Skip to content

SPRINTER® LEGEND SPL25015XRecalls & reported problems

Medtronic, Inc. · BLN SPL25015X SPR LEG 2.5X15 RX US M · Class 2

All models of SPRINTER LEGEND (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00613994902597), received between Jul 4, 2018 and Mar 7, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel2
  • Deflation Problem2
  • Activation, Positioning or Separation Problem1
  • Fracture1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Chest Pain1
  • Device Embedded In Tissue or Plaque1
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Pain1
  • Patient Problem/Medical Problem1
  • Perforation1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 7, 2024Burst Container or Vessel; Burst Container or Vessel

    AN ATTEMPT WAS MADE TO USE ONE SPRINTER LEGEND RX CORONARY BALLOON CATHETER TO TREAT A LESION IN THE LEFT MAIN (LM) CORONARY ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT THE BALLOON BROKE AND PERFORATED THE LM ARTERY. THE BALLOON'S INITIAL INFLATION RESULTED IN THE PERFORATION OF THE LM ARTERY. IT WAS STATED THAT PART OF THE BALLOON REMAINS IN THE BODY. A STENT WAS PLACED AND PINNED UP THE PIECE OF BALLOON LEFT IN THE CORONARY ARTERY. THE PATIENT IS ALIVE WITH NO FURTHER INJURY.

    Patient problems: Device Embedded In Tissue or Plaque; Insufficient Information; Perforation

    MAUDE report 18855164 ↗
  • MalfunctionAug 19, 2020Activation, Positioning or Separation Problem

    ELDERLY FEMALE WITH HISTORY OF CHEST PAIN. PRESENTED WITH BILATERAL ARM PAIN AS WELL AS JAW PAIN. WENT TO CATH LAB AND WHILE TESTING THE DEVICE, THE METAL STYLET WOULD NOT LEAVE THAT BALLOON AS USUAL.

    Patient problems: Pain

    MAUDE report 10424799 ↗
  • MalfunctionMar 2, 2020Deflation Problem; Deflation Problem

    DURING A PROCEDURE, ONE SPRINTER LEGEND RX PTCA BALLOON CATHETER WAS USED TO TREAT A LESION. IT WAS REPORTED THAT THERE WAS INFLATION DIFFICULTIES DURING BALLOON INFLATION. IT WAS ALSO REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED AND THE DEVICE WOULD NOT DEFLATE AT THE LESION SITE. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9773359 ↗
  • MalfunctionJan 30, 2020Burst Container or Vessel; Burst Container or Vessel

    IT WAS REPORTED THAT A SPRINTER LEGEND RX PTCA BALLOON RUPTURED WHILE INSIDE THE PATIENT. NO PATIENT INJURY REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9646604 ↗
  • MalfunctionDec 18, 2019Fracture

    AN ELDERLY MALE PATIENT PRESENTS WITH A HISTORY SIGNIFICANT FOR TYPE 1 DIABETES, HYPERTENSION, HYPERLIPIDEMIA, CORONARY ARTERY DISEASE STATUS POST CORONARY ARTERY BYPASS GRAFTING (CABG), GASTROESOPHAGEAL REFLUX DISEASE (GERD), AND ACUTE RESPIRATORY FAILURE. WHILE PERFORMING A HEART CATHETERIZATION, SYNERGY STENT DEPLOYED ANGIOGRAMS DONE. STENT APPEARED TO HAVE FRACTURE. INTRAVASCULAR ULTRASOUND (IVUS) PROCEDURE DONE TO VERIFY. SECOND STENT DEPLOYED OVER FRACTURED STENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9488668 ↗
  • MalfunctionJul 4, 2018Deflation Problem; Deflation Problem

    THE PATIENT PRESENTED WITH ACS, CRUSHING CHEST PAIN AND ST ELEVATION. ONE SPRINTER LEGEND PTCA BALLOON CATHETER WAS ATTEMPTED TO BE USED TO TREAT A LESION IN THE PROXIMAL CIRCUMFLEX (CX) ARTERY WITH MILD TORTUOSITY AND CALCIFICATION AND 95% STENOSIS. THERE WERE NO ABNORMALITIES REPORTED IN RELATION TO PATIENT'S ANATOMY. THERE WAS NO DAMAGE TO THE PACKAGING. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICES FROM THE HOOP/TRAY. THE DEVICES WERE INSPECTED WITH NO ISSUES. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. NO RESISTANCE WAS NOTED WHILE ADVANCING THE BALLOON. THE DEVICE WAS NOT MOVED OR REPOSITIONED IN THE LESION. THERE WAS NO INFLATION DIFFICULTY. IT WAS REPORTED THAT THE BALLOON FAILED TO DEFLATE AT THE LESION SITE, AFTER FIRST INFLATION. AN ATTEMPT WAS MADE TO PERFORATE THE BALLOON WITH A STIFF WIRE. A GUIDELINER WAS THEN USED TO PARTIALLY PULL THE BALLOON BACK INTO THE GUIDELINER AND ALL WAS REMOVED. THE PATIENT EXPERIENCED SOME CHEST DISCOMFORT DURING THE EVENT BUT IT SUBSIDED ONCE THE BALLOON WAS REMOVED. THE WHOLE EVENT LASTED APPROX. 10 MINUTES. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: Chest Pain; Patient Problem/Medical Problem

    MAUDE report 7661850 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →