Protecta™ DR D334DRG (UDI 00613994678157)Recalls & reported problems
Medtronic, Inc. · ICD D334DRG PROTECTA DR US MR BCP · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 1
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00613994678157), received between Feb 9, 2017 and Apr 8, 2022 — classified as 1 death, 7 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Battery Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Battery Problem3
- Delayed Charge Time3
- Premature Discharge of Battery3
- Failure to Charge2
- Defibrillation/Stimulation Problem1
- Electrical /Electronic Property Problem1
- Electromagnetic Interference1
- Inappropriate or Unexpected Reset1
Patient problems
- No Known Impact Or Consequence To Patient8
- Cardiac Arrest1
- Death1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryApr 8, 2022Failure to Charge; Electrical /Electronic Property Problem; Failure to Charge; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS ALERTING FOR A CHARGE-CIRCUIT TIMEOUT AND CHARGE-CIRCUIT INACTIVE, THOUGH IT WAS NOTED THAT IT WAS PAST END OF SERVICE (EOS) AND ALL ALERTS WERE PROGRAMMED OFF. THE ICD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14055347 ↗InjuryAug 2, 2019Failure to Charge; Battery Problem; Failure to Charge; Battery Problem
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) HAD A CHARGE CIRCUIT TIMEOUT AFTER DECLARING RECOMMENDED REPLACEMENT TIME (RRT). THE PATIENT ELECTED TO KEEP THE DEVICE INTACT AFTER THE RRT AND PROGRAMMED THERAPIES ¿OFF¿. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8854219 ↗MalfunctionJul 11, 2019Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED THAT THERE WAS AN AUDIBLE ALERT FROM THE DEVICE DUE TO THE PATIENT BEING TOO CLOSE TO A MAGNET. FOLLOW UP CONCLUDED NO INTERVENTION WAS PERFORMED AND THE PATIENT IS NOW AWARE TO BE COGNIZANT WHEN AROUND MAGNETS AS A TONE WILL BE DELIVERED. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8782240 ↗InjuryJun 23, 2019Battery Problem; Inappropriate or Unexpected Reset; Battery Problem; Inappropriate or Unexpected Reset
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) WAS AT END OF SERVICE (EOS) AND EXPERIENCED A POWER ON RESET (POR) RESULTING IN LOST TRENDS/DATA. ALL DATA WAS CLEARED FROM THE DEVICE UPON INTERROGATION DUE TO ELECTRICAL RESET. THE DEVICE WAS REMOVED AND A NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8725537 ↗InjuryFeb 4, 2019Delayed Charge Time; Battery Problem; Delayed Charge Time; Battery Problem
IT WAS REPORTED THAT THE DEVICE REACHED RECOMMENDED REPLACEMENT TIME (RRT) AND THEN HAD AN EXCESSIVE CHARGE TIME ALERT WITH A CHARGE TIME OF GREATER THAN 16 SECONDS. THE DEVICE WAS REMOVED AND A NEW DEVICE WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8302699 ↗InjuryNov 1, 2018Delayed Charge Time; Delayed Charge Time
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED EXCESSIVE CHARGE TIME AND WENT STRAIGHT TO END OF SERVICE (EOS). THE DEVICE IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8028976 ↗InjuryApr 25, 2018Premature Discharge of Battery; Delayed Charge Time; Premature Discharge of Battery; Delayed Charge Time
IT WAS REPORTED THAT THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) PREMATURELY REACHED END OF SERVICE (EOS) DUE TO EXCESSIVE CHARGE TIME. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7460107 ↗InjuryFeb 8, 2018Premature Discharge of Battery; Premature Discharge of Battery
THE DEVICE WAS REMOVED AND A NEW DEVICE WAS IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7254953 ↗DeathApr 20, 2017Defibrillation/Stimulation Problem; Defibrillation/Stimulation Problem
IT WAS REPORTED THE PATIENT IS DECEASED. THE DECEDENT¿S SPOUSE STATED THE PATIENT HAD DIFFICULTY BREATHING DURING THE AMBULANCE RIDE, THE PATIENT HAD A CARDIAC ARREST, AND THE DEVICE NEVER SHOCKED. THE PATIENT WENT INTO A COMA. THE SPOUSE STATED A PHYSICIAN SAID, ¿INTERROGATION OF THE DEVICE FOUND IT WAS SET TOO HIGH.¿ A REQUEST FOR ADDITIONAL INFORMATION WAS MADE, BUT WAS NOT AVAILABLE.
Patient problems: Cardiac Arrest; Death
MAUDE report 6507795 ↗MalfunctionFeb 9, 2017Premature Discharge of Battery; Premature Discharge of Battery
IT WAS REPORTED THAT THE PATIENT'S IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) TRIGGERED ELECTIVE REPLACEMENT INDICATOR (ERI) EARLIER THAN ANTICIPATED. THE DEVICE CURRENTLY REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6318115 ↗
Other Implantable cardioverter defibrillator (non-CRT) devices
- EMBLEM™ MRI S-ICD A219Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)2 recalls36 deaths15,404 reports
- EMBLEM™ S-ICD A209Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)1 recall7 deaths6,788 reports
- EMBLEM™ S-ICD 3501Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)7 deaths4,936 reports
- Cobalt™ XT DR MRI SureScan™ DDPA2D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)3 recalls30 deaths1,603 reports
- ENDOTAK RELIANCE® G 0185Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,632 reports
- VIGILANT™ EL ICD DR D233Boston Scientific Corporation · Implantable cardioverter defibrillator (non-CRT)9 deaths1,563 reports
- Evera MRI™ XT DR SureScan® DDMB1D4Medtronic, Inc. · Implantable cardioverter defibrillator (non-CRT)1 recall35 deaths1,426 reports
- Durata™ 7122St. Jude Medical, Inc. · Implantable cardioverter defibrillator (non-CRT)27 deaths1,377 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page