NIM-Neuro® 8253402 (UDI 00613994529077)Recalls & reported problems

Medtronic Xomed, Inc. · MAINFRAME 8253402 NIM NEURO 3.0 INTL · Class 2

All models of NIM-Neuro (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic NIM-Neuro 3.0 mainframe is the nerve-monitoring stimulation unit surgeons use to check nerve function during operations. The 54 reports mostly describe the unit not registering or picking up nerve responses, giving wrong or missing readings, or breaking, and nearly all were found in pre-op checks with no patient involvement; 2 injuries and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
53
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 53 adverse event reports that list this exact device (DI 00613994529077), received between Jun 30, 2017 and Mar 11, 2026 — classified as 2 injuries, 51 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (18 reports).

3 other recalls cover devices cleared under the same FDA submission (K083124); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 2 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings18
  • Use of Device Problem12
  • Break5
  • Device Operates Differently Than Expected5
  • Device Sensing Problem3
  • Failure to Sense3
  • False Negative Result3
  • Application Program Freezes or Fails to Launch2

Patient problems

  • No Clinical Signs, Symptoms or Conditions35
  • No Consequences Or Impact To Patient14
  • No Known Impact Or Consequence To Patient3
  • Complaint, Ill-Defined1
  • Facial Nerve Paralysis1
  • Nerve Damage1
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 53 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT, THE NIM MAINFRAME WAS LOOSING SIGNAL. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24566208 ↗
  • MalfunctionJan 12, 2026Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED DURING PRE-OP THAT THE DEVICE HAD NOT WORKING. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24033207 ↗
  • MalfunctionMar 10, 2025No Device Output; Use of Device Problem; No Device Output; Use of Device Problem

    IT WAS REPORTED THAT PRE-OP THE MAINFRAME WAS NOT GETTING RESPONSE AND NOT WORKING. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21557513 ↗
  • MalfunctionMar 10, 2025False Negative Result; False Negative Result

    IT WAS REPORTED THAT BEFORE OPERATION, IT WAS FOUND THAT THERE WAS NO RESPONSE IN NIM MAINFRAME. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21556995 ↗
  • MalfunctionMar 6, 2025Incorrect, Inadequate or Imprecise Result or Readings; Use of Device Problem; Incorrect, Inadequate or Imprecise Result or Readings; Use of Device Problem

    IT WAS REPORTED THAT DURING PRE CHECK, THE MAINFRAME WAS NOT WORKING PROPERLY AND HAD SENSING ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21536867 ↗
  • MalfunctionFeb 26, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THERE WAS NO STIMULATION RESPONSE FOR MAINFRAME. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21467428 ↗
  • MalfunctionFeb 24, 2025Incorrect, Inadequate or Imprecise Result or Readings; Connection Problem; Incorrect, Inadequate or Imprecise Result or Readings; Connection Problem

    IT WAS REPORTED THAT DURING PREOP THERE WAS NO STIMULATION RESPONSE AND STIM 1 CANNOT INSERT ELECTRODE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21443287 ↗
  • MalfunctionFeb 12, 2025Inappropriate Audible Prompt/Feedback; Inappropriate Audible Prompt/Feedback

    IT WAS REPORTED THAT THE MAINFRAME/PATIENT INTERFACE WAS MAKING NOISES AFTER USAGE. THERE WAS NO PATIENT INVOLVEMENT ASSOCIATED WITH THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21362500 ↗
  • MalfunctionOct 21, 2024Output Problem; Output Problem

