1-HOUR DELAY TO THE PROCEDURE AND NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8137250 ↗MalfunctionJan 4, 2018Communication or Transmission Problem; Device Operates Differently Than Expected; Material Integrity Problem; Communication or Transmission Problem; Device Operates Differently Than Expected; Material Integrity Problem
A SITE REPRESENTATIVE REPORTED THAT DURING FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), A 70 DEGREE CURVED SUCTION COULD NOT BE RECOGNIZED. A BACK UP SUCTION WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY TO PROCEDURE. NO IMPACT ON PATIENT OUTCOME. AFTER PROCEDURE, A SITE REPRESENTATIVE WAS ABLE TO REPLICATE THE REPORTED ISSUE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7165460 ↗MalfunctionOct 12, 2017Material Deformation; Material Deformation
A MEDTRONIC REPRESENTATIVE REPORTED THAT DURING A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), A 70 DEGREE SUCTION WAS DIFFICULT TO VERIFY. SURGEON WAS ABLE TO GET THE INSTRUMENT TO VERIFY BY MANIPULATING THE INSTRUMENT. THE PROCEDURE WAS COMPLETED WITH THE USE OF NAVIGATION. THERE WAS NO DELAY TO PROCEDURE. NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6945456 ↗MalfunctionJun 23, 2017Communication or Transmission Problem; Device Operates Differently Than Expected; Communication or Transmission Problem; Device Operates Differently Than Expected
A MEDTRONIC REPRESENTATIVE REPORTED THAT, WHILE IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), THE SITE WAS UNABLE TO TRACK THE 70 DEGREE SUCTION. IT WAS REPORTED THAT THE INSTRUMENT COULD BE SEEN IN THE FIELD OF VIEW, BUT NAVIGATION WAS NOT POSSIBLE. WHEN ATTEMPTING TO RE-VERIFY, THE DIVOT ERROR WAS TOO HIGH AND THE INSTRUMENT WOULD NOT COMPLETE THE VERIFICATION. ALL OTHER INSTRUMENTS COULD BE VERIFIED AND THE EMITTER WAS PROPERLY PLACED. THE SITE ELECTED TO USE A SEPARATE SUCTION TO COMPLETE THE PROCEDURE. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6663086 ↗MalfunctionJun 16, 2017Material Deformation; Material Deformation
A MEDTRONIC REPRESENTATIVE REPORTED THAT, WHILE OUTSIDE OF A PROCEDURE, THE INSTRUMENT TRACKER WOULD NO LONGER REMAIN ON A 70 DEGREE SUCTION. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Patient Involvement
MAUDE report 6646960 ↗MalfunctionMay 25, 2017Material Deformation; Material Deformation
A MEDTRONIC REPRESENTATIVE REPORTED THAT, WHILE IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), THE CURVED SUCTION BECAME BENT. THE SITE ELECTED TO CONTINUE USING THE SUCTION IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED. THERE WAS NO REPORTED DELAY TO THE PROCEDURE DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6591177 ↗MalfunctionApr 5, 2017Break; Material Deformation; Break; Material Deformation
A MEDTRONIC REPRESENTATIVE REPORTED THAT WHILE IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), THE (B)(6) DEGREE CURVED SUCTION INSTRUMENT WAS DAMAGED. THE SURGEON CONTINUED THE SURGERY USING ANOTHER INSTRUMENT. THERE WAS A REPORTED DELAY TO THE PROCEDURE OF LESS THAN 1 HOUR DUE TO THIS ISSUE. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6461787 ↗MalfunctionMar 30, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
A SITE REPRESENTATIVE REPORTED THAT THE CURVED SUCTION WAS BENT BUT IT WAS STILL POSSIBLE TO VERIFY IT WHEN PUSHING/PRESSING IT INTO THE DIVOT. NO PATIENT WAS PRESENT DURING THE TIME OF THE REPORTED INCIDENT.
Patient problems: No Patient Involvement
MAUDE report 6446379 ↗MalfunctionDec 15, 2016Imprecision; Imprecision
A MEDTRONIC REPRESENTATIVE REPORTED THAT, WHILE IN A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS), THE SURGEON ALLEGED AN INACCURACY OCCURRED. THE SURGEON ALLEGED A 4 MILLIMETER INACCURACY WITH THE CURVED 70 DEGREE SUCTION. ALL OTHER INSTRUMENTS WERE ACCURATE WITHIN 0.5 MILLIMETERS. SUCTION VERIFIED WITH NO ISSUE, HOWEVER, THE SURGEON DEEMED BEING INACCURATE USING THEM. THE DOCTOR STOPPED USING THESE INSTRUMENTS AND CONTINUED THE PROCEDURE. THE SURGEON COMPLETED THE PROCEDURE WITH THE USE OF THE NAVIGATION SYSTEM. DELAY IN THERAPY WAS LESS THAN ONE HOUR. THERE WAS NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6175743 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K001284. One submission often covers many models, so check the product and lot details in each record.
- Z-2317-2016 · Jul 18, 2016 · TerminatedFusion ENT Application. The software application is sent in CD format with an IFU, wrapped in plastic with a label for shipping purposes. Product Usage: The StealthStation Syste
- Z-2319-2016 · Jul 18, 2016 · TerminatedMACH AxiEM Cranial Treon. The software application is sent in CD format with an IFU, wrapped in plastic with a label for shipping purposes. Product Usage: The StealthStation Sys
- Z-1592-2014 · Apr 24, 2014 · TerminatedPIN, 9733235, 100MM, STERILE, PERC REF
- Z-1591-2014 · Apr 24, 2014 · TerminatedPATIENT TRACKER 9733534XOM ENT 1PK
- Z-1586-2014 · Apr 24, 2014 · TerminatedINSTRUMENT TRACKER 9733533XOM ENT 1PK
- Z-1589-2014 · Apr 24, 2014 · TerminatedPACK 9731975 TUMOR RESECTION AXIEM CR
- Z-1602-2014 · Apr 24, 2014 · TerminatedUPGRD PRGM 9734638 ADD AxiEM NAVIGATION
- Z-1587-2014 · Apr 24, 2014 · TerminatedKIT, 9731427, THORACIC TACTILE PROBES
- Z-1595-2014 · Apr 24, 2014 · TerminatedSPHERES 8801071 1/TRAY 12PK
- Z-1579-2014 · Apr 24, 2014 · TerminatedBIOPSY NEEDLE KIT, 9733068, PASSIVE
- Z-1590-2014 · Apr 24, 2014 · TerminatedPAK 9733497 NAV NIM-SPINE PEDICLE ACC KT
- Z-1601-2014 · Apr 24, 2014 · TerminatedTUMOR RESECTION KIT 9733608 SKULL MOUNT
- Z-1582-2014 · Apr 24, 2014 · TerminatedDISP 9730951 STRL SPHERE, 5PK INSERT
- Z-1600-2014 · Apr 24, 2014 · TerminatedTUMOR RESECTION KIT 9733607 NON-INVASIVE
- Z-1594-2014 · Apr 24, 2014 · TerminatedSHUNT KIT 9733605 NON-INVASIVE
…and 13 more.
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page