SPRINTER® LEGEND SPL25012XRecalls & reported problems

Medtronic, Inc. · BLN SPL25012X SPR LEG 2.5X12 RX US M · Class 2

All models of SPRINTER LEGEND (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00613994902832), received between Jun 27, 2018 and Jan 8, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Deflation Problem” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Deflation Problem3
  • Material Deformation2
  • Detachment of Device or Device Component1
  • Inflation Problem1
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Consequences Or Impact To Patient1
  • Non specific EKG/ECG Changes1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 8, 2024Inflation Problem; Leak/Splash; Material Deformation; Inflation Problem; Leak/Splash; Material Deformation

    AN ATTEMPT WAS MADE TO USE ONE SPRINTER LEGEND RX BALLOON CATHETER TO TREAT A LESION. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT INFLATION DIFFICULTIES OCCURRED DURING BALLOON INFLATION AT 8ATM. IT WAS DETAILED THAT THE BALLOON WOULD NOT INFLATE DURING INFLATION, IT WAS THEN OBSERVED UNDER FLUOROSCOPY THAT THE BALLOON WAS LEAKING. THE PATIENT IS ALIVE WITH NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18462247 ↗
  • MalfunctionJun 1, 2022Deflation Problem; Material Deformation; Deflation Problem; Material Deformation

    ONE SPRINTER LEGEND BALLOON CATHETER WAS ATTEMPTED TO BE USED TO TREAT A LESION WITH 90% STENOSIS IN THE DISTAL RIGHT CORONARY ARTERY (RCA). THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. THE BALLOON WAS INFLATED TO 12 ATM FOR 22 SECONDS AND THEN FOR 21 SECONDS.  IT WAS REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED AND THE BALLOON WOULD NOT DEFLATE AT THE LESION SITE. THE BALLOON WAS INFLATED TO 12 ATM, AN ATTEMPT WAS MADE TO COME DOWN AND GO BACK UP TO 12 ATM AND THEN ATTEMPTED TO COME DOWN AGAIN. THE INFLATION DEVICE WAS CHANGED BUT THAT DID NOT WORK SO THE BALLOON WAS THEN PULLED BACK INTO THE GUIDE AND REMOVED FROM THE PATIENT. A RESOLUTE ONYX STENT WAS DEPLOYED IN THE DISTAL RCA DURING THE PROCEDURE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14562535 ↗
  • MalfunctionMay 13, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    ONE SPRINTER LEGEND RX BALLOON CATHETER WAS ATTEMPTED TO BE USED TO TREAT A 75% CALCIFIED, NON TORTUOUS LESION IN THE PROXIMAL CIRCUMFLEX. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP. IT WAS DIFFICULT TO REMOVE THE PROTECTIVE SHEATH, THERE WAS NO DIFFICULTY REMOVING THE PACKAGING STYLETTE. THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. THE LESION WAS NOT PRE-DILATED. THE TRAVERSE GUIDEWIRE WAS 300CM. THE GUIDEWIRE WAS USED SUCCESSFULLY WITH OTHER DEVICES PRIOR TO THE DIFFICULTY OCCURRING. THE GUIDEWIRE WAS FRONT LOADED THROUGH THE WIRE ENTRY PORT. THERE WERE NO DIFFICULTIES NOTED WHEN LOADING THE DEVICE ONTO THE GUIDEWIRE. IT WAS REPORTED THAT THERE WAS POOR GUIDE WIRE MOVEMENT DURING THE PROCEDURE. IT WAS ATTEMPTED TO RE-POSITION THE WIRE AND BRING IT BACK TO THE BALLOON. IT WAS STATED THAT THE TRAVERSE WIRE GOT STUCK IN THE BALLOON. THE WIRE AND THE BALLOON WERE THEN REMOVED AS ONE UNIT. IT WAS INDICATED THAT THE SPRINTER LEGEND DEVICE LOOKED DAMAGED DUE TO THE ATTEMPTS TO REMOVE THE WIRE. NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11818399 ↗
  • InjuryMay 4, 2021Deflation Problem; Deflation Problem

    DURING A PROCEDURE, AN ATTEMPT WAS MADE TO USE ONE SPRINTER LEGEND RX PTCA BALLOON CATHETER TO TREAT A NON-TORTUOUS, NON-CALCIFIED LESION LOCATED IN THE MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE DEVICE WAS INSPECTED WITH NO ISSUES NOTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES NOTED. RESISTANCE WAS NOT ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT BALLOON DEFLATION DIFFICULTIES OCCURRED AND THE DEVICE WOULD NOT DEFLATE AT THE LESION SITE. THE DEVICE MOVED IN THE LESION WHILE INFLATED. THE DEVICE WAS PULLED OUT. THE PATIENT WAS REPORTED TO BE ALIVE WITH INJURY. THE EVENT RESULTED IN STELEVATION AND THE PATIENT IS IN HOSPITAL ON ECMO.

    Patient problems: Non specific EKG/ECG Changes; Unspecified Tissue Injury

    MAUDE report 11768456 ↗
  • MalfunctionJun 27, 2018Deflation Problem; Deflation Problem

    ONE SPRINTER LEGEND PTCA BALLOON CATHETER WAS ATTEMPTED TO BE USED TO PRE-DILATE A LESION. THERE WAS NO DAMAGE NOTED TO THE DEVICE PACKAGING. THERE WERE NO ISSUES NOTED WHEN REMOVING THE DEVICE FROM THE HOOP/TRAY. THE DEVICE WAS NOT INSPECTED. NEGATIVE PREP WAS PERFORMED WITH NO ISSUES. NO DIFFICULTIES NOTED WHEN REMOVING THE PROTECTIVE SHEATH OR STYLETTE FROM THE DEVICE. THE LESION WAS PRE-DILATED. THE DEVICE DID NOT PASS THROUGH A PREVIOUSLY DEPLOYED STENT. NO RESISTANCE WAS ENCOUNTERED WHEN ADVANCING THE DEVICE. EXCESSIVE FORCE WAS NOT USED DURING DELIVERY. IT WAS REPORTED THAT THE BALLOON WAS INFLATED TO NOMINAL PRESSURE AND WAS UNABLE TO DEFLATE AT THE LESION SITE AFTER THE FIRST INFLATION. THE DEVICE HAD NOT BEEN MOVED OR RE-POSITIONED PRIOR TO DEFLATION DIFFICULTIES. A 50/50 CONTRAST TO SALINE SOLUTION WAS USED. THE ENTIRE DELIVERY SYSTEM WAS THEN PULLED OUT, WITH THE WIRE, FROM THE VESSEL. IT WAS INFORMED THAT THE LARGE ARTERY ALLOWED FOR THE SUCCESSFUL REMOVAL OF THE DEVICE WITH NO ISSUES. A NON-MEDTRONIC BALLOON CATHETER WAS THEN USED TO PRE-DILATE AND STENT WAS SUCCESSFULLY DEPLOYED AT THE LESION SITE. NO PATIENT INJURY IS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7641335 ↗

Other Catheters, transluminal coronary angioplasty, percutaneous devices

All Catheters, transluminal coronary angioplasty, percutaneous devices →