Endurant® II ETLW1610C156E (UDI 00613994991102)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT ETLW1610C156E ENDUR II LIMB · Class 3 · Implant

All models of Endurant II (114)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994991102), received between Jul 12, 2017 and Apr 21, 2021 — classified as 2 injuries.

The most frequently reported specific device problem is “Leak/Splash” (1 report).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Leak/Splash1
  • Migration or Expulsion of Device1

Patient problems

  • Injury1
  • Obstruction/Occlusion1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ELECTIVE ENDOVASCULAR TREATMENT OF A 50MM AAA. IT WAS REPORTED DURING THE INDEX PROCEDURE, THE ETBF3216C145E WAS EXTENDED WITH AN ETLW1624C124E TO SEAL PROXIMAL TO THE RIGHT INTERNAL ILIAC AND THEN ETLW1610C156E EXTENDED WITH ETEW1010C82E INTO THE LEFT EXTERNAL AFTER LEFT INTERNAL EMBOLIZATION WITH AN NON-MDT PLUG. APPROX. 6 YEARS 4 MONTHS POST IMPLANT, THE PATIENT PRESENTED WITH A SHRINKING 4.4CM FUSIFORM AAA WITH AN OCCLUDED LEFT LIMB THAT EXTENDS INTO THE TORTUOUS LEFT EXTERNAL ILIAC.  THE RIGHT COMMON ILIAC MEASURED APPROX. 23MM , WHICH IS SEALED WITH A 24MM FLARED LIMB PROXIMAL TO THE RIGHT INTERNAL ILIAC OSTIA. PER THE PHYSICIAN THE CAUSE OF THE OCCLUSION OCCURRED DUE TO COMPRESSION OF THE ETEW1010C82E IN TORTUOUS ANATOMY. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT WILL BE MONITORED

    Patient problems: Obstruction/Occlusion

    MAUDE report 11700437 ↗
  • InjuryJul 12, 2017Leak/Splash; Migration or Expulsion of Device; Leak/Splash; Migration or Expulsion of Device

    AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN THE PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY AT IMPLANT WAS NOT REPORTED. IT WAS REPORTED THAT AT APPROXIMATELY 23 MONTHS POST IMPLANT AN ANGIOGRAM REVEALED A TYPE II ENDOLEAK. THE PHYSICIAN ELECTED TO COIL THE HYPOGASTRIC ARTERY AND IMPLANT A 16 X 10 X 156 ENDURANT LIMB THAT WAS EXTENDED INTO THE RIGHT EXTERNAL ILIAC ARTERY. THE LEFT ILIAC LIMB WAS EXTENDED WITH A 16 X 24 X 93 ENDURANT LIMB ABOVE THE LEFT HYPOGASTRIC ARTERY AND A 32 X 32 X 49 ENDURANT CUFF WAS IMPLANTED PROXIMALLY. IT WAS REPORTED THAT 24 MONTHS LATER ADDITIONAL A RECENT ULTRASOUND REVEALED THAT THERE WAS ANEURYSM ENLARGEMENT THAT WAS DUE TO AN UNKNOWN ENDOLEAK. A CT WAS NOT PERFORMED. THE NEXT DAY AN AORTOGRAM REVEALED A DISTAL TYPE III SEPARATION ENDOLEAK, THE 16 X 10 X 156 ENDURANT STENT GRAFT HAD BEEN PULLED OUT AND SEPARATED FROM THE 16 X 24 X 93 ENDURANT LIMB ON THE RIGHT ILIAC LIMB FROM THE ORIGINAL IMPLANT. THE PHYSICIAN IMPLANTED A 16 X 16 X 124 ENDURANT ILIAC LIMB TO RECONNECT AND CLOSE THE GAP BETWEEN THE LIMBS. THIS WAS SUCCESSFULLY DONE AND THE ILIAC LIMB WAS BALLOONED. ANOTHER AORTOGRAM WAS PERFORMED AND

    Patient problems: Injury

    MAUDE report 6707669 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoili
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infra
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166