CAPSTONE® Spinal System 2990926Recalls & reported problems

Medtronic Sofamor Danek, Inc. · SPACER 2990926 CAPSTONE PEEK 09X26 · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994387370), received between Nov 27, 2018 and Jan 26, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Crack1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2025Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT USED IN L5-S1 POSTERIOR DECOMPRESSION AND INSTRUMENTED FUSION WITH TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) SPINAL THERAPY FOR SPINAL STENOSIS. IT WAS REPORTED THAT WHEN SURGEON WANTED TO BRING BACK THE CAGE BY 2MM AND WAS TAPPING THE CAGE OUT WITH THE INSERTER HANDLE AND A MALLET. THE INSERTER SEEMED TO LET GO AND WHEN IT WAS PULLED OUT OF THE PATIENT, THE PIECE THAT THREADS ON WAS STILL ATTACHED TO THE CAGE AND THE END OF THE CAGE REMAINED. X-RAY WAS TAKEN AND THE CAGE APPEARED TO BE IN THE RIGHT SPACE, SO NO INTERVENTIONS WERE CARRIED TO RETRIEVE THAT PIECE OUT OF THE PATIENT THERE ARE NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THERE WERE NO ISSUES WITH THE INSERTER. IT WAS ENGAGED WITH THE CAGE PROPERLY AND NO CONCERNS WERE VOICED BY THE SURGEON. NO REVISION SURGERY TO BE PERFORMED AT THIS TIME AND THE REMAINING PART OF THE CAGE WAS DISCARDED BY STAFF.

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 21236959 ↗
  • MalfunctionNov 25, 2020Crack; Crack

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER VIA MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH AN INDICATION OF LUMBAR DEGENERATION IN NEED OF POSTERIOR LUMBAR DECOMPRESSION AND FUSION USED IN SPINAL THERAPY. IT WAS REPORTED THAT DURING THE IMPLANTATION PROCESS, THE PROSTHESIS CRACKED, SO IT WAS REMOVED AFTER IMPLANTATION AND IMMEDIATELY REPLACED WITH A NEW PROSTHESIS. THERE WAS A DELAY OF LESS THAN 60 MINS IN OVERALL PROCEDURE TIME REPORTED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FRAGMENTS IN THE PATIENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10901947 ↗
  • MalfunctionNov 27, 2018Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT LUMBAR INTERBODY FUSION AT L3/4/5 DUE TO DEGENERATIVE SCOLIOSIS. INTRA-OP, THE CAGE WAS INSERTED FROM THE LEFT OF L4/5, WHILE BEING HAMMERED, PHYSICIAN CONFIRMED FROM THE PERSPECTIVE IMAGE THAT THE MARKER LOOKED DIFFERENT FROM USUAL. WHEN THE PHYSICIAN ATTEMPTED TO ADJUST THE MARKER TO BE REFLECTED ON THE TRUE SAGITTAL, CAGE BROKE ALONG WITH A SNAP SOUND. A NEW CAGE OF ONE SIZE SMALLER WAS OPENED TO DEAL WITH ISSUE. NO FRAGMENT REMAINS IN THE PATIENT. NO PATIENT SYMPTOMS OR COMPLICATION WERE REPORTED AS RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8104996 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →