Skip to content

CAPSTONE® Spinal System 2990722Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SPACER 2990722 CAPSTONE PEEK 07X22 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994387295), received between Jun 21, 2018 and Jun 21, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Malposition of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Malposition of Device1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2018Malposition of Device; Malposition of Device

    IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR FUSION AT TH9-S2AI DUE TO SCOLIOSIS. INTRA-OP, WHILE PERFORMING POSTERIOR LUMBAR INTERBODY FUSION BETWEEN L5/S, THE CAGE WAS DEEPLY INSERTED IN THE INTERVERTEBRAL DISC SPACE AND IT WAS IN THE STATE OF BACKING OUT IN ANTERIOR OF THE VERTEBRAL BODY, AS CONFIRMED LATER WITH THE IMAGE. THE DOCTOR JUDGED THAT THERE WAS NO INJURY TO THE BLOOD VESSEL ETC., AND THE CAGE WAS INDWELT IN THE PATIENT¿S BODY WITHOUT BEING REMOVED; A CAGE OF A BIGGER SIZE WAS INSTALLED INSTEAD. THE ENTRANCE OF THE INTERVERTEBRAL SPACE WAS NARROW AND THE INNER PART BECAME WIDENED SLIGHTLY. PATIENT COMPLICATIONS WERE REPORTED AS UNKNOWN.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7624858 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page