Infinity Internal Hex System 203713Recalls & reported problems

Ace Surgical Supply CO., Inc. · infinity INTERNAL 3.7x13MM Ø3.5HD HEX IMPLANT 6AL-4V ELI · Class 2 · Implant

All models of Infinity Internal Hex System (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This ACE Surgical Infinity internal hex implant is a small endosseous screw used in dental implant procedures. The 75 reports describe the implant failing to fuse with the surrounding bone, becoming loose or mobile, along with pain, numbness, swelling, infection and fistula. There were no deaths, and 75 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
75
Deaths reported
0
Injuries reported
75
Recalls naming it
0

Summary

The FDA's MAUDE database contains 75 adverse event reports that list this exact device (DI 00614950004652), received between Dec 22, 2015 and Dec 13, 2024 — classified as 75 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (57 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem57
  • Failure to Osseointegrate42
  • Insufficient Device Problem Information8
  • Patient Device Interaction Problem4
  • Device Operates Differently Than Expected1
  • Loss of Osseointegration1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • Failure of Implant75
  • Pain18
  • Unspecified Infection3
  • Fistula2
  • Numbness2
  • Swelling1

Recent adverse event reports

The 25 most recent of 75 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 13, 2024Loss of Osseointegration; Loss of Osseointegration

    IMPLANT BECAME LOOSE.

    Patient problems: Failure of Implant

    MAUDE report 20925047 ↗
  • InjuryOct 10, 2023Failure to Osseointegrate; Failure to Osseointegrate

    SAME DAY SPINNER.

    Patient problems: Failure of Implant

    MAUDE report 17904845 ↗
  • InjuryOct 5, 2023Failure to Osseointegrate; Failure to Osseointegrate

    SAME DAY SPINNER.

    Patient problems: Failure of Implant

    MAUDE report 17882423 ↗
  • InjuryMay 19, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    SAME DAY SPINNER.

    Patient problems: Failure of Implant

    MAUDE report 16965466 ↗
  • InjuryMar 22, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    SAME DAY SPINNER.

    Patient problems: Failure of Implant

    MAUDE report 16591756 ↗
  • InjuryDec 22, 2021Failure to Osseointegrate; Therapeutic or Diagnostic Output Failure

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. MOBILITY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13059850 ↗
  • InjuryApr 12, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11648038 ↗
  • InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain

    MAUDE report 11423158 ↗
  • InjuryFeb 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND ASYMPTOMATIC WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11282544 ↗
  • InjuryJan 14, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAD RADIATION TX-HEAD/NECK AREA.

    Patient problems: Failure of Implant

    MAUDE report 11170354 ↗
  • InjuryNov 8, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 10806310 ↗
  • InjuryNov 3, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 10777662 ↗
  • InjuryNov 3, 2020Osseointegration Problem; Patient Device Interaction Problem; Osseointegration Problem; Patient Device Interaction Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 10777425 ↗
  • InjuryApr 2, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY, PAIN AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Pain

    MAUDE report 9916868 ↗
  • InjuryMar 27, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER AND HAS A DRUG/ALCOHOL ABUSE.

    Patient problems: Failure of Implant

    MAUDE report 9889701 ↗
  • InjuryFeb 20, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND NUMBNESS WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Numbness

    MAUDE report 9733030 ↗
  • InjuryJan 31, 2020Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. PAIN, BLEEDING, SWELLING, NUMBNESS AND INFECTION WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. PATIENT HAD RADIATION HEAD/NECK AREA.

    Patient problems: Failure of Implant; Pain; Swelling; Unspecified Infection

    MAUDE report 9655400 ↗
  • InjuryNov 15, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND PAIN WERE REPORTED. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE (NUOSS) WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Pain

    MAUDE report 9330667 ↗
  • InjurySep 9, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8975127 ↗
  • InjurySep 3, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Pain

    MAUDE report 8954907 ↗
  • InjurySep 3, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant

    MAUDE report 8954903 ↗
  • InjurySep 3, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Pain

    MAUDE report 8954902 ↗
  • InjuryJun 27, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8738608 ↗
  • InjuryJun 20, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 8715607 ↗
  • InjuryJun 11, 2019Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY, PAIN, PERI-IMPLANTITIS AND INFECTION WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO 1 YEAR AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Pain; Unspecified Infection

    MAUDE report 8686648 ↗