VERTEX® Reconstruction System G7755124Recalls & reported problems
Medtronic Sofamor Danek, Inc. · ROD G7755124 ADJ ROD 3.5MM X 220MM · Class 2 · Implant
All models of VERTEX Reconstruction System (34)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00613994667199), received between Sep 27, 2017 and Nov 23, 2023 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Neck Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 23, 2023Migration; Migration
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PREOPERATIVE DIAGNOSIS FOR POSTERIOR OCCIPITOCERVICAL FUSION. PROCEDURE PERFORMED WAS POSTERIOR OCCIPITOCERVICAL FUSION WAS PERFORMED ON (B)(6) DUE TO C2 METASTASIS. LEVEL AT WHICH IMPLANT WAS PERFORMED O/C3. IT WAS REPORTED THAT THE ROD ALIGNMENT CHANGED POSTOPERATIVELY AND THE O-C2 ANGLE CHANGED, RESULTING IN A REOPERATION. DATE OF INITIAL SURGERY ON (B)(6) 2023 THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18202922 ↗InjuryAug 8, 2023Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH PREOPERATIVE DIAGNOSIS FOR CEREBRAL PALSY. PROCEDURE PERFORMED WAS POSTERIOR CERVICAL SPINE FUSION. IT WAS REPORTED THAT THE AFTER C2 TO C6 WERE ALREADY FIXED WITH JJ'S SYNAPSES (ABOUT 10 YEARS AGO), DEPLOYMENT SURGERY WAS PERFORMED IN (B)(6) 2023 USING A CRANIAL PLATE (VS). AFTERWARD, THERE WAS A POSSIBILITY THAT SOME KIND OF MALFUNCTION OCCURRED WITH THE ADJUSTABLE ROD. HENCE, IT WAS CONFIRMED BY PERFORMING A REOPERATION. THE JOINT OF THE ADJUSTABLE ROD BROKE, AND IT WAS REPLACED WITH A PREVENT ROD (CCM) (ONLY THE RIGHT SIDE). THE INVOLUNTARY MOVEMENT OF CEREBRAL PALSY WAS ACCOMPANIED BY A LOAD APPLIED TO THE ROD, WHICH LED TO THE BREAKAGE OF THE ROD JOINT. PROCEDURE WAS PERFORMED FOR THE REMOVAL OF BROKEN ROD. THERE WAS NO FRAGMENT LEFT INSIDE THE PATIENT. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THIS EVENT. MDT PRODUCT WAS NOT USED IN THE INITIAL SURGERY. THERE WAS NO DELAY IN THE OVERALL PROCEDURE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477905 ↗InjurySep 27, 2017Break; Break
THE PATIENT UNDERWENT POSTERIOR CERVICAL FUSION AT OC-C6. POST-OP, THE PATIENT COMPLAINED OF PAIN AT CERVICAL. PATIENT UNDERWENT X-RAY WHICH REVEALED BREAKAGE OF CONNECTOR PART OF THE ADJUSTABLE ROD AT OC-C2. THIS CAUSED PROTRUSION OF ROD AND PAIN. THE BROKEN ROD WAS SCHEDULED TO BE REMOVED ON (B)(6) 2017. NO INFORMATION REGARDING REVISION SURGERY WAS MADE AVAILABLE.
Patient problems: Neck Pain
MAUDE report 6897711 ↗
Other ORTHOSIS, SPINAL PEDICLE FIXATION devices
- NA 04.614.508Synthes GmbH · ORTHOSIS, SPINAL PEDICLE FIXATION277 reports
- CREO 1134.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION112 reports
- CREO 1119.0010Globus Medical, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION80 reports
- VIPER 186731745Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION43 reports
- CD HORIZON® Spinal System 1553201090Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION38 reports
- EXPEDIUM 179902000Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION37 reports
- CD HORIZON® Spinal System 7540220Medtronic Sofamor Danek, Inc. · ORTHOSIS, SPINAL PEDICLE FIXATION36 reports
- EXPEDIUM VERSE 199721640Medos International Sàrl · ORTHOSIS, SPINAL PEDICLE FIXATION35 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page