Skip to content

Endurant® II ETLW1613C124E (UDI 00613994991546)Recalls & reported problems

Medtronic, Inc. · STENT GRAFT ETLW1613C124E ENDUR II LIMB · Class 3 · Implant

All models of Endurant II (114)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00613994991546), received between Jun 27, 2017 and Apr 5, 2018 — classified as 2 injuries.

The most frequently reported specific device problem is “Leak/Splash” (2 reports).

6 other recalls cover devices cleared under the same FDA submission (P100021); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Leak/Splash2

Patient problems

  • Injury1
  • Pain1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 5, 2018Leak/Splash; Leak/Splash

    AN ENDURANT II STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM OF UNKNOWN SIZE. IT WAS REPORTED THAT APPROXIMATELY 4.5 YEARS POST INDEX PROCEDURE A TYPE IB ENDOLEAK WAS NOTED ON ANGIOGRAM. INTERVENTION WAS PERFORMED WITH THE IMPLANTATION OF AN ENDURANT LIMB (16X20X93CM) TO TREAT A TYPE IB ENDOLEAK IN THE RIGHT COMMON ILIAC. THE PHYSICIAN NOTED THAT HE OBTAINED ADEQUATE SEAL DISTAL TO THE EXISTING LIMB BY EXTENSION OF ABOUT 15MM DOWN INTO THE RIGHT INTERNAL ILIAC ARTERY. THE EVENT WAS RESOLVED. AS PER THE PHYSICIAN, THE CAUSE OF THE EVENT WAS ANATOMY RELATED DUE TO DILATION OF THE RIGHT COMMON ILIAC SINCE THE INDEX PROCEDURE. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: Injury

    MAUDE report 7399109 ↗
  • InjuryJun 27, 2017Leak/Splash; Leak/Splash

    AN ENDURANT II STENT GRAFT SYSTEM WAS SELECTED FOR USE IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT THE PATIENT PRESENTED EMERGENTLY WITH BACK AND ABDOMINAL PAIN. THE CT REVEALED THAT THERE WAS A TYPE III SEPARATION ENDOLEAK BETWEEN THE ENDURANT BIFURCATED STENT GRAFT AND THE ENDURANT 16X13X124. CURRENTLY THE ABDOMINAL AORTIC ANEURYSM IS 7.3 CM IN DIAMETER AND 9.3 CM IN LENGTH. THE PHYSICIAN ELECTED TO IMPLANT AN ENDURANT 16X16X156 ILIAC LIMB. THE FINAL ANGIOGRAM DEMONSTRATED THAT THE TYPE III SEPARATION ENDOLEAK WAS RESOLVED. TWO DAYS LATER THE FOLLOW UP CT REVEALED THAT THERE WAS AN UNKNOWN ENDOLEAK IN THE AREA OF THE JUNCTION OF THE BIFURCATED STENT GRAFT AND THE ENDURANT 16X20X156 ILIAC LIMB ON THE OTHER SIDE. THE PHYSICIAN ELECTED TO IMPLANT AN ENDURANT 16X20X124 ILIAC LIMB. THE UNKNOWN ENDOLEAK WAS RESOLVED. THE PHYSICIAN STATED THAT THE CAUSE OF BOTH, THE TYPE III SEPARATION ENDOLEAK AND THE UNKNOWN ENDOLEAK WERE RELATED TO THE PATIENT¿S ANATOMY; DISEASE PROGRESSION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

    Patient problems: Pain

    MAUDE report 6671738 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P100021. One submission often covers many models, so check the product and lot details in each record.

  • Z-0292-2022 · Oct 15, 2021 · Open, Classified
    Endurant IIs Stent Graft System REF/Description: ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STENT GRAFT ESBF2814C103EE ENDUR IIS BIF ESBF2814C103EE STEN
  • Z-0290-2022 · Oct 15, 2021 · Open, Classified
    Endurant Stent Graft System REF/Description: ENTF2525C70EE STENT GRAFT ENTF2525C70EE ENDURANT TUBE
  • Z-0291-2022 · Oct 15, 2021 · Open, Classified
    Endurant II Stent Graft System REF/Description: ETBF2513C145EE STENT GRAFT ETBF2513C145EE ENDUR II BIF ETBF2820C166EE STENT GRAFT ETBF2820C166EE ENDUR II BIF ETBF2820C166EE STENT G
  • Z-2135-2021 · Jun 7, 2021 · Open, Classified
    Endurant II/Endurant IIs Stent Graft System: Medtronic Endurant II Stent Graft System REF ETUF3214C102EE, ETCF3636C49EE, ETBF3216C166EE, and ETUF3614C102E. Used to treat infra
  • Z-2136-2021 · Jun 7, 2021 · Open, Classified
    Medtronic Endurant IIs Stent Graft System Medtronic Endurant IIs Stent Graft System REF ESBF3614C103EE and ESBF3214C103E. Used to treat infrarenal abdominal aortic or aortoili
  • Z-1543-2017 · Feb 27, 2017 · Terminated
    Endurant / Endurant II 23mm and 25mm Bifurcated Stent Graft Systems Model Numbers: ETBF2516C145E ETBF2516C166E ETBF2316C145E ETBF2313C145E ETBF2513C166E ETBF2313C166E ETBF2316C166

Other SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices

All SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page