Skip to content

Integrity INT40018UXRecalls & reported problems

Medtronic, Inc. · Stent INT40018UX INTEGRITY 4.00X18UX · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
1
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994798640), received between Feb 23, 2017 and Feb 23, 2017 — classified as 1 death.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Burst Container or Vessel1
  • Detachment of Device or Device Component1

Patient problems

  • Death1
  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • DeathFeb 23, 2017Burst Container or Vessel; Detachment of Device or Device Component; Burst Container or Vessel; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PHYSICIAN IMPLANTED A 4.0 X 12 INTEGRITY BARE METAL STENT TO TREAT A MODERATELY CALCIFIED AND MILDLY TORTUOUS OSTIUM OF THE LEFT MAIN CORONARY ARTERY AND CX. THE LESION WAS PRE-DILATED. DURING THE PROCEDURE A CUTTING BALLOON WAS USED AND AN NC EUPHORA BALLOON WAS USED TO PRE-DILATE THE PROXIMAL CIRC AND ALSO TO SUPPORT ADVANCEMENT OF A NON MDT GUIDE EXTENSION CATHETER. THE GUIDE EXTENSION CATHETER WAS ADVANCED DISTAL TO THE LESION AND THE BALLOON WAS REMOVED. THE 4.0 X 12 INTEGRITY STENT WAS SUCCESSFULLY DELIVERED TO THE MID CIRC. THE 4.0 X 18 INTEGRITY STENT WAS INSERTED AND POSITIONED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY-DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED UPON ADVANCEMENT OF THE DEVICE BUT NO EXCESSIVE FORCE WAS USED. THE GUIDE EXTENSION CATHETER WAS PULLED BACK WHICH EXPOSED THE STENT AND THEN STENT DEPLOYMENT BEGAN. UPON INFLATING THE BALLOON DELIVERY SYSTEM TO APPROXIMATELY 7 ATM, RESISTANCE WAS LOST AND BLOOD TRAVELED BACK TO INFLATION DEVICE AND THE TECH REPORTED A BURST BALLOON. INITIALLY THE OPERATOR ATTEMPTED TO PULL BACK THE ENTIRE SYSTEM BUT IT WAS ¿STUCK¿ . A FEW SECONDS LATER AS THEY PULLED BACK AGAIN THE DELIVER Y SYSTEM WAS

    Patient problems: Death; Device Embedded In Tissue or Plaque

    MAUDE report 6352592 ↗

Other STENT, CORONARY devices

All STENT, CORONARY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page