ATLANTIS® Anterior Cervical Plate System 7200085Recalls & reported problems
Medtronic Sofamor Danek, Inc. · PLATE 7200085 ATL VISION ELITE 85MM · Class 2 · Implant
All models of ATLANTIS Anterior Cervical Plate System (46)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994624949), received between Nov 23, 2018 and Nov 23, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Dislodged or Dislocated1
- Use of Device Problem1
Patient problems
- Dysphagia/ Odynophagia1
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 23, 2018Use of Device Problem; Device Dislodged or Dislocated; Use of Device Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT ON (B)(6) 2017, THE PATIENT UNDERWENT ANTERIOR CERVICAL DISCECTOMY AND FUSION AT C3-T1 DUE TO DEGENERATIVE DISC DISEASE AND HERNIATED DISC. THE POSTERIOR PORTION OF THE SURGERY WAS PERFORMED ON (B)(6) 2017. ON AN UNKNOWN DATE, POST-OP, THE PATIENT COMPLAINED OF PAIN, DYSPHAGIA AND DIFFICULTY EATING. THE PATIENT WAS NOT FOLLOWED UP BY THE OPERATING PHYSICIAN AND SUBSEQUENTLY ENDED UP IN A REHAB FACILITY FOR OVER A YEAR AND LOST CONSIDERABLE WEIGHT DUE TO INABILITY TO EAT. AS PER THE REPORTED INFORMATION, IN THE INITIAL OPERATION, THE OPERATING SURGEON HAD TRAPPED THE ESOPHAGUS UNDER THE TOP PART OF THE PLATE AND INSERTED ONE SCREW WHICH WENT DIRECTLY THROUGH THE ESOPHAGUS. ONLY ONE SCREW WAS PLACED AT THE TOP OF THE PLATE (AT C3) AND WAS NOT LOCKED DOWN. THIS SCREW BACKED OUT BY 3MM. HENCE, ON (B)(6) 2018, THE REVISION ACDF WAS PERFORMED AT C3-T1, IN WHICH THE PLATE WAS EXPLANTED AND WAS REPLACED WITH PLATE FROM ANOTHER MANUFACTURER. THE SURGEON (PERFORMING THE REVISION) REPORTED THAT THE CAUSE OF DYSPHAGIA WAS THAT THE SCREW WAS INSERTED INTO THE ESOPHAGUS AFTER THE PLATE TRAPPED THE ESOPHAGUS. NO PATIENT COMPLICATIONS WERE REPORTED AFTER THE REVISION SURGERY.
Patient problems: Dysphagia/ Odynophagia; Pain
MAUDE report 8100002 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page