NA 1898430Recalls & reported problems

Medtronic Xomed, Inc. · FOOT SWITCH 1898430 ROHS · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00613994436405), received between Apr 28, 2017 and Feb 19, 2024 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Self-Activation or Keying” (6 reports).

1 other recall cover devices cleared under the same FDA submission (K081277); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Self-Activation or Keying6
  • Device Remains Activated1
  • Material Split, Cut or Torn1

Patient problems

  • No Consequences Or Impact To Patient4
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 19, 2024Self-Activation or Keying; Material Split, Cut or Torn; Self-Activation or Keying; Material Split, Cut or Torn

    IT WAS REPORTED THAT PRE-OP, THE FOOT PEDAL GRIP TAPE WAS LIFTING FROM THE PEDAL THE PEDAL ACTIVATES WHEN NOT BEING PRESSED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18730576 ↗
  • MalfunctionJan 8, 2024Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT HANDPIECE WOULD NOT STOP. THERE IS NO KNOWN PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18464236 ↗
  • MalfunctionJun 30, 2023Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING SURGERY AND DURING BIOMED TESTING, THE FOOT PEDAL WOULD ACTIVATE THE HAND PIECE ON ITS OWN AND DISPLAYED  CODE 17 ERROR ON THE SCREEN. DURING BIOMED TESTING, AFTER TURNING ON THE DEVICE, THE HAND-PIECE WAS LEFT ON THE WORK BENCH AT REST AND THE HAND-PIECE ACTIVATED ON ITS OWN (WITHOUT STEPPING ON THE FOOT SWITCH) AFTER A COUPLE MINUTES OF INACTIVITY. THE ISSUE WAS NOT DISCOVERED PRIOR TO USE ON THE PATIENT AND THE THE USER DID NOT ACTIVATE THE DEVICE IN ERROR. NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17236194 ↗
  • MalfunctionJan 12, 2018Self-Activation or Keying; Self-Activation or Keying

    THE CUSTOMER REPORTED THE HANDPIECE STALLED, THEN OPERATED AT LOW RPM WHEN THE FOOT SWITCH WAS NOT PRESSED. THE CONSOLE (EC300) WAS USED WITH THE FOOT SWITCH WHEN THE EVENT OCCURRED. THERE WAS NO PATIENT IMPACT. REPORT 1 OF 2 - FOOT SWITCH.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7187552 ↗
  • MalfunctionNov 14, 2017Self-Activation or Keying; Self-Activation or Keying

    THE FACILITY REPORTED THAT THE FOOT PEDAL IS ACTIVATING HANDPIECES WHEN IT IS NOT BEING PRESSED ON. THE FAULT WAS CONFIRMED WITH MULTIPLE UNITS. A BACKUP FOOT PEDAL WAS USED FOR THIS PROCEDURE. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7028634 ↗
  • MalfunctionSep 15, 2017Self-Activation or Keying; Self-Activation or Keying

    THE CUSTOMER REPORTED THAT DURING SETUP, WHEN THE HANDPIECE WAS AT REST/INACTIVE AND THE FOOT SWITCH WAS RELEASED, THE HANDPIECE TURNED ON BY ITSELF. ALSO, WHEN USING THE FOOT SWITCH, THERE WAS NO ACTION FROM THE CONSOLE. THERE WAS NO PATIENT IMPACT OR INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6868407 ↗
  • MalfunctionApr 28, 2017Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT AN ATTACHMENT RUNS WHEN THE FOOT SWITCH IS NOT PRESSED. THERE WAS NO PATIENT IMPACT OR INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6530758 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K081277. One submission often covers many models, so check the product and lot details in each record.

  • Z-0961-2022 · Dec 15, 2021 · Terminated
    Software 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND E

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →