Stealthstation® S7™ 9733857RRecalls & reported problems
Medtronic Navigation, Inc. · CART 9733857R S7 SURGEON LCD REFURB · Class 2
All models of Stealthstation S7 (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994603685), received between Feb 10, 2021 and Nov 15, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Application Program Freezes or Fails to Launch” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Application Program Freezes or Fails to Launch2
- Display or Visual Feedback Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 15, 2021Display or Visual Feedback Problem; Display or Visual Feedback Problem
ADDITIONAL INFORMATION RECEIVED FROM A MANUFACTURER REPRESENTATIVE INDICATED THE MONITOR CABLE HAD BEEN REPLACED AND WAS IN THE PROCESS OF BEING RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12814691 ↗MalfunctionApr 12, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED OUTSIDE OF A PROCEDURE THAT THE SYSTEM WOULD NOT BOOT UP AND SHOWS NO SIGNAL MESSAGE. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS THE COMPUTER. THE NEXT STEP WAS TO REPLACE THE COMPUTER. NO PATIENT WAS PRESENT AT THE TIME OF THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DURING THE SITE VISIT THE MANUFACTURER REPRESENTATIVE FOUND OUT THAT THE COMPUTER DOESN'T POWER ON DESPITE THE FACT THAT AC POWER IS COMING IN. MOST LIKELY THE POWER SUPPLY OF THE COMPUTER IS OUT OF ORDER. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE MANUFACTURER REPRESENTATIVE FOUND THE ISSUE WHEN THEY WERE CHECKING UP ON THE SYSTEM THE DAY BEFORE PERFORMING A PROCEDURE. TROUBLESHOOTING WAS PERFORMED AND THE SYSTEM WAS SHOWING NO SIGNAL. THE SYSTEM WAS RESTARTED TWICE WITHOUT SUCCESS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11648486 ↗MalfunctionFeb 10, 2021Application Program Freezes or Fails to Launch; Application Program Freezes or Fails to Launch
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED IN A CRANIAL RESECTION. IT WAS REPORTED THAT THE COMPUTER INTERMITTENTLY FREEZES AND HAS TO BE RESTARTED TO FUNCTION. THE NEXT STEP WAS TO REPLACE THE COMPUTER. THERE WAS LESS THAN AN HOUR DELAY IN THE CASE. NO IMPACT ON PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11307881 ↗
Other Neurological stereotaxic Instrument devices
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument4 deaths4,668 reports
- Stealthstation® S7™ 9733856Medtronic Navigation, Inc. · Neurological stereotaxic Instrument6 deaths3,908 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death1,896 reports
- StealthStation S8 9735669Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1,831 reports
- StealthStation S8 Platform 9735665Medtronic Navigation, Inc. · Neurological stereotaxic Instrument2 deaths1,542 reports
- Fusion™ ENT Image Guidance System 9733560XOMMedtronic Navigation, Inc. · Neurological stereotaxic Instrument1,316 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument8 deaths984 reports
- Stealthstation® S7™ 9733858Medtronic Navigation, Inc. · Neurological stereotaxic Instrument1 death695 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page