Medtronic Reusable Instruments 9411905Recalls & reported problems

Medtronic Sofamor Danek, Inc. · RACK 9411905 19 22 25MM E · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994730107), received between Mar 25, 2021 and Mar 25, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Material Deformation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 25, 2021Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH L3/4 VERTEBRAL BODY COLLAPSE SUGGESTED FOR SPINAL THERAPY. POSTERIOR FIXATION WAS PERFORMED AT L1-S2. LEVELS IMPLANTED WAS L3- 4. EVENT OCCURRED INTRA-OP. IT WAS REPORTED THAT L3/4 VERTEBRAL SUBTOTAL REMOVAL WAS PERFORMED AND THE REPORTED IMPLANTS WERE PLACED. THE CAGE WAS LENGTHENED AFTER BEING INSERTED BETWEEN THE VERTEBRAL BODIES, BUT IT SHRANK BEFORE IT WAS FULLY LENGTHENED. WHEN THE CAGE WAS CHECKED OUTSIDE THE OPERATIVE FIELD, IT COULD BE LENGTHENED TO THE END WITHOUT ANY PROBLEM, BUT THE SAME EVENT WAS CONFIRMED IN THE OPERATIVE FIELD. IT WAS ATTEMPTED TO LENGTHEN THE CAGE BY INSERTING AN ADJUSTER INTO THE OPENING FOR FILLING BONE GRAFT, BUT THE CAGE COULDN'T BE LENGTHENED UNTIL THE END. THEREFORE, THE CAGE WAS REPLACED WITH ANOTHER ONE WITH ONE INCREASED SIZE. THE TORQUE LIMITING DRIVER WAS NOT USED. DELAY OF OVER 60 MINUTES WAS REPORTED. IMPLANT WAS EXPLANTED AND BEING RETURNED. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WAS REPORTED. NO DEFECT WAS FOUND IN INSERTER IN APPEARANCE. LOT#CA12G066 AND CA12J027 OF PINION DRIVER HAD SCRATCHES AT TIP. CA13F029-B HAD SCR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11567361 ↗
  • InjuryMar 25, 2021Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROFESSIONAL VIA MANUFACTURING REPRESENTATIVE REGARDING PATIENT WITH L3/4 VERTEBRAL BODY COLLAPSE SUGGESTED FOR SPINAL THERAPY. POSTERIOR FIXATION WAS PERFORMED AT L1-S2. LEVELS IMPLANTED WAS L3- 4. EVENT OCCURRED INTRA-OP. IT WAS REPORTED THAT L3/4 VERTEBRAL SUBTOTAL REMOVAL WAS PERFORMED AND THE REPORTED IMPLANTS WERE PLACED. THE CAGE WAS LENGTHENED AFTER BEING INSERTED BETWEEN THE VERTEBRAL BODIES, BUT IT SHRANK BEFORE IT WAS FULLY LENGTHENED. WHEN THE CAGE WAS CHECKED OUTSIDE THE OPERATIVE FIELD, IT COULD BE LENGTHENED TO THE END WITHOUT ANY PROBLEM, BUT THE SAME EVENT WAS CONFIRMED IN THE OPERATIVE FIELD. IT WAS ATTEMPTED TO LENGTHEN THE CAGE BY INSERTING AN ADJUSTER INTO THE OPENING FOR FILLING BONE GRAFT, BUT THE CAGE COULDN'T BE LENGTHENED UNTIL THE END. THEREFORE, THE CAGE WAS REPLACED WITH ANOTHER ONE WITH ONE INCREASED SIZE. THE TORQUE LIMITING DRIVER WAS NOT USED. DELAY OF OVER 60 MINUTES WAS REPORTED. IMPLANT WAS EXPLANTED AND BEING RETURNED. NO PATIENT SYMPTOMS OR FURTHER COMPLICATIONS WAS REPORTED. NO DEFECT WAS FOUND IN INSERTER IN APPEARANCE. LOT#CA12G066 AND CA12J027 OF PINION DRIVER HAD SCRATCHES AT TIP. CA13F029-B HAD SCR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11567353 ↗