Launcher LA6AL75Recalls & reported problems
Medtronic, Inc. · CATHETER LA6AL75 LA 6F 100CM AL75 · Class 2
All models of Launcher (64)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994820624), received between Mar 4, 2021 and Feb 1, 2023 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
52 other recalls cover devices cleared under the same FDA submission (K021256); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Detachment of Device or Device Component1
- Material Deformation1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Embolism/Embolus1
- Insufficient Information1
- Vascular Dissection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryFeb 1, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE INDEX PROCEDURE TWO LAUNCHER GUIDE CATHETERS WERE USED. DURING THE PROCEDURE A DISSECTION OCCURRED. IT WAS STATED THAT A MEDTRONIC LAUNCHER DEVICE WAS THE CAUSE OF THE DISSECTION.
Patient problems: Vascular Dissection
MAUDE report 16276217 ↗InjuryJan 20, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
DURING THE INDEX PROCEDURE ONE LAUNCHER GUIDE CATHETER WAS USED TO TREAT A LESION IN THE RIGHT CORONARY ARTERY (RCA). ON THE SAME DAY AS THE PROCEDURE IT WAS REPORTED THAT A DISTAL EMBOLIZATION POTENTIALLY OCCURRED.
Patient problems: Embolism/Embolus
MAUDE report 16213362 ↗InjuryMay 17, 2022Detachment of Device or Device Component; Material Deformation; Detachment of Device or Device Component; Material Deformation
DURING A PROCEDURE, AN ATTEMPT WAS MADE TO USE ONE 6F LAUNCHER GUIDE CATHETER TO TREAT A MODERATELY TORTUOUS, SEVERELY CALCIFIED LESION EXHIBITING 75-90% STENOSIS LOCATED IN THE PROXIMAL RIGHT CORONARY ARTERY (RCA). THERE WAS NO DAMAGE NOTED TO THE PACKAGING. THE DEVICE WAS REMOVED FROM PACKAGING PER IFU WITH NO ISSUES NOTED. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED DURING INSERTION/DELIVERY. THE DEVICE WAS NOT EXCESSIVELY TORQUED. IT WAS REPORTED THAT THE CATHETER KINKED / SNAPPED IN THE ILIAC AREA WHILE IN THE PATIENT. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 14424743 ↗MalfunctionMar 4, 2021Use of Device Problem; Use of Device Problem
DURING A PROCEDURE, AN ATTEMPT WAS MADE TO USE ONE 6 FRENCH LAUNCHER GUIDE CATHETER TO TREAT A MODERATELY TORTUOUS, SEVERELY CALCIFIED LESION EXHIBITING 90-95% STENOSIS LOCATED IN THE MID RIGHT CORONARY ARTERY (RCA). THE LAUNCHER WAS BEING USED FOR ORBITAL ARTHERECTOMY AND IVUS. THERE WAS NO DAMAGE NOTED TO THE PACKAGING. THE DEVICE WAS REMOVED FROM PACKAGING AND PREPPED PER IFU WITH NO ISSUES NOTED. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE AND EXCESSIVE FORCE WAS NOT USED DURING INSERTION/DELIVERY. IT WAS REPORTED THAT THE LAUNCHER GUIDE CATHETER WAS USED ONE WEEK AFTER THE PRODUCT EXPIRED. IT WAS STATED THAT NO ISSUES WERE OBSERVED OR ENCOUNTERED DURING THE PROCEDURE. THE PATIENT WAS REPORTED TO BE ALIVE WITH NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11412330 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K021256. One submission often covers many models, so check the product and lot details in each record.
- Z-1248-2021 · Jan 15, 2021 · TerminatedMedtronic 6F Launcher Guide Catheter 100 cm EBU 3.0 Model Number: LA6EBU30
- Z-1245-2021 · Jan 15, 2021 · TerminatedMedtronic 6F Launcher Guide Catheter 100 cmJL 3.5 Model Number: LA6JL35
- Z-1247-2021 · Jan 15, 2021 · TerminatedMedtronic 6F Launcher Guide Catheter 100cm EBU 3.75 Model Number: LA6EBU375
- Z-1246-2021 · Jan 15, 2021 · TerminatedMedtronic 6F Launcher Guide Catheter 100 cm JR 4.0 Model Number: LA6JR40
- Z-0205-2015 · Oct 13, 2014 · TerminatedMedtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K
- Z-1564-2013 · Jun 12, 2013 · TerminatedMedtronic, Launcher Guiding Catheter, Model # LA8EBU45SH, Lot # 0006722866, 8F and Model # LA6IMAD, Lot # 0006724224, 6F Product Usage: The Medtronic Guiding Catheter is designe
- Z-0950-06 · Apr 5, 2006 · TerminatedCATHETER LA6JL45A 6F 110CM JL4 CATHETER LA6JL45A 6F 110CM JL45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are intr
- Z-0965-06 · Apr 5, 2006 · TerminatedCATH. GUIDE 6F SR5.0 LAUNCHER CATHETER LA6SR50 LA 6F 100CM SR50 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are int
- Z-0946-06 · Apr 5, 2006 · TerminatedCATH. GUIDE 6F JL3.5 110CM LAU CATHETER LA6JL35A LA 6F 110CM JL35 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are i
- Z-0966-06 · Apr 5, 2006 · TerminatedCATH. GUIDE 6F SR5.0 W/SH LAUN CATHETER LA6SR50SH LA 6F 100CM SR50 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are
- Z-0944-06 · Apr 5, 2006 · TerminatedCATH GUIDE 6F LAUNCHER EBU 4.5 CATHETER LA6EBU45A LA 6F 110CM EB45 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are
- Z-0941-06 · Apr 5, 2006 · TerminatedCATH GUIDE 6F LAUNCHER AL 3.0 CATHETER LA6AL30A LA 6F 110CM AL30 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are in
- Z-0943-06 · Apr 5, 2006 · TerminatedCATH. GUIDE 6F EBU4.0 110CM LA CATHETER LA6EBU40A LA 6F 110CM EB40 Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are
- Z-1000-06 · Apr 5, 2006 · TerminatedCATHETER SA6SR50 SA 6F 100CM Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri
- Z-1003-06 · Apr 5, 2006 · TerminatedCATHETER SB6SR30 SB 6F 100CM S Coronary Guiding catheter designed to provide a pathway through which therapeutic and/or diagnostic devices are introduced into the coronary or peri
…and 37 more.
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page