Infinity Internal Hex System 203708Recalls & reported problems

Ace Surgical Supply CO., Inc. · infinity INTERNAL 3.7x8.0MM Ø3.5HD HEX IMPLANT 6AL-4V ELI · Class 2 · Implant

Quick summary

This ACE Surgical Infinity internal hex dental implant has 139 adverse event reports, with no deaths and 139 injuries. Most describe the implant failing to fuse with the bone, with some reports citing poor bone quality, smoking or hygiene, and others describing mobility or mild infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
139
Deaths reported
0
Injuries reported
139
Recalls naming it
0

Summary

The FDA's MAUDE database contains 139 adverse event reports that list this exact device (DI 00614950004621), received between Dec 22, 2015 and Jun 27, 2025 — classified as 139 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (109 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem109
  • Failure to Osseointegrate99
  • Insufficient Device Problem Information12
  • Loss of Osseointegration9
  • Patient Device Interaction Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Device Issue1

Patient problems

  • Failure of Implant139
  • Pain37
  • Unspecified Infection9
  • Swelling8
  • Numbness7
  • Fistula3

Recent adverse event reports

The 25 most recent of 139 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 27, 2025Failure to Osseointegrate; Failure to Osseointegrate

    FAILURE TO OSSEOINTEGRATE.

    Patient problems: Failure of Implant

    MAUDE report 22345186 ↗
  • InjuryFeb 3, 2025Loss of Osseointegration; Loss of Osseointegration

    BECAME MOBILE WITH SLIGHT INFECTION

    Patient problems: Failure of Implant

    MAUDE report 21297942 ↗
  • InjuryJun 5, 2024Loss of Osseointegration; Loss of Osseointegration

    LOST OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 19472458 ↗
  • InjuryMay 20, 2024Loss of Osseointegration; Loss of Osseointegration

    IMPLANT LOSS DUE TO POOR BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 19350435 ↗
  • InjuryApr 2, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT WAS SPININING AND SLIGHT INFECTION.

    Patient problems: Failure of Implant

    MAUDE report 19023289 ↗
  • InjuryJan 9, 2024Osseointegration Problem; Osseointegration Problem

    IMPLANT LOSS DUE TO POOR HYGEINE AND HEAVY SMOKING

    Patient problems: Failure of Implant

    MAUDE report 18475803 ↗
  • InjuryJan 5, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT LOSS DUE TO POOR BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 18451053 ↗
  • InjuryJan 4, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT LOSS DUE TO POOR BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 18447380 ↗
  • InjuryOct 19, 2023Failure to Osseointegrate; Failure to Osseointegrate

    NO OSSEOINTEGRATION AND SLIGHT INFECTION.

    Patient problems: Failure of Implant

    MAUDE report 17969577 ↗
  • InjuryAug 11, 2023Loss of Osseointegration; Loss of Osseointegration

    INADEQUATE BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 17513685 ↗
  • InjuryAug 10, 2023Loss of Osseointegration; Loss of Osseointegration

    POOR BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 17499576 ↗
  • InjuryAug 8, 2023Loss of Osseointegration; Loss of Osseointegration

    POOR BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 17484135 ↗
  • InjuryJul 24, 2023Loss of Osseointegration; Loss of Osseointegration

    BONE LOSS.

    Patient problems: Failure of Implant

    MAUDE report 17385981 ↗
  • InjuryDec 22, 2021Failure to Osseointegrate; Failure to Osseointegrate

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS NOT ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. MOBILITY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13059851 ↗
  • InjuryOct 21, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    PERI-IMPLANTITIS AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE (ALLOGRAFT) WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12677155 ↗
  • InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    PAIN, BLEEDING, SWELLING, INFECTION AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain; Unspecified Infection

    MAUDE report 12567733 ↗
  • InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY, INFECTION AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 12567713 ↗
  • InjuryJul 21, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12201846 ↗
  • InjuryJun 1, 2021Adverse Event Without Identified Device or Use Problem

    BONE QUALITY WAS NOT IDENTIFIED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS NOT IDENTIFIED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS NOT ACHIEVED AT SURGERY. OSSEOINTEGRATION AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11913601 ↗
  • InjuryJun 1, 2021Failure to Osseointegrate

    BONE QUALITY WAS NOT IDENTIFIED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED AT THE TIME OF SURGERY BUT IMPLANT DID NOT INTEGRATE WITH BONE AT RESTORATION. OSSEOINTEGRATION AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11913475 ↗
  • InjuryApr 22, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11711803 ↗
  • InjuryApr 22, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11711794 ↗
  • InjuryMar 29, 2021Loss of Osseointegration; Loss of Osseointegration

    LOSS OF OSSEOINTEGRATION, PAIN AND BLEEDING WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Pain

    MAUDE report 11581168 ↗
  • InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant

    MAUDE report 11423156 ↗
  • InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant

    MAUDE report 11423149 ↗

Other IMPLANT, ENDOSSEOUS, ROOT-FORM devices