NA 1898430Recalls & reported problems
Medtronic Xomed, Inc. · FOOT SWITCH 1898430 ROHS · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00613994436405), received between Apr 28, 2017 and Feb 19, 2024 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Self-Activation or Keying” (6 reports).
1 other recall cover devices cleared under the same FDA submission (K081277); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Self-Activation or Keying6
- Device Remains Activated1
- Material Split, Cut or Torn1
Patient problems
- No Consequences Or Impact To Patient4
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 19, 2024Self-Activation or Keying; Material Split, Cut or Torn; Self-Activation or Keying; Material Split, Cut or Torn
IT WAS REPORTED THAT PRE-OP, THE FOOT PEDAL GRIP TAPE WAS LIFTING FROM THE PEDAL THE PEDAL ACTIVATES WHEN NOT BEING PRESSED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18730576 ↗MalfunctionJan 8, 2024Device Remains Activated; Device Remains Activated
IT WAS REPORTED THAT HANDPIECE WOULD NOT STOP. THERE IS NO KNOWN PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18464236 ↗MalfunctionJun 30, 2023Self-Activation or Keying; Self-Activation or Keying
IT WAS REPORTED THAT DURING SURGERY AND DURING BIOMED TESTING, THE FOOT PEDAL WOULD ACTIVATE THE HAND PIECE ON ITS OWN AND DISPLAYED CODE 17 ERROR ON THE SCREEN. DURING BIOMED TESTING, AFTER TURNING ON THE DEVICE, THE HAND-PIECE WAS LEFT ON THE WORK BENCH AT REST AND THE HAND-PIECE ACTIVATED ON ITS OWN (WITHOUT STEPPING ON THE FOOT SWITCH) AFTER A COUPLE MINUTES OF INACTIVITY. THE ISSUE WAS NOT DISCOVERED PRIOR TO USE ON THE PATIENT AND THE THE USER DID NOT ACTIVATE THE DEVICE IN ERROR. NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17236194 ↗MalfunctionJan 12, 2018Self-Activation or Keying; Self-Activation or Keying
THE CUSTOMER REPORTED THE HANDPIECE STALLED, THEN OPERATED AT LOW RPM WHEN THE FOOT SWITCH WAS NOT PRESSED. THE CONSOLE (EC300) WAS USED WITH THE FOOT SWITCH WHEN THE EVENT OCCURRED. THERE WAS NO PATIENT IMPACT. REPORT 1 OF 2 - FOOT SWITCH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7187552 ↗MalfunctionNov 14, 2017Self-Activation or Keying; Self-Activation or Keying
THE FACILITY REPORTED THAT THE FOOT PEDAL IS ACTIVATING HANDPIECES WHEN IT IS NOT BEING PRESSED ON. THE FAULT WAS CONFIRMED WITH MULTIPLE UNITS. A BACKUP FOOT PEDAL WAS USED FOR THIS PROCEDURE. THERE WAS NO PATIENT IMPACT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7028634 ↗MalfunctionSep 15, 2017Self-Activation or Keying; Self-Activation or Keying
THE CUSTOMER REPORTED THAT DURING SETUP, WHEN THE HANDPIECE WAS AT REST/INACTIVE AND THE FOOT SWITCH WAS RELEASED, THE HANDPIECE TURNED ON BY ITSELF. ALSO, WHEN USING THE FOOT SWITCH, THERE WAS NO ACTION FROM THE CONSOLE. THERE WAS NO PATIENT IMPACT OR INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6868407 ↗MalfunctionApr 28, 2017Self-Activation or Keying; Self-Activation or Keying
IT WAS REPORTED THAT AN ATTACHMENT RUNS WHEN THE FOOT SWITCH IS NOT PRESSED. THERE WAS NO PATIENT IMPACT OR INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6530758 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081277. One submission often covers many models, so check the product and lot details in each record.
- Z-0961-2022 · Dec 15, 2021 · TerminatedSoftware 1898072 IPC upgrade v 2.7.3.0, IPC System: a) CONSOLE 1898001 IPC, Product Number 1898001; b) CONSOLE IPC REFURB, Product Number EC300RF; c) MASTER CONSOLE EC300 LEGEND E
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026