CD HORIZON® Spinal System 1604200500Recalls & reported problems

Medtronic Sofamor Danek, Inc. · ROD 1604200500 6.0 TIAL NS STRT LN 500MM · Class 2 · Implant

All models of CD HORIZON Spinal System (342)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00613994990693), received between Jun 2, 2017 and Nov 18, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Break” (4 reports).

1 other recall cover devices cleared under the same FDA submission (K113174); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Insufficient Device Problem Information1

Patient problems

  • Pain3
  • No Clinical Signs, Symptoms or Conditions1
  • Nonunion/Delayed-union Bone Fracture1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 18, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING FUSION SPINAL T HERAPY. IT WAS REPORTED THAT THE TULIP FROM THE MAS-SCREW HAD COME LOOSE FROM THE SCREW AND THE RODS IN THE CONSTRUCTION WERE BROKEN. THE BROKEN TULIP WAS ON LEFT L4 AND THE ROD ON THE LEFT SIDE WAS BROKEN ON TWO SPOTS WHILE THE ROD ON THE RIGHT SIDE WAS BROKEN IN ONE POSITION. THERE WAS PAIN IN THE AREA OF THE BACK WHERE THE BROKEN SCREW WAS LOCATED NEAR L4. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE ROD BELONGS TO MDT.

    Patient problems: Pain

    MAUDE report 23584762 ↗
  • InjuryNov 18, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING FUSION SPINAL T HERAPY. IT WAS REPORTED THAT THE TULIP FROM THE MAS-SCREW HAD COME LOOSE FROM THE SCREW AND THE RODS IN THE CONSTRUCTION WERE BROKEN. THE BROKEN TULIP WAS ON LEFT L4 THERE WAS PAIN IN THE AREA OF THE BACK WHERE THE BROKEN SCREW WAS LOCATED NEAR L4. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION RECEIVED FROM MANUFACTURER REPRESENTATIVE THAT THE ROD BELONGS TO MDT.

    Patient problems: Pain

    MAUDE report 23584265 ↗
  • InjuryAug 21, 2024Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING HARDWARE REMOVA L/REVISION. PRE-OPERATIVE DIAGNOSIS FOR THIS PROCEDURE WAS NON-UNION DUE TO HARDWARE ISSUES. IT WAS REPORTED THAT WHEN THE DRIVER WAS USED TO REMOVE A SOLERA SCREW, THE TIP OF THE DRIVER WAS BROKEN. THE BROKEN TIP WAS RETRIEVED.  THE PATIENT/PROCEDURE WAS NOT IMPACTED IN ANY WAY. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Nonunion/Delayed-union Bone Fracture

    MAUDE report 20020330 ↗
  • InjuryJun 2, 2017Break; Break

    IT WAS REPORTED THAT PATIENT WITH ADULT DEGENERATIVE DISEASE AND SCOLIOSIS UNDERWENT SPINAL SURGERY FROM T10 - PELVIS. POST-OP ROD BREAKAGE AT L5-S1 WAS OBSERVED CAUSING PAIN TO THE PATIENT. IT IS NOT CONFIRMED WHETHER SURGEON WILL PERFORM REVISION SURGERY ON THE PATIENT.

    Patient problems: Pain

    MAUDE report 6607197 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113174. One submission often covers many models, so check the product and lot details in each record.

  • Z-2603-2018 · Jul 9, 2018 · Terminated
    Medtronic CD HORIZON Spinal System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, or sacral spine.

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →