Skip to content

Integrity INT22508UXRecalls & reported problems

Medtronic, Inc. · Stent INT22508UX INTEGRITY 2.25X08UX · Class 3 · Implant

All models of Integrity (27)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994798688), received between Dec 18, 2018 and Dec 18, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1

Patient problems

  • Patient Problem/Medical Problem1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 18, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    AN ATTEMPT WAS MADE TO USE ONE INTEGRITY RX CORONARY BARE METAL STENT TO TREAT A SEVERELY TORTUOUS, NON-CALCIFIED LESION IN THE LEFT MID CEREBRAL ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT WHILE ATTEMPTING TO DEPLOY THE STENT IN THE LEFT MID CEREBRAL ARTERY THAT RESISTANCE WAS MET AND THE STENT COULD NOT REACH THE LESION. WHILE REMOVING THE STENT AND CATHETER AS A UNIT, RESISTANCE WAS ENCOUNTERED AND FOLLOWING A TUG ON THE STENT, THE STENT DISLODGED INTO THE CAROTID ARTERY. THE DISLODGED STENT WAS REMOVED USING A SNARE. THE PATIENT IS REPORTED TO BE ALIVE WITH NO FURTHER INJURY POST PROCEDURE.

    Patient problems: Patient Problem/Medical Problem

    MAUDE report 8173804 ↗

Other STENT, CORONARY devices

All STENT, CORONARY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page