    ADDITIONAL INFORMATION RECEIVED STATING THE CAUSE OF THE ISSUE WAS IDENTIFIED AND RESOLVED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20493100 ↗
  • MalfunctionSep 23, 2024Break; Incorrect, Inadequate or Imprecise Result or Readings; Break; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING PRE-OP, THE NIM 3.0 NEUROTRANSMITTER SIGNAL APPEARS WEAK AND NOT TRUTHFUL, BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20277377 ↗
  • MalfunctionSep 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT MAINFRAME WAS NOT GETTING STIMULATION AND ELECTRODE OFF ERROR. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20166595 ↗
  • MalfunctionAug 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE MACHINE IS PLAYING UP. CANNOT IDENTIFY THE PLACEMENT OF NEEDLES, AS SUPPOSED TO BE TICKED AS CHECK, BUT SHOWING "X". WHEN PROMPTED TO CONTINUE, IT SENSES THE NEEDLE PLACEMENTS IN THE NERVES THROUGH SOUNDS. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19997881 ↗
  • MalfunctionJun 17, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING THYROID PROCEDURE, NIM CONSOLE LOST RESPONSE ADJUSTED THE TUBE AND GOT RESPONSE. AGAIN LOST RESPONSE READJUSTED THE TUBE AGAIN BUT STILL NO RESPONSE COULD BE OBTAINED. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19547445 ↗
  • MalfunctionJun 17, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DEVICE NOT WORKING PROPERLY. NO KNOWN IMPACT OR CONSEQUENCE TO PATIENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19546687 ↗
  • MalfunctionApr 30, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    ADDITIONAL INFORMATION RECEIVED THAT CUSTOMER WAS NOT SENDING UNIT IN AS THEY HAVE RECTIFIED ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19210659 ↗
  • MalfunctionFeb 20, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRE-OP, MAINFRAME WAS NOT WORKING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18737020 ↗
  • MalfunctionJan 30, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THE MAINFRAME AND PATIENT INTERFACE DOES NOT WORK CORRECTLY. SOMETIMES IT DOES NOT SENSE OR IT WORKS INCORRECTLY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18606567 ↗
  • MalfunctionJan 8, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRE-OP, THE CONSOLE AND PATIENT INTERFACE WERE NOT WORKING. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18463894 ↗
  • MalfunctionDec 26, 2023Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    ADDITIONAL INFORMATION RECEIVED. TROUBLESHOOTING WAS PERFORMED BY RESTARTING THE DEVICE UNTIL IT WORKS PROPERLY. THE CUSTOMER INDICATED THE DEVICE CAN SOMETIMES NOT BE OPERATED EVEN AFTER RESTARTING. THERE WAS NOTHING DISPLAYED ON THE SCREEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18392346 ↗
  • MalfunctionDec 13, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING INTRA-OP, THE MAINFRAME AND PATIENT INTERFACE WERE NOT WORKING CONSISTENTLY WHEN MONITORING. DEVICES WERE USING THE HEAD NECK, PAROTID 2-CHANNEL SETTING. OCULAR MONITORING WAS NOT STABLE WHEN READING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18315998 ↗
  • MalfunctionNov 28, 2023False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING SET-UP, THE EQUIPMENT IS NOT WORKING AND DOES NOT EMIT ANY SOUND. THE CURRENT STIM DID NOT RETURN SHOW "0", INDICATING THAT THE CURRENT IS NOT BEING DELIVERED. THERE WAS NO VISUAL AND/OR AUDIBLE INDICATOR THAT ALERTED THE USER OF THE ISSUE. THE ISSUE WAS NOT RESOLVED BY REPOSITIONING OR TROUBLESHOOTING. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18213831 ↗
  • MalfunctionNov 8, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRE-OP, THE CURRENT WAS NOT DELIVERED DURING STIMULATION. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18089575 ↗
  • MalfunctionOct 26, 2023Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch

    ADDITIONAL INFORMATION RECEIVED AS THE CUSTOMER RESTARTED THE DEVICE MULTIPLE TIMES UNTIL THE ISSUE WAS RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18011221 ↗
  • MalfunctionOct 24, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRE-OP, THE MAINFRAME WAS NOT WORKING, THE FAULT WAS NOT SPECIFIED BY NURSES. MISSING PATIENT INTERFACE AND MUTING DETECTOR. THERE IS NO PATIENT INVOLVED IN THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18000376 ↗
  • MalfunctionAug 31, 2023Noise, Audible; Noise, Audible

    IT WAS REPORTED THAT NIM MONITOR MAKING INTERFERENCE NOISES EVEN AFTER TROUBLESHOOTING (CHECKING MUTING CORD, ALL PINS IN RIGHT PORT, SWITCHING ON AND OFF AGAIN). USED A DIFFERENT NIM MONITOR ON THE SAME PATIENT, USING THE SAME PINS, DIFFERENT MONITOR MADE NO NOISES. THERE IS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17659674 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083124. One submission often covers many models, so check the product and lot details in each record.

  • Z-2343-2012 · Sep 27, 2010 · Terminated
    Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253410 (Neuro). LABELING: Box Label: REF 8253410 x1***NIM-Response 3.0 Patie
  • Z-2342-2012 · Sep 27, 2010 · Terminated
    Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces 8253200 (Response) LABELING: Box Label: REF 8253200 x1***NIM-Response 3.0 Patient
  • Z-2344-2012 · Sep 27, 2010 · Terminated
    Interfaces subject to recall: Medtronic NIM (Nerve Integrity Monitor) 3.0 Patient Interfaces REF 8253210 (Pulse). LABELING: Box Label: REF 8253210 x1***NIM-Response 3.0 Patie

Other STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices

All STIMULATOR, ELECTRICAL, EVOKED RESPONSE devices →