Skip to content

StealthStation® S7® 9733886Recalls & reported problems

Medtronic Navigation, Inc. · FRAME 9733886 SMALL ACTIVE HC · Class 2

All models of Stealthstation S7 (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994480668), received between Nov 24, 2024 and Nov 24, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Device Sensing Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Sensing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2024Device Sensing Problem; Device Sensing Problem

    MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT WHILE CALIBRATING THE NAVIGATION SYSTEM AND IMAGING SYSTEM, THE ACTIVE FRAME STOPPED TRACKING ON THE NAVIGATION SYSTEM. JUST RECEIVED THE ACTIVE FRAME ON FRIDAY, (B)(6) 2023. PLUGGED INTO DIFFERENT PORTS AND WAS STILL UNABLE TO SEE IT. CHECKED INSTRUMENT TAB AS WELL AND IT WAS NOT TRACKING. THE CONCAVE POINTER WAS TRACKING THOUGH. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED. THE MANUFACTURER REPRESENTATIVE REMOVED THE SIDE PANEL TO MAKE SURE THE BOX FOR THE CONNECTION LOOKED OKAY. THE MORE THE MANUFACTURER REPRESENTATIVE TROUBLESHOOTED, THEY NOTICED THAT THE ACTIVE FRAME WAS GOING IN AND OUT OF CONNECTION. WHEN THE CORD WAS MOVED INTO A CERTAIN WAY, IT WOULD WORK. THE MANUFACTURER REPRESENTATIVE TAPED THE CORD DOWN TO KEEP A SOLID CONNECTION WHICH WORKED. THE MANUFACTURER REPRESENTATIVE ALSO USED IT FOR ANOTHER CALIBRATION (B)(6)2023 AND IT SEEMED TO WORK WITHOUT ANY LOSS OF CONNECTION. THE MOST LIKELY / SUSPECTED CAUSE OF THE EVENT WAS THE CORD MAY BE WEARING DOWN SINCE IT WAS GOING IN AND OUT OF CONNECTION, AND THE MANUFACTURER REPRESENTAT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20768878 ↗

Other Neurological stereotaxic Instrument devices

All Neurological stereotaxic Instrument devices →