TALENT® OCCLUDER OCL20USRecalls & reported problems
Medtronic, Inc. · STENT GRAFT OCL20US 20MM OCCLUDER · Class 3 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 1
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994573063), received between Mar 20, 2018 and May 17, 2018 — classified as 1 death, 1 injury.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Death1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 17, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
A TALENT OCCLUDER DEVICE WAS IMPLANTED IN A PATIENT DURING THE ENDOVASCULAR TREATMENT OF A UNKNOWN SIZED ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT DURING THE INDEX PROCEDURE, THE PHYSICIAN IMPLANTED THE TALENT OCCULDER INTO THE LEFT COMMON ILIAC ARTERY. A RETROGRADE INJECTION WAS DONE TO SEE IF THE LEFT COMMON ILIAC ARTERY HAD BEEN OCCLUDED WHICH THEN SHOWED EXTRAVASATION INTO THE RETROPERITONEAL SPACE. THE PHYSICIAN PLACED A 16MM OCCLUDER PLUG TO STOP THE BLEEDING, HOWEVER, IT DID NOT WORK. A FLANK INCISION WAS MADE TO VISUALIZE THE LEFT COMMON ILIAC ARTERY AND THE 20MM TALENT OCCLUDER PLUG WAS REMOVED FROM THE VESSEL AS IT HAD TORN THROUGH THE ARTERY. THE 16MM OCCLUDER PLUG WAS ALSO REMOVED. THROUGH SOME DIFFICULTY DUE THE DAMAGED VESSEL, PROLENE SUTURES WERE SOWN OVER THE ARTERY TO STOP THE BLEEDING AND HOMEOSTASIS WAS ACHIEVED. AT THE END OF THE PROCEDURE A FINAL ANGIOGRAM WAS PERFORMED AND NO EXTRAVASATION WAS APPARENT IN THE RETROPERITONEAL SPACE. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS ANATOMY RELATED DUE TO THE CALCIFICATION IN THE LEFT COMMON ILIAC. IT WAS NOTED BY THE PHYSICIAN THAT THERE IS A POSSIBILITY THAT A DISSECTION WAS CREATED IN THE ILIAC AS THE OC
Patient problems: Vascular Dissection
MAUDE report 7521473 ↗DeathMar 20, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF A RUPTURED UNKNOWN DIAMETER ABDOMINAL AORTIC ANEURYSM AND SUBSEQUENT IMPLANTATION OF ANOTHER ANEURX STENT GRAFT AND ENDURANT STENT GRAFT WAS ALSO PERFORMED. IT WAS REPORTED THAT THE PATIENT PRESENTED EMERGENTLY WITH A RUPTURED ABDOMINAL AORTIC ANEURYSM. THE PATIENT WAS RECEIVING TREATMENT WITH THE ATTEMPTED IMPLANTATION OF AN AUI DEVICE BUT THEN EXPIRED DURING THIS PROCEDURE. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS DUE TO MIGRATION OF THE STENT GRAFT. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED.
Patient problems: Death
MAUDE report 7353710 ↗
Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page