T2™ Spinal System 9392221Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CENTERPIECE 9392221 22 X 21MM 22A · Class 2 · Implant
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994676597), received between Jul 1, 2019 and Jul 1, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Migration or Expulsion of Device1
Patient problems
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2019Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT VERTEBRAL COLUMN RESECTION AT T10-L2. BONE QUALITY OF THE PATIENT WAS REPORTED AS POOR. POST-OP, THE SCREW PLACED AT L1-L2 LOOSENED; DUE TO WHICH THE CAGE BECAME UNSTABLE AND INCLINED. THE PATIENT ALSO EXPERIENCED PAIN. HENCE, A REVISION SURGERY WAS PERFORMED AND THE LOOSENED SCREW WAS REMOVED AND REPLACED WITH A BIGGER SCREW. THE FUSION WAS ALSO EXTENDED TO L4 AND THE CAGE POSITION WAS ADJUSTED. THE PAIN DISAPPEARED AFTER THE REVISION SURGERY.
Patient problems: Pain
MAUDE report 8747166 ↗
Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices
- X-CORE 7160014Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7220016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 23-32 mm, 2-5 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7180018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 33-54 mm, 7-10 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- GIZA Corpectomy implant Giza, Diam 14 mm, 18-22 mm, 5-8 DegEden Spine Europe SA · Spinal Vertebral Body Replacement Device1 recall0 reports
- X-CORE 7220018Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
- X-CORE 7180016Nuvasive, Inc. · SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.