Clydesdale™ Spinal System 2968045Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CAGE 2968045 CLYDESDALE PEEK IB 6 10X45 · Class 2 · Implant
All models of Clydesdale Spinal System (25)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994485861), received between Feb 13, 2017 and Feb 5, 2021 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (K100175, P000058); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration2
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Crack1
- Migration or Expulsion of Device1
Patient problems
- Neurological Deficit/Dysfunction1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryFeb 5, 2021Crack; Migration; Crack; Migration
INFORMATION WAS RECEIVED FROM THE HEALTHCARE PROFESSIONAL (HCP) VIA THE MANUFACTURER REPRESENTATIVE REGARDING AN EVENT WHICH OCCURRED DURING AN OBLIQUE APPROACH INTERBODY FUSION PROCEDURE IN A PATIENT DIAGNOSED WITH LUMBAR DISC HERNIATION. IT WAS REPORTED THAT THE FUSION CAGE WAS CRACKED, WHICH CAUSED THE PROTRUDING AND DISPLACEMENT OF THE FUSION CAGE AFTER THE SURGERY, AND REVISION SURGERY WAS PERFORMED. PRODUCT WILL BE RETURNED. NO PATIENT OTHER PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11284002 ↗InjuryOct 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MDR DECISION CORRECTED TO NOT REPORTABLE. NO ADDITIONAL SUPPLEMENTALS REQUIRED UNLESS ADDITIONAL INFORMATION RECEIVED INDICATES REPORTABLE EVENT.
Patient problems: Neurological Deficit/Dysfunction
MAUDE report 10733628 ↗InjuryJan 2, 2019Migration; Migration
PRE-OPERATIVE DIAGNOSIS: ADULT SPINAL DEFORMITY PROCEDURE: OBLIQUE LUMBAR INTERBODY FUSION (OLIF) IT WAS REPORTED THAT POST-OP, TRANSFERENCE TO THE LATERAL SIDE OCCURRED TO THE CAGE, THE NEXT DAY. THE LATERAL TRANSFERENCE WAS ABOUT 5 MM NOW. THE PHYSICIAN OBSERVED THAT THE EVENT OCCURRED DUE TO INTERVERTEBRAL DISC SPACE WAS NOT CLEANED UP WELL AND CAGE OF A WRONG SIZE WAS SELECTED. AT THAT TIME, IT WAS DECIDED TO PERFORM POSTERIOR FIXATION TWO WEEKS LATER. ON (B)(6) 2018, IT WAS PLANNED TO HAMMER THE CAGE IN ADDITIONALLY AND TO PERFORM POSTERIOR FIXATION ON (B)(6) 2018. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. PATIENT HAD PAIN IN THE LOWER LIMB.
Patient problems: Pain
MAUDE report 8210767 ↗MalfunctionJun 18, 2018Break; Break
THE PATIENT WAS PRE-OPERATIVELY DIAGNOSED WITH LUMBAR SPONDYLOLISTHESIS AND UNDERWENT OBLIQUE LUMBAR INTERBODY FUSION(OLIF) SURGERY. IT WAS REPORTED THAT DURING SURGERY, THE CAGE BROKE DURING IMPLANTATION. THE PRODUCT CAME IN CONTACT WITH THE PATIENT.NO FRAGMENT OF THE BROKEN CAGE WERE REMAINING IN THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7608923 ↗InjuryFeb 13, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT WITH LUMBAR CANAL STENOSIS UNDERWENT ANTERIOR AND POSTERIOR INTERBODY FUSION SURGERY, IN WHICH, CAGE, SCREWS AND RODS WERE IMPLANTED AT L4-5. POST-OP, THE PATIENT UNDERWENT REVISION SURGERY DUE TO MIGRATION OF THE CAGE INSIDE THE PATIENT. IN THIS SURGERY, THE CAGE, SCREWS AND RODS WERE EXPLANTED. THE MIGRATED CAGE WAS REPLACED WITH A NEW CAGE IN THIS SURGERY. IT WAS ALSO REPORTED THAT THERE WAS A POSSIBILITY OF INFECTION IN THE PATIENT DUE TO THIS EVENT.
Patient problems: No Code Available
MAUDE report 6325063 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K100175, P000058. One submission often covers many models, so check the product and lot details in each record.
- Z-0440-2019 · Sep 27, 2018 · TerminatedINFUSE Bone Graft X SMALL KIT REF 7510100
- Z-1595-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft, LARGE II KIT, REF 7510800AUS, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Z-1592-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft,LARGE II KIT, REF 7510800, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Z-1590-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft, MEDIUM KIT, REF 7510400, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Z-1593-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft, SMALL KIT, REF 7510200AUS, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Z-1594-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft, MEDIUM KIT, REF 7510400AUS, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Z-1589-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft, SMALL KIT, REF 7510200, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
- Z-1591-2013 · May 28, 2013 · TerminatedINFUSE(R) Bone Graft, LARGE KIT, REF 7510600, Medtronic Sofamore Danek USA, Inc., 1800 Pyramid Place, Memphis, Tennessee 38132 orthopaedic
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page