Skip to content

Integrity INT35030UXRecalls & reported problems

Medtronic, Inc. · Stent INT35030UX INTEGRITY 3.50X30UX · Class 3 · Implant

All models of Integrity (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994798602), received between Apr 24, 2017 and Apr 24, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Deformation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 24, 2017Material Deformation; Material Deformation

    THE PHYSICIAN INTENDED TO USE AN INTEGRITY BARE METAL STENT TO TREAT A LESION LOCATED IN THE PROXIMAL SAPHENOUS VEIN GRAFT . IT WAS REPORTED THAT THE LESION WAS SEVERELY TORTUOUS, SEVERELY CALCIFIED WITH 70% STENOSIS. IN STENT RESTENOSIS WAS NOTED. NO DAMAGE WAS NOTED TO PACKAGING AND NO ISSUES WERE ENCOUNTERED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS NOT PERFORMED. IT WAS REPORTED THAT THE STENT WOULD NOT CROSS THE LESION DESPITE MULTIPLE ATTEMPTS TO POSITION THE INTEGRITY STENT. THE STENT DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT HOWEVER RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE BUT NO EXCESSIVE FORCE WAS USED. THE PHYSICIAN ALSO PRE-DILATED THE LESION THREE TIMES USING NC BALLOONS. PRIOR TO LAST ATTEMPT TO POSITION THE STENT, THE STENT HAD A 4X4 GAUZE LAYING ON IT. WHEN THE GAUZE WAS REMOVED IT CAUGHT ON THE STENT CAUSING IT TO BE STRETCHED AND DEFORMED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6515569 ↗