Skip to content

KYPHON® HV-R® Bone Cement C01A (UDI 00613994583215)Recalls & reported problems

Medtronic Sofamor Danek, Inc. · CEMENT C01A KYPHX HV-R BONE CEMENT EN · Class 2 · Implant

All models of KYPHON HV-R Bone Cement (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994583215), received between Apr 17, 2017 and Oct 16, 2017 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K041584); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Migration or Expulsion of Device1

Patient problems

  • Inadequate Pain Relief3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 16, 2017Migration or Expulsion of Device; Migration or Expulsion of Device

    IT WAS REPORTED THAT THE PATIENT UNDERWENT BALLOON KYPHOPLASTY PROCEDURE ON T12 SPINE LEVEL DUE TO PRIMARY OSTEOPOROSIS AND COMPRESSION FRACTURE. INTRA-OP, DURING CEMENT INJECTION, THE CEMENT LEAKED OUT ALONG THE SIDE WALL OF THE VERTEBRAL BODY AND LEAKED FROM THE DISEASED VERTEBRA (T12) TO THE SIDE WALL OF THE L1. THE SURGEON CONSIDERS THAT THE CEMENT LEAKAGE OCCURRED BECAUSE THE WIDENING OF THE BALLOON MIGHT HAVE BEEN EXCESSIVE. THE SURGEON JUDGED THAT THIS EVENT HAS NO INFLUENCE ON THE PATIENT. THE CEMENT WAS STORED AT A PROPER TEMPERATURE BEFORE AND DURING THE PROCEDURE; AND WAS MIXED FOR LESS THAN 9 MINUTES. THE CEMENT WAS DOUGHY AND HOMOGENEOUS PRIOR TO DELIVERY INTO THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6950752 ↗
  • InjuryAug 1, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PRIMARY DIAGNOSIS: PRIMARY OSTEOPOROSIS LEVELS IMPLANTED: T12 TYPE OF FRACTURE: COMPRESSION FRACTURE DATE OF ADMISSION: (B)(6) 2011 DATE OF DISCHARGE: (B)(6) 2011 IT WAS REPORTED THAT PATIENT FELL DOWN THE STAIRS AND WAS INJURED. SYNOSTOSIS CAN NOT BE OBTAINED WITH INPATIENT REST FOR MORE THAN 1 MONTH AND BALLOON KYPHOPLASTY IS ENFORCED FOR SYMPTOM EXACERBATION. HOSPITALIZATION PERIOD BECAME LONGER THAN USUAL. POST OP REMAINING NRS SCORE WAS REPORTED.

    Patient problems: Inadequate Pain Relief

    MAUDE report 6757952 ↗
  • InjuryApr 28, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT PRESENTED WITH PRE-OP DIAGNOSIS: PRIMARY OSTEOPOROSIS (TYPE OF FRACTURE-COMPRESSION FRACTURE). FOR WHICH, ON (B)(6) 2011, PATIENT UNDERWENT BALLOON KYPHOPLASTY PROCEDURE AT LEVEL L3. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE ORIGINAL PRODUCT. REPORTEDLY, ON AN UNKNOWN DATE, POST-OP, DURING A POST-OP NRS (NUMERIC RATING SCALE) ASSESSMENT THE PATIENT REPORTED BACK PAIN BUT DID NOT REPORT ANY SCORE. IT IS UNKNOWN WHETHER PATIENT HAD GET ADEQUATE PAIN RELIEF.

    Patient problems: Inadequate Pain Relief

    MAUDE report 6528732 ↗
  • InjuryApr 17, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PRE-OPERATIVE DIAGNOSIS: PRIMARY OSTEOPOROSIS (TYPE OF FRACTURE -COMPRESSION FRACTURE) LEVEL: T12 IT WAS REPORTED THAT THE PATIENT WAS ENROLLED IN FOR STUDY. THE PATIENT WAS HOSPITALIZED ON (B)(6) 2011 AND DISCHARGED ON (B)(6) 2011. IT WAS REPORTED THAT ON UNKNOWN DATE, POST-OP, PATIENT COMPLAINED OF BACK PAIN. PATIENT BACK PAIN DID NOT RESOLVE COMPLETELY EVEN 12 MONTHS AFTER THE SURGERY.

    Patient problems: Inadequate Pain Relief

    MAUDE report 6496708 ↗

FDA clearance and classification

Cleared through 510(k) K041584 on Jul 7, 2004 (decision: Substantially Equivalent). Submitted by Kyphon, Inc. as “Kyphx Hv-R Bone Cement, Model C01a”. The FDA review took 23 days. FDA record ↗

The FDA regulates product code NDN (cement, bone, vertebroplasty) under 21 CFR 888.3027 as a Class II device. 21 CFR 888.3027 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K041584. One submission often covers many models, so check the product and lot details in each record.

  • Z-1625-2011 · Nov 10, 2009 · Terminated
    Medtronic brand KyphX HV-R Bone Cement with Kyphon mixer, a polymethyl bone cement kit consisting of a quantity of powder (20 g) and a volume of liquid (10 ml/9 g) mixed in a Kypho
  • Z-1624-2011 · Nov 10, 2009 · Terminated
    Medtronic brand KyphX HV-R Bone Cement, a polymethyl bone cement consisting of a quantity of powder (20 g) and a volume of liquid (10 ml/9 g) mixed in a Kyphon mixer for up to two

Common questions about the KYPHON® HV-R® Bone Cement C01A

Has the KYPHON® HV-R® Bone Cement C01A been recalled?

No FDA recall record in our data names this exact device identifier (DI 00613994583215), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the KYPHON® HV-R® Bone Cement C01A?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Apr 17, 2017 and Oct 16, 2017. In those reports, the most common specific device problem is “Migration or Expulsion of Device” (1); the most common patient problem is “Inadequate Pain Relief” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the KYPHON® HV-R® Bone Cement C01A?

None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the KYPHON® HV-R® Bone Cement C01A FDA approved?

It was cleared through 510(k) K041584 on Jul 7, 2004. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the KYPHON® HV-R® Bone Cement C01A MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the KYPHON® HV-R® Bone Cement C01A, and is it still sold?

The FDA's device database lists Medtronic Sofamor Danek, Inc. as the company responsible for it (brand KYPHON® HV-R® Bone Cement, model C01A). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CEMENT, BONE, VERTEBROPLASTY devices

All CEMENT, BONE, VERTEBROPLASTY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page