Infinity Internal Hex System 204111Recalls & reported problems

Ace Surgical Supply CO., Inc. · infinity INTERNAL 4.1x11.5MM Ø3.5HD HEX IMPLANT 6AL-4V ELI · Class 2 · Implant

All models of Infinity Internal Hex System (24)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This ACE Surgical Infinity internal hex dental implant has 138 adverse event reports, with no deaths and 138 injuries. Most describe the implant failing to fuse with the bone, with some reports citing poor bone quality, pain and infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
138
Deaths reported
0
Injuries reported
138
Recalls naming it
0

Summary

The FDA's MAUDE database contains 138 adverse event reports that list this exact device (DI 00614950004706), received between Nov 24, 2015 and Jun 6, 2025 — classified as 138 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (109 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem109
  • Failure to Osseointegrate97
  • Insufficient Device Problem Information9
  • Loss of Osseointegration7
  • Patient Device Interaction Problem5
  • Therapeutic or Diagnostic Output Failure2
  • Microbial Contamination of Device1
  • Missing Value Reason1

Patient problems

  • Failure of Implant138
  • Pain29
  • Unspecified Infection10
  • Numbness8
  • Fistula2
  • Swelling2

Recent adverse event reports

The 25 most recent of 138 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 6, 2025Loss of Osseointegration; Loss of Osseointegration

    IMPLANT #9 SPINNED WITH SLIGHT INFECTION.

    Patient problems: Failure of Implant

    MAUDE report 22158918 ↗
  • InjuryJan 5, 2024Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT LOSS DUE TO POOR BONE QUALITY.

    Patient problems: Failure of Implant

    MAUDE report 18454598 ↗
  • InjuryOct 19, 2023Failure to Osseointegrate; Failure to Osseointegrate

    NO OSSEOINTEGRATION.

    Patient problems: Failure of Implant

    MAUDE report 17969588 ↗
  • InjuryAug 22, 2023Microbial Contamination of Device; Microbial Contamination of Device

    CONTAMINATION OF STERILE IMPLANT SURFACE.

    Patient problems: Failure of Implant

    MAUDE report 17599488 ↗
  • InjuryAug 21, 2023Failure to Osseointegrate; Failure to Osseointegrate

    NO PRIMARY STABILITY ON INSERTION.

    Patient problems: Failure of Implant

    MAUDE report 17586663 ↗
  • InjuryAug 9, 2023Loss of Osseointegration; Loss of Osseointegration

    PAIN & MOBILITY 1.5 YEARS AFTER FINAL RESTORATION.

    Patient problems: Failure of Implant; Pain

    MAUDE report 17494308 ↗
  • InjuryAug 1, 2023Loss of Osseointegration; Loss of Osseointegration

    TYPE OF BONE.

    Patient problems: Failure of Implant

    MAUDE report 17438633 ↗
  • InjuryJul 13, 2023Failure to Osseointegrate; Loss of Osseointegration; Osseointegration Problem; Patient Device Interaction Problem; Failure to Osseointegrate; Loss of Osseointegration; Osseointegration Problem; Patient Device Interaction Problem

    NO INTERGRATION.

    Patient problems: Failure of Implant

    MAUDE report 17316398 ↗
  • InjuryDec 22, 2021Failure to Osseointegrate; Therapeutic or Diagnostic Output Failure

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS NOT ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. MOBILITY WAS REPORTED.

    Patient problems: Failure of Implant; Unspecified Infection

    MAUDE report 13059863 ↗
  • InjuryDec 22, 2021Therapeutic or Diagnostic Output Failure

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS UP TO ONE TO FIVE YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS ALSO ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. MOBILITY WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 13059856 ↗
  • InjuryDec 13, 2021Failure to Osseointegrate; Failure to Osseointegrate

    MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE IS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED.

    Patient problems: Failure of Implant

    MAUDE report 12984650 ↗
  • InjuryOct 18, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE IV. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12647541 ↗
  • InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    PAIN, PERI-IMPLANTITIS , BLEEDING, INFECTION, NUMBNESS AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Numbness; Pain

    MAUDE report 12567722 ↗
  • InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant

    MAUDE report 12567719 ↗
  • InjuryOct 4, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    PAIN, PERI-IMPLANTITIS, BLEEDING, INFECTION AND INADEQUATE BONE QUALITY/BONE QUANTITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE I. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS 1 TO 5 YEARS AFTER PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant; Numbness; Pain; Unspecified Infection

    MAUDE report 12567710 ↗
  • InjuryAug 6, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 12290901 ↗
  • InjuryAug 6, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY AND PAIN WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT HAS ALLERGIES (PENICILLIN) AND IS A SMOKER.

    Patient problems: Failure of Implant; Pain

    MAUDE report 12290860 ↗
  • InjuryJun 21, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    MOBILITY WAS REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant

    MAUDE report 12035128 ↗
  • InjuryJun 8, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    BONE QUALITY WAS IDENTIFIED AS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS PERFORMED AT TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11956887 ↗
  • InjuryJun 2, 2021Failure to Osseointegrate

    BONE QUALITY WAS TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS NOT ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11923264 ↗
  • InjuryMay 28, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    BONE QUALITY WAS TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED AT SURGERY. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11904042 ↗
  • InjuryMay 25, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    BONE QUALITY AT THE TIME OF IMPLANT FAILURE WAS POOR (III). TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11879187 ↗
  • InjuryApr 28, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY AND OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11737601 ↗
  • InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/BONE QUANTITY AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE II. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY.

    Patient problems: Failure of Implant

    MAUDE report 11423216 ↗
  • InjuryMar 5, 2021Failure to Osseointegrate; Osseointegration Problem; Failure to Osseointegrate; Osseointegration Problem

    INADEQUATE BONE QUALITY/ BONE QUANTITY, NUMBNESS AND MOBILITY WERE REPORTED. BONE QUALITY AT THE TIME OF IMPLANT FAILURE TYPE III. TIME OF LOSS IN RELATION TO THE IMPLANTATION WAS BEFORE PROSTHETIC RESTORATION. PRIMARY STABILITY WAS ACHIEVED. OSSEOINTEGRATION WAS NOT ACHIEVED. AUGMENTATION PROCEDURE WAS NOT PERFORMED AT THE TIME OF SURGERY. PATIENT IS A SMOKER.

    Patient problems: Failure of Implant; Numbness

    MAUDE report 11423188 ↗