Skip to content

N/A 9733534XOMRecalls & reported problems

Medtronic Navigation, Inc. · PATIENT TRACKER 9733534XOM ENT 1PK · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994380029), received between Jun 9, 2020 and Jun 9, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

28 other recalls cover devices cleared under the same FDA submission (K001284); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Communication or Transmission Problem1
  • Human-Device Interface Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 9, 2020Communication or Transmission Problem; Human-Device Interface Problem; Communication or Transmission Problem; Human-Device Interface Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED FOR A FUNCTIONAL ENDOSCOPIC SINUS SURGERY (FESS). IT WAS REPORTED THAT THE SYSTEM WOULD NOT RECOGNIZE MULTIPLE PATIENT TRACKERS. HE WAS IN A FACETIME CALL WITH THE FACILITY AT THE TIME. HE SAW THAT THE EMITTER DETAILS SHOWED IT WAS CONNECTED BUT WAS NOT GIVING ANY VALUE. THERE WAS A REPORTED DELAY OF LESS THAN ONE HOUR AND NO REPORTED IMPACT TO PATIENT OUTCOME. ADDITIONAL INFORMATION RECEIVED. THE MANUFACTURER REPRESENTATIVE WAS ON SITE AND WAS ABLE TO GET A NIPT TRACKING WITH NO ISSUE. THE HEAD FRAME PATIENT TRACKER WAS ONLY WORKING INTERMITTENTLY. EM INTERFACE SHOWED THAT COIL 1 OF THE PATIENT TRACKER WAS INTERMITTENTLY NOT GETTING PICKED UP, BUT COIL 2 AND 3 WERE WORKING. SITE REUSES THE PATIENT TRACKERS. SITE WAS INFORMED THAT THE INSTRUMENT IS NOT INTENDED FOR MORE THAN ONE USE AND ISSUE WOULD LIKELY NOT OCCUR AGAIN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10135833 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K001284. One submission often covers many models, so check the product and lot details in each record.

  • Z-2319-2016 · Jul 18, 2016 · Terminated
    MACH AxiEM Cranial Treon. The software application is sent in CD format with an IFU, wrapped in plastic with a label for shipping purposes. Product Usage: The StealthStation Sys
  • Z-2317-2016 · Jul 18, 2016 · Terminated
    Fusion ENT Application. The software application is sent in CD format with an IFU, wrapped in plastic with a label for shipping purposes. Product Usage: The StealthStation Syste
  • Z-1591-2014 · Apr 24, 2014 · Terminated
    PATIENT TRACKER 9733534XOM ENT 1PK
  • Z-1583-2014 · Apr 24, 2014 · Terminated
    DRAPE 9732722 TUBE STERILE O-ARM 20PK
  • Z-1598-2014 · Apr 24, 2014 · Terminated
    TRAJ GUIDE KIT, 9733066, BIOPSY, INT
  • Z-1589-2014 · Apr 24, 2014 · Terminated
    PACK 9731975 TUMOR RESECTION AXIEM CR
  • Z-1587-2014 · Apr 24, 2014 · Terminated
    KIT, 9731427, THORACIC TACTILE PROBES
  • Z-1602-2014 · Apr 24, 2014 · Terminated
    UPGRD PRGM 9734638 ADD AxiEM NAVIGATION
  • Z-1600-2014 · Apr 24, 2014 · Terminated
    TUMOR RESECTION KIT 9733607 NON-INVASIVE
  • Z-1592-2014 · Apr 24, 2014 · Terminated
    PIN, 9733235, 100MM, STERILE, PERC REF
  • Z-1578-2014 · Apr 24, 2014 · Terminated
    BAG 9732315 MOUSE STERILE 15 PK
  • Z-1596-2014 · Apr 24, 2014 · Terminated
    SPHERES 8801075 5/TRAY 12PK
  • Z-1581-2014 · Apr 24, 2014 · Terminated
    DISP 9730950 STRL SPHERE, 1 PK INSERT
  • Z-1590-2014 · Apr 24, 2014 · Terminated
    PAK 9733497 NAV NIM-SPINE PEDICLE ACC KT
  • Z-1579-2014 · Apr 24, 2014 · Terminated
    BIOPSY NEEDLE KIT, 9733068, PASSIVE

…and 13 more.

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page