Skip to content

Launcher LA6JR40SH (UDI 00613994817631)Recalls & reported problems

Medtronic, Inc. · CATHETER LA6JR40SH LA 6F 100CM JR40 · Class 2

All models of Launcher (64)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00613994817631), received between Mar 6, 2020 and Mar 6, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Material Deformation” (1 report).

52 other recalls cover devices cleared under the same FDA submission (K021256); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Deformation1

Patient problems

  • Vascular Dissection1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 6, 2020Material Deformation; Material Deformation

    A LAUNCHER GUIDE CATHETER WAS USED DURING A PROCEDURE TO TREAT A NON-CALCIFIED, NON-TORTUOUS MID BRACHIAL ARTERY. THERE WAS NO DAMAGE NOTED TO THE PACKAGING, THE DEVICE WAS REMOVED FROM PACKAGING PER IFU WITH NO ISSUES. THE DEVICE WAS INSPECTED WITH NO ISSUES. THE DEVICE WAS PREPPED PER IFU WITH NO ISSUES. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT THE TIP OF THE DEVICE WAS NOT CYLINDRICAL, BUT HAD A SHARP CORNER THAT CAUSED VESSEL DAMAGE/DISSECTION. THE PRODUCT WAS REPLACED WITH A NON MEDTRONIC DEVICE. THE PATIENT STATUS POST PROCEDURE IS ALIVE WITH NO INJURY.

    Patient problems: Vascular Dissection

    MAUDE report 9797372 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K021256. One submission often covers many models, so check the product and lot details in each record.

  • Z-1248-2021 · Jan 15, 2021 · Terminated
    Medtronic 6F Launcher Guide Catheter 100 cm EBU 3.0 Model Number: LA6EBU30
  • Z-1246-2021 · Jan 15, 2021 · Terminated
    Medtronic 6F Launcher Guide Catheter 100 cm JR 4.0 Model Number: LA6JR40
  • Z-1245-2021 · Jan 15, 2021 · Terminated
    Medtronic 6F Launcher Guide Catheter 100 cmJL 3.5 Model Number: LA6JL35
  • Z-1247-2021 · Jan 15, 2021 · Terminated
    Medtronic 6F Launcher Guide Catheter 100cm EBU 3.75 Model Number: LA6EBU375
  • Z-0205-2015 · Oct 13, 2014 · Terminated
    Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K
  • Z-1564-2013 · Jun 12, 2013 · Terminated
    Medtronic, Launcher Guiding Catheter, Model # LA8EBU45SH, Lot # 0006722866, 8F and Model # LA6IMAD, Lot # 0006724224, 6F Product Usage: The Medtronic Guiding Catheter is designe
  • Z-0959-06 · Apr 5, 2006 · Terminated
    CATH. GUIDE 6F SR3.5 W/SH LAUN CATHETER LA6SR35SH LA 6F 100CM SR35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are
  • Z-0963-06 · Apr 5, 2006 · Terminated
    CATH. GUIDE 6F SR4.5 LAUNCHER CATHETER LA6SR45 LA 6F 100CM SR45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int
  • Z-1008-06 · Apr 5, 2006 · Terminated
    CATHETER SB6SR50 SB 6F 100CM S Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri
  • Z-0999-06 · Apr 5, 2006 · Terminated
    CATHETER SA6SR40 SA 6F 100CM Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or perip
  • Z-0947-06 · Apr 5, 2006 · Terminated
    CATH. GUIDE 6F JL4.0 5 PACK LA Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri
  • Z-1004-06 · Apr 5, 2006 · Terminated
    CATHETER SB6SR35 SB 6F 100CM S Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri
  • Z-0944-06 · Apr 5, 2006 · Terminated
    CATH GUIDE 6F LAUNCHER EBU 4.5 CATHETER LA6EBU45A LA 6F 110CM EB45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are
  • Z-1003-06 · Apr 5, 2006 · Terminated
    CATHETER SB6SR30 SB 6F 100CM S Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri
  • Z-0950-06 · Apr 5, 2006 · Terminated
    CATHETER LA6JL45A 6F 110CM JL4 CATHETER LA6JL45A 6F 110CM JL45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are intr

…and 37 more.

Other CATHETER, PERCUTANEOUS devices

All CATHETER, PERCUTANEOUS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page