KYPHON® HV-R® Bone Cement C01A (UDI 00613994583215)Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CEMENT C01A KYPHX HV-R BONE CEMENT EN · Class 2 · Implant
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994583215), received between Apr 17, 2017 and Oct 16, 2017 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K041584); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Migration or Expulsion of Device1
Patient problems
- Inadequate Pain Relief3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 16, 2017Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT BALLOON KYPHOPLASTY PROCEDURE ON T12 SPINE LEVEL DUE TO PRIMARY OSTEOPOROSIS AND COMPRESSION FRACTURE. INTRA-OP, DURING CEMENT INJECTION, THE CEMENT LEAKED OUT ALONG THE SIDE WALL OF THE VERTEBRAL BODY AND LEAKED FROM THE DISEASED VERTEBRA (T12) TO THE SIDE WALL OF THE L1. THE SURGEON CONSIDERS THAT THE CEMENT LEAKAGE OCCURRED BECAUSE THE WIDENING OF THE BALLOON MIGHT HAVE BEEN EXCESSIVE. THE SURGEON JUDGED THAT THIS EVENT HAS NO INFLUENCE ON THE PATIENT. THE CEMENT WAS STORED AT A PROPER TEMPERATURE BEFORE AND DURING THE PROCEDURE; AND WAS MIXED FOR LESS THAN 9 MINUTES. THE CEMENT WAS DOUGHY AND HOMOGENEOUS PRIOR TO DELIVERY INTO THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6950752 ↗InjuryAug 1, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PRIMARY DIAGNOSIS: PRIMARY OSTEOPOROSIS LEVELS IMPLANTED: T12 TYPE OF FRACTURE: COMPRESSION FRACTURE DATE OF ADMISSION: (B)(6) 2011 DATE OF DISCHARGE: (B)(6) 2011 IT WAS REPORTED THAT PATIENT FELL DOWN THE STAIRS AND WAS INJURED. SYNOSTOSIS CAN NOT BE OBTAINED WITH INPATIENT REST FOR MORE THAN 1 MONTH AND BALLOON KYPHOPLASTY IS ENFORCED FOR SYMPTOM EXACERBATION. HOSPITALIZATION PERIOD BECAME LONGER THAN USUAL. POST OP REMAINING NRS SCORE WAS REPORTED.
Patient problems: Inadequate Pain Relief
MAUDE report 6757952 ↗InjuryApr 28, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT PRESENTED WITH PRE-OP DIAGNOSIS: PRIMARY OSTEOPOROSIS (TYPE OF FRACTURE-COMPRESSION FRACTURE). FOR WHICH, ON (B)(6) 2011, PATIENT UNDERWENT BALLOON KYPHOPLASTY PROCEDURE AT LEVEL L3. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE ORIGINAL PRODUCT. REPORTEDLY, ON AN UNKNOWN DATE, POST-OP, DURING A POST-OP NRS (NUMERIC RATING SCALE) ASSESSMENT THE PATIENT REPORTED BACK PAIN BUT DID NOT REPORT ANY SCORE. IT IS UNKNOWN WHETHER PATIENT HAD GET ADEQUATE PAIN RELIEF.
Patient problems: Inadequate Pain Relief
MAUDE report 6528732 ↗InjuryApr 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PRE-OPERATIVE DIAGNOSIS: PRIMARY OSTEOPOROSIS (TYPE OF FRACTURE -COMPRESSION FRACTURE) LEVEL: T12 IT WAS REPORTED THAT THE PATIENT WAS ENROLLED IN FOR STUDY. THE PATIENT WAS HOSPITALIZED ON (B)(6) 2011 AND DISCHARGED ON (B)(6) 2011. IT WAS REPORTED THAT ON UNKNOWN DATE, POST-OP, PATIENT COMPLAINED OF BACK PAIN. PATIENT BACK PAIN DID NOT RESOLVE COMPLETELY EVEN 12 MONTHS AFTER THE SURGERY.
Patient problems: Inadequate Pain Relief
MAUDE report 6496708 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041584. One submission often covers many models, so check the product and lot details in each record.
- Z-1625-2011 · Nov 10, 2009 · TerminatedMedtronic brand KyphX HV-R Bone Cement with Kyphon mixer, a polymethyl bone cement kit consisting of a quantity of powder (20 g) and a volume of liquid (10 ml/9 g) mixed in a Kypho
- Z-1624-2011 · Nov 10, 2009 · TerminatedMedtronic brand KyphX HV-R Bone Cement, a polymethyl bone cement consisting of a quantity of powder (20 g) and a volume of liquid (10 ml/9 g) mixed in a Kyphon mixer for up to two
Other CEMENT, BONE, VERTEBROPLASTY devices
- CONFIDENCE SPINAL CEMENT SYSTEM 283910000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY5 deaths163 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283913000Medos International Sàrl · CEMENT, BONE, VERTEBROPLASTY8 deaths146 reports
- KYPHON® Xpede™ Bone Cement and KYPHON® Mixer Pack CX01BMedtronic Sofamor Danek, Inc. · CEMENT, BONE, VERTEBROPLASTY3 deaths67 reports
- SYNFLATE 03.804.701SSynthes GmbH · CEMENT, BONE, VERTEBROPLASTY60 reports
- KYPHON® HV-R® Bone Cement C01AMedtronic Sofamor Danek, Inc. · CEMENT, BONE, VERTEBROPLASTY57 reports
- iVAS 0705315000Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall3 reports
- Cortoss 2101-0005Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall3 reports
- iVAS 0705310000Stryker Corporation · CEMENT, BONE, VERTEBROPLASTY1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